Right ventricular migration of a recovery IVC filter's fractured wire with subsequent pericardial tamponade

被引:34
作者
Saeed, Imran
Garcia, Mark
McNicholas, Kathleen
机构
[1] Christiana Hosp, MAP 2, Newark, DE 19713 USA
[2] Christiana Hosp, Dept Radiol, Newark, DE 19713 USA
[3] Christiana Hosp, MAP 1, Newark, DE 19713 USA
关键词
filter complication; filter fracture; interventional radiology; pericardial tamponade; retrievable IVC filters;
D O I
10.1007/s00270-005-0136-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A Recovery filter (C.R. Bard, Tempe, AZ, USA) is a device for pulmonary embolism prophylaxis. There have been few case reports involving the migration of this particular filter or of a broken wire migrating to the heart. We report a case of right ventricular migration of a fractured wire from this filter in a patient who subsequently developed pericardial tamponade and required open heart surgery to extract the fractured wire. We discuss the current US Food and Drug Administration (FDA)-approved nonpermanent inferior vena cava filters and their reported complications. These complications can be life-threatening and may require immediate surgical intervention.
引用
收藏
页码:685 / 686
页数:2
相关论文
共 10 条
[1]   Initial experience in humans with a new retrievable inferior vena cava filter [J].
Asch, MR .
RADIOLOGY, 2002, 225 (03) :835-844
[2]   Right atrial migration and percutaneous retrieval of a Gunther Tulip inferior vena cava filter [J].
Bochenek, KM ;
Aruny, JE ;
Tal, MG .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (09) :1207-1209
[3]   A new optional vena cava filter: Retrieval at 12 weeks in an animal model [J].
Brountzos, EN ;
Kaufman, JA ;
Venbrux, AC ;
Brown, R ;
Harry, J ;
Kinst, TF ;
Kleshinski, S ;
Ravenscroft, AC .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (06) :763-772
[4]  
Izutani H, 2004, CAN J CARDIOL, V20, P233
[5]  
Kercher K, 2003, AM SURGEON, V69, P643
[6]   Indications, management, and complications of temporary inferior vena cava filters [J].
Linsenmaier, U ;
Rieger, J ;
Schenk, F ;
Rock, C ;
Mangel, E ;
Pfeifer, KJ .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1998, 21 (06) :464-469
[7]   The role of temporary inferior vena cava filters in critically ill surgical inferior patients [J].
Offner, PJ ;
Hawkes, A ;
Madayag, R ;
Seale, F ;
Maines, C .
ARCHIVES OF SURGERY, 2003, 138 (06) :591-594
[8]   Intracardiac migration of nitinol TrapEase™ vena cava filter and paradoxical embolism [J].
Porcellini, M ;
Stassano, P ;
Musumeci, A ;
Bracale, G .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2002, 22 (03) :460-461
[9]   Fatal outcome in atrial migration of the tempofilter [J].
Rossi, P ;
Arata, FM ;
Bonaiuti, P ;
Pedicini, V .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1999, 22 (03) :227-231
[10]   Outcome and complications of retrievable inferior vona cava filters [J].
Stein, PD ;
Nas, MA ;
Skaf, E ;
Kayali, F ;
Siddiqui, T ;
Olson, RE ;
Patel, K .
AMERICAN JOURNAL OF CARDIOLOGY, 2004, 94 (08) :1090-1093