Accuracy of CGM Systems During Continuous and Interval Exercise in Adults with Type 1 Diabetes

被引:12
作者
Da Prato, G. [1 ]
Pasquini, S. [1 ]
Rinaldi, E. [1 ]
Lucianer, T. [1 ]
Dona, S. [1 ]
Santi, L. [1 ]
Negri, C. [1 ]
Bonora, E. [1 ]
Moghetti, P. [1 ]
Trombetta, M. [1 ]
机构
[1] Univ & Hosp Verona, Div Endocrinol Diabet & Metab, Dept Med, Verona, Italy
关键词
CGM; accuracy; exercise; T1D; DEXCOM G4 PLATINUM; PHYSICAL-EXERCISE; AEROBIC EXERCISE; GLUCOSE; SENSOR;
D O I
10.1177/19322968211023522
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: continuous glucose monitoring systems (CGMs) play an important role in the management of T1D, but their accuracy may reduce during rapid glucose excursions. The aim of study was to assess the accuracy of recent rt-CGMs available in Italy, in subjects with T1D during 2 sessions of physical activity: moderate continuous (CON) and interval exercise (IE). Method: we recruited 22 patients with T1D, on CSII associated or integrated with a CGM, to which a second different sensor was applied. Data recorded by CGMs were compared with the corresponding plasma glucose (PG) values, measured every 5minutes with the glucose analyzer. To assess the accuracy of the CGMs, we evaluated the Sensor Bias (SB), the Mean Absolute Relative Difference (MARD) and the Clarke error grid (CEG). Results: a total of 2355 plasma-sensor glucose paired points were collected. Both average plasma and interstitial glucose concentrations did not significantly differ during CON and IE. During CON: 1. PG change at the end of exercise was greater than during IE (P = .034); 2. all sensors overestimated PG more than during IE, as shown by SB (P < .001) and MARD (P < .001) comparisons. Classifying the performance according to the CEG, significant differences were found between the 2 sessions in distribution of points in A and B zones. Conclusions: the exercise affects the accuracy of currently available CGMs, especially during CON, suggesting, in this circumstance, the need to maintain blood glucose in a "prudent" range, above that generally recommended. Further studies are needed to investigate additional types of activities.
引用
收藏
页码:1436 / 1443
页数:8
相关论文
共 30 条
[1]   Continuous glucose monitoring system during physical exercise in adolescents with type 1 diabetes [J].
Adolfsson, P. ;
Nilsson, S. ;
Lindblad, B. .
ACTA PAEDIATRICA, 2011, 100 (12) :1603-1609
[2]   Evaluation of glucose control when a new strategy of increased carbohydrate supply is implemented during prolonged physical exercise in type 1 diabetes [J].
Adolfsson, Peter ;
Mattsson, Stig ;
Jendle, Johan .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2015, 115 (12) :2599-2607
[3]   Accuracy of continuous glucose monitoring during differing exercise conditions [J].
Bally, Lia ;
Zueger, Thomas ;
Pasi, Nicola ;
Carlos, Ciller ;
Paganini, Daniela ;
Stettler, Christoph .
DIABETES RESEARCH AND CLINICAL PRACTICE, 2016, 112 :1-5
[4]  
Basu Ananda, 2015, J Diabetes Sci Technol, V9, P63, DOI 10.1177/1932296814554797
[5]   Accuracy of Continuous Glucose Monitoring before, during, and after Aerobic and Anaerobic Exercise in Patients with Type 1 Diabetes Mellitus [J].
Biagi, Lyvia ;
Bertachi, Arthur ;
Quiros, Carmen ;
Gimenez, Marga ;
Conget, Ignacio ;
Bondia, Jorge ;
Vehi, Josep .
BIOSENSORS-BASEL, 2018, 8 (01)
[6]   EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE [J].
CLARKE, WL ;
COX, D ;
GONDERFREDERICK, LA ;
CARTER, W ;
POHL, SL .
DIABETES CARE, 1987, 10 (05) :622-628
[7]   Physical Activity/Exercise and Diabetes: A Position Statement of the American Diabetes Association [J].
Colberg, Sheri R. ;
Sigal, Ronald J. ;
Yardley, Jane E. ;
Riddell, Michael C. ;
Dunstan, David W. ;
Dempsey, Paddy C. ;
Horton, Edward S. ;
Castorino, Kristin ;
Tate, Deborah F. .
DIABETES CARE, 2016, 39 (11) :2065-2079
[8]   Effect of exercise-mediated acidosis on determination of glycemia using CGMS [J].
Davey, Raymond J. ;
Ferreira, Luis D. ;
Jones, Timothy W. ;
Fournier, Paul A. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2006, 8 (04) :516-518
[9]   Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring [J].
Deis, Dorothee ;
Bolinder, Jan ;
Riveline, Jean-Pierre ;
Battelino, Tadej ;
Bosi, Emanuele ;
Tubiana-Rufi, Nadia ;
Kerr, David ;
Phillip, Moshe .
DIABETES CARE, 2006, 29 (12) :2730-2732
[10]   Continuous Glucose Monitoring Sensors: Past, Present and Future Algorithmic Challenges [J].
Facchinetti, Andrea .
SENSORS, 2016, 16 (12)