An integrated system for planning, navigation and robotic assistance for skull base surgery

被引:106
作者
Xia, Tian
Baird, Clint [2 ]
Jallo, George [2 ]
Hayes, Kathryn [3 ]
Nakajima, Nobuyuki [3 ]
Hata, Nobuhiko [3 ]
Kazanzides, Peter [1 ]
机构
[1] Johns Hopkins Univ, CSEB 120, Dept Comp Sci, Baltimore, MD 21218 USA
[2] Johns Hopkins Med Inst, Dept Neurosurg, Baltimore, MD 21205 USA
[3] Brigham & Womens Hosp, Surg Planning Lab, Cambridge, MA USA
关键词
medical robotics; skull base surgery; image-guided surgery;
D O I
10.1002/rcs.213
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background We developed an image-guided robot system to provide mechanical assistance for skull base drilling, which is performed to gain access for some neurosurgical interventions, such as tumour resection. The motivation for introducing this robot was to improve safety by preventing the surgeon from accidentally damaging critical neurovascular structures during the drilling procedure. Methods We integrated a Stealthstation (R) navigation system, a NeuroMate (R) robotic arm with a six-degree-of-freedom force sensor, and the 3D Slicer visualization software to allow the robotic arm to be used in a navigated, cooperatively-controlled fashion by the surgeon. We employed virtual fixtures to constrain the motion of the robot-held cutting tool, so that it remained in the safe zone that was defined on a preoperative CT scan. Results We performed experiments on both foam skull and cadaver heads. The results for foam blocks cut using different registrations yielded an average placement error of 0.6 mm and an average dimensional error of 0.6 mm. We drilled the posterior porus acusticus in three cadaver heads and concluded that the robot-assisted procedure is clinically feasible and provides some ergonomic benefits, such as stabilizing the drill. We obtained postoperative CT scans of the cadaver heads to assess the accuracy and found that some bone outside the virtual fixture boundary was cut. The typical overcut was 1-2 mm, with a maximum overcut of about 3 mm. Conclusions The image-guided cooperatively-controlled robot system can improve the safety and ergonomics of skull base drilling by stabilizing the drill and enforcing virtual fixtures to protect critical neurovascular structures. The next step is to improve the accuracy so that the overcut can be reduced to a more clinically acceptable value of about 1 mm. Copyright (C) 2008 John Wiley & Sons, Ltd.
引用
收藏
页码:321 / 330
页数:10
相关论文
共 25 条
  • [1] Virtual fixture architectures for telemanipulation
    Abbott, JJ
    Okamura, AM
    [J]. 2003 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS 1-3, PROCEEDINGS, 2003, : 2798 - 2805
  • [2] LEAST-SQUARES FITTING OF 2 3-D POINT SETS
    ARUN, KS
    HUANG, TS
    BLOSTEIN, SD
    [J]. IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1987, 9 (05) : 699 - 700
  • [3] Brisson G., 2006, COMPUT ASSIST ORTHOP, P72
  • [4] An automated robotic approach with redundant navigation for minimal invasive extended transsphenoidal skull base surgery
    Bumm, K
    Wurm, J
    Rachinger, J
    Dannenmann, T
    Bohr, C
    Fahlbusch, R
    Iro, H
    Nimsky, C
    [J]. MINIMALLY INVASIVE NEUROSURGERY, 2005, 48 (03) : 159 - 164
  • [5] CHAN F, 2007, 3 AS C COMP AID SURG
  • [6] DEVOS P, 2005, COMPUT ASSIST ORTHOP, P88
  • [7] Development of the first force-controlled robot for otoneurosurgery
    Federspil, PA
    Geisthoff, UW
    Henrich, D
    Plinkert, PK
    [J]. LARYNGOSCOPE, 2003, 113 (03) : 465 - 471
  • [8] HAIDEGGER T, 2008, IEEE INT C IN PRESS
  • [9] Handini D, 2004, 2004 IEEE CONFERENCE ON ROBOTICS, AUTOMATION AND MECHATRONICS, VOLS 1 AND 2, P43
  • [10] The hands-on orthopaedic robot "Acrobot": Early clinical trials of total knee replacement surgery
    Jakopec, M
    Rodriguez y Baena, F
    Harris, SJ
    Gomes, P
    Cobb, J
    Davies, BL
    [J]. IEEE TRANSACTIONS ON ROBOTICS AND AUTOMATION, 2003, 19 (05): : 902 - 911