Safety for Human MR Scanners at 7T

被引:7
作者
Okada, Tomohisa [1 ,2 ]
Akasaka, Thai [1 ]
Thuy, Dinh H. D. [1 ]
Isa, Tadashi [1 ]
机构
[1] Kyoto Univ, Human Brain Res Ctr, Grad Sch Med, Kyoto, Kyoto, Japan
[2] Kyoto Univ, Human Brain Res Ctr, Grad Sch Med, 54 Shogoin Kawaharacho,Sakyo ku, Kyoto, Kyoto 6068507, Japan
关键词
safety; 7; tesla; magnetic resonance imaging; DOUBLE INVERSION-RECOVERY; STATIC MAGNETIC-FIELD; TEMPORAL-LOBE EPILEPSY; DEEP GRAY-MATTER; FOCUS LATERALITY; GRAVID UTERUS; EXPOSURE; BRAIN; TESLA; BODY;
D O I
10.2463/mrms.rev.2021-0063
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
After introduction of the first human 7 tesla (7T) system in 1999, 7T MR systems have been employed as one of the most advanced platforms for human MR research for more than 20 years. Currently, two 7T MR models are approved for clinical use in the U.S.A. The approval facilitated introduction of the 7T system, summing up to around 100 worldwide. The approval in Japan is much awaited. As a clinical MR scanner, the 7T MR system is drawing attention in terms of safety.Several large-sized studies on bioeffects have been reported for vertigo, dizziness, motion disturbances, nausea, and others. Such effects might also be found in MR workers and researchers. Frequency and severity of reported bioeffects will be presented and discussed, including their variances. The high resonance frequency and shorter RF wavelength of 7T increase the concern about the safety. Homogeneous RF pulse excitation is difficult even for the brain, and a multi-channel parallel transmit (pTx) system is considered mandatory. However, pTx may create a hot spot, which makes the estimation of specific absorption rate (SAR) to be difficult. The stronger magnetic field of 7T causes a large force of displacement and heating on metallic implants or devices, and the scan of patients with them should not be conducted at 7T. However, there are some opinions that such patients might be scanned even at 7T, if certain criteria are met. This article provides a brief review on the effect of the static magnetic field on humans (MR subjects, workers, and researchers) and neurons, in addition to scan sound, SAR, and metal implants and devices. Understanding and avoiding adverse effects will contribute to the reduction in safety risks and the prevention of incidents.
引用
收藏
页码:531 / 537
页数:7
相关论文
共 50 条
  • [31] Evaluation of Intracranial Aneurysms with 7T versus 1.5T Time-of-Flight MR Angiography - Initial Experience
    Moenninghoff, C.
    Maderwald, S.
    Theysohn, J. M.
    Kraff, O.
    Ladd, S. C.
    Ladd, M. E.
    Forsting, M.
    Quick, H. H.
    Wanke, I.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2009, 181 (01): : 16 - 23
  • [32] Short Echo Spectroscopic Imaging of the Human Brain at 7T Using Transceiver Arrays
    Avdievich, N. I.
    Pan, J. W.
    Baehring, J. M.
    Spencer, D. D.
    Hetherington, H. P.
    MAGNETIC RESONANCE IN MEDICINE, 2009, 62 (01) : 17 - 25
  • [33] Measurements of RF power reflected and radiated by multichannel transmit MR coils at 7T
    Gerd Weidemann
    Frank Seifert
    Werner Hoffmann
    Harald Pfeiffer
    Reiner Seemann
    Bernd Ittermann
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2016, 29 : 371 - 378
  • [34] High SNR full brain relaxometry at 7T by accelerated MR-STAT
    Versteeg, Edwin
    Liu, Hongyan
    van der Heide, Oscar
    Fuderer, Miha
    van den Berg, Cornelis A. T.
    Sbrizzi, Alessandro
    MAGNETIC RESONANCE IN MEDICINE, 2024, 92 (01) : 226 - 235
  • [35] SWI by 7T MR Imaging for the Microscopic Imaging Diagnosis of Astrocytic and Oligodendroglial Tumors
    Natsumeda, M.
    Matsuzawa, H.
    Watanabe, M.
    Motohashi, K.
    Gabdulkhaev, R.
    Tsukamoto, Y.
    Kanemaru, Y.
    Watanabe, J.
    Ogura, R.
    Okada, M.
    Kurabe, S.
    Okamoto, K.
    Kakita, A.
    Igarashi, H.
    Fujii, Y.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2022, 43 (11) : 1575 - 1581
  • [36] Subjective perception of safety in healthy individuals working with 7 T MRI scanners: a retrospective multicenter survey
    Fatahi, Mahsa
    Demenescu, Liliana Ramona
    Speck, Oliver
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2016, 29 (03) : 379 - 387
  • [37] In vivo 1D and 2D correlation MR spectroscopy of the soleus muscle at 7T
    Ramadan, Saadallah
    Ratai, Eva-Maria
    Wald, Lawrence L.
    Mountford, Carolyn E.
    JOURNAL OF MAGNETIC RESONANCE, 2010, 204 (01) : 91 - 98
  • [38] Measurements of RF power reflected and radiated by multichannel transmit MR coils at 7T
    Weidemann, Gerd
    Seifert, Frank
    Hoffmann, Werner
    Pfeiffer, Harald
    Seemann, Reiner
    Ittermann, Bernd
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2016, 29 (03) : 371 - 378
  • [39] Safety issues and updates under MR environments
    Kim, Soo Jung
    Kim, Kyung Ah
    EUROPEAN JOURNAL OF RADIOLOGY, 2017, 89 : 7 - 13
  • [40] Intracranial atherosclerosis on 7T MRI and cognitive functioning The SMART-MR study
    Zwartbol, Maarten H. T.
    van der Kolk, Anja G.
    Ghaznawi, Rashid
    van der Graaf, Yolanda
    Hendrikse, Jeroen
    Geerlings, Mirjam, I
    NEUROLOGY, 2020, 95 (10) : E1351 - E1361