Safety for Human MR Scanners at 7T

被引:7
作者
Okada, Tomohisa [1 ,2 ]
Akasaka, Thai [1 ]
Thuy, Dinh H. D. [1 ]
Isa, Tadashi [1 ]
机构
[1] Kyoto Univ, Human Brain Res Ctr, Grad Sch Med, Kyoto, Kyoto, Japan
[2] Kyoto Univ, Human Brain Res Ctr, Grad Sch Med, 54 Shogoin Kawaharacho,Sakyo ku, Kyoto, Kyoto 6068507, Japan
关键词
safety; 7; tesla; magnetic resonance imaging; DOUBLE INVERSION-RECOVERY; STATIC MAGNETIC-FIELD; TEMPORAL-LOBE EPILEPSY; DEEP GRAY-MATTER; FOCUS LATERALITY; GRAVID UTERUS; EXPOSURE; BRAIN; TESLA; BODY;
D O I
10.2463/mrms.rev.2021-0063
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
After introduction of the first human 7 tesla (7T) system in 1999, 7T MR systems have been employed as one of the most advanced platforms for human MR research for more than 20 years. Currently, two 7T MR models are approved for clinical use in the U.S.A. The approval facilitated introduction of the 7T system, summing up to around 100 worldwide. The approval in Japan is much awaited. As a clinical MR scanner, the 7T MR system is drawing attention in terms of safety.Several large-sized studies on bioeffects have been reported for vertigo, dizziness, motion disturbances, nausea, and others. Such effects might also be found in MR workers and researchers. Frequency and severity of reported bioeffects will be presented and discussed, including their variances. The high resonance frequency and shorter RF wavelength of 7T increase the concern about the safety. Homogeneous RF pulse excitation is difficult even for the brain, and a multi-channel parallel transmit (pTx) system is considered mandatory. However, pTx may create a hot spot, which makes the estimation of specific absorption rate (SAR) to be difficult. The stronger magnetic field of 7T causes a large force of displacement and heating on metallic implants or devices, and the scan of patients with them should not be conducted at 7T. However, there are some opinions that such patients might be scanned even at 7T, if certain criteria are met. This article provides a brief review on the effect of the static magnetic field on humans (MR subjects, workers, and researchers) and neurons, in addition to scan sound, SAR, and metal implants and devices. Understanding and avoiding adverse effects will contribute to the reduction in safety risks and the prevention of incidents.
引用
收藏
页码:531 / 537
页数:7
相关论文
共 50 条
  • [1] Assessment of Sensations Experienced by Subjects during MR Imaging Examination at 7T
    Uwano, Ikuko
    Metki, Tsuyoshi
    Sendai, Fusako
    Yoshida, Ryoko
    Kudo, Kohsuke
    Yamashita, Fumio
    Higuchi, Satomi
    Ito, Kenji
    Harada, Taisuke
    Goodwin, Jonathan
    Ogawa, Akira
    Sasaki, Makoto
    MAGNETIC RESONANCE IN MEDICAL SCIENCES, 2015, 14 (01) : 35 - 41
  • [2] Shielded Microstrip Array for 7T Human MR Imaging
    Wu, Bing
    Wang, Chunsheng
    Kelley, Douglas A. C.
    Xu, Duan
    Vigneron, Daniel B.
    Nelson, Sarah J.
    Zhang, Xiaoliang
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (01) : 179 - 184
  • [3] Insertable inductively coupled volumetric coils for MR microscopy in a human 7T MR system
    Okada, Tomohisa
    Handa, Shinya
    Ding, Bill
    Urayama, Shin-ichi
    Fujimoto, Koji
    Shima, Atsushi
    Yoshii, Daisuke
    Ayaki, Takashi
    Sawamoto, Nobukatsu
    Takahashi, Ryosuke
    Onoe, Hirotaka
    Isa, Tadashi
    Petropoulos, Labros
    MAGNETIC RESONANCE IN MEDICINE, 2022, 87 (03) : 1613 - 1620
  • [4] Rapid in vivo musculoskeletal MR with parallel imaging at 7T
    Banerjee, Suchandrima
    Krug, Roland
    Carballido-Gamio, Julio
    Kelley, Douglas A. C.
    Xu, Duan
    Vigneron, Daniel B.
    Majumdar, Sharmila
    MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (03) : 655 - 660
  • [5] 7T Transmit/Receive Arrays Using ICE Decoupling for Human Head MR Imaging
    Yan, Xinqiang
    Zhang, Xiaoliang
    Feng, Baotong
    Ma, ChuangXin
    Wei, Long
    Xue, Rong
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2014, 33 (09) : 1781 - 1787
  • [6] Advantages of fluid and white matter suppression (FLAWS) with MP2RAGE compared with double inversion recovery turbo spin echo (DIR-TSE) at 7T
    Urushibata, Yuta
    Kuribayashi, Hideto
    Fujimoto, Koji
    Kober, Tobias
    Grinstead, John W.
    Isa, Tadashi
    Okada, Tomohisa
    EUROPEAN JOURNAL OF RADIOLOGY, 2019, 116 : 160 - 164
  • [7] Impact of the Static and Radiofrequency Magnetic Fields Produced by a 7T MR Imager on Metallic Dental Materials
    Oriso, Kenta
    Kobayashi, Takuya
    Sasaki, Makoto
    Uwano, Ikuko
    Kihara, Hidemichi
    Kondo, Hisatomo
    MAGNETIC RESONANCE IN MEDICAL SCIENCES, 2016, 15 (01) : 26 - 33
  • [8] A closer look: pediatric neuroimaging at 7T
    Barkovich, Emil Jernstedt
    Cortes-Albornoz, Maria Camila
    Machado-Rivas, Fedel
    Ferraciolli, Suely Fazio
    Afacan, Onur
    Jaimes, Camilo
    PEDIATRIC RADIOLOGY, 2025, : 1071 - 1087
  • [9] MR-safety in clinical practice at 7T: Evaluation of a multistep screening process in 1819 subjects
    Hansson, B.
    Simic, M.
    Olsrud, J.
    Bloch, K. Markenroth
    Owman, T.
    Sundgren, P. C.
    Bjorkman-Burtscher, I. M.
    RADIOGRAPHY, 2022, 28 (02) : 454 - 459
  • [10] 31P MR Spectroscopic Imaging Combined with 1H MR Spectroscopic Imaging in the Human Prostate Using a Double Tuned Endorectal Coil at 7T
    Luttje, Mariska P.
    Italiaander, Michel G. M.
    de Castro, Catalina S. Arteaga
    van der Kemp, Wybe J. M.
    Luijten, Peter R.
    van Vulpen, Marco
    van der Heide, Uulke A.
    Klomp, Dennis W. J.
    MAGNETIC RESONANCE IN MEDICINE, 2014, 72 (06) : 1516 - 1521