The utility of measurement uncertainty in medical laboratories

被引:72
作者
Braga, Federica [1 ,2 ]
Panteghini, Mauro [1 ]
机构
[1] Univ Milan, Res Ctr Metrol Traceabil Lab Med CIRME, Milan, Italy
[2] ASST Fatebenefratelli Sacco, UOC Patol Clin, Via GB Grassi 74, I-20157 Milan, Italy
关键词
measurement uncertainty; metrological traceability; performance specifications; standardization; METROLOGICAL TRACEABILITY; PERFORMANCE SPECIFICATIONS; QUALITY; STANDARDIZATION; IMPLEMENTATION; GOALS;
D O I
10.1515/cclm-2019-1336
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The definition and enforcement of reference measurement systems, based on the implementation of metrological traceability of patient results to higher-order (reference) methods and/or materials, together with a clinically acceptable level of measurement uncertainty (MU), are fundamental requirements to produce accurate and equivalent laboratory results. The MU associated with each step of the traceability chain should be governed to obtain a final combined MU on clinical samples fulfilling the requested performance specifications. MU is useful for a number of reasons: (a) for giving objective information about the quality of individual laboratory performance; (b) for serving as a management tool for the medical laboratory and in vitro diagnostics (IVD) manufacturers, forcing them to investigate and eventually fix the identified problems; (c) for helping those manufacturers that produce superior products and measuring systems to demonstrate the superiority of those products; (d) for identifying analytes that need analytical improvement for their clinical use and ask IVD manufacturers to work for improving the quality of assay performance and (e) for abandoning assays with demonstrated insufficient quality. Accordingly, the MU should not be considered a parameter to be calculated by medical laboratories just to fulfill accreditation standards, but it must become a key quality indicator to describe both the performance of an IVD measuring system and the laboratory itself.
引用
收藏
页码:1407 / 1413
页数:7
相关论文
共 34 条
[1]  
[Anonymous], 1002008 JCGM
[2]  
[Anonymous], 2019, Medical laboratories -practical guidance for the estimation of measurement uncertainty, V1st
[3]   State of Harmonization of 24 Serum Albumin Measurement Procedures and Implications for Medical Decisions [J].
Bachmann, Lorin M. ;
Yu, Min ;
Boyd, James C. ;
Bruns, David E. ;
Miller, W. Greg .
CLINICAL CHEMISTRY, 2017, 63 (03) :770-779
[4]   Defining acceptable limits for the metrological traceability of specific measurands [J].
Bais, Renze ;
Armbruster, Dave ;
Jansen, Rob T. P. ;
Klee, George ;
Panteghini, Mauro ;
Passarelli, Joseph ;
Sikaris, Ken A. .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (05) :973-979
[5]   Trueness evaluation and verification of inter-assay agreement of serum folate measuring systems [J].
Braga, Federica ;
Frusciante, Erika ;
Ferraro, Simona ;
Panteghini, Mauro .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2020, 58 (10) :1697-1705
[6]   Trueness Evaluation and Verification of Interassay Agreement of 11 Serum IgA Measuring Systems: Implications for Medical Decisions [J].
Braga, Federica ;
Infusino, Ilenia ;
Frusciante, Erika ;
Ceriotti, Ferruccio ;
Panteghini, Mauro .
CLINICAL CHEMISTRY, 2019, 65 (03) :473-483
[7]   The role of external quality assessment in the verification of in vitro medical diagnostics in the traceability era [J].
Braga, Federica ;
Pasqualetti, Sara ;
Panteghini, Mauro .
CLINICAL BIOCHEMISTRY, 2018, 57 :23-28
[8]   Defining permissible limits for the combined uncertainty budget in the implementation of metrological traceability [J].
Braga, Federica ;
Panteghini, Mauro .
CLINICAL BIOCHEMISTRY, 2018, 57 :7-11
[9]   Performance criteria for combined uncertainty budget in the implementation of metrological traceability [J].
Braga, Federica ;
Infusino, Ilenia ;
Panteghini, Mauro .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (06) :905-912
[10]   Verification of in vitro medical diagnostics (IVD) metrological traceability: Responsibilities and strategies [J].
Braga, Federica ;
Panteghini, Mauro .
CLINICA CHIMICA ACTA, 2014, 432 :55-61