Scheimpflug Corneal Power Measurements for Intraocular Lens Power Calculation in Cataract Surgery

被引:48
作者
Saad, Elie [2 ]
Shammas, Maya C. [2 ]
Shammas, H. John [1 ,2 ]
机构
[1] Univ So Calif, Keck Sch Med, Lynwood, CA USA
[2] Shammas Eye Med Ctr, Lynwood, CA USA
关键词
AUTOMATED KERATOMETRY; PHOTOGRAPHY; ACCURACY;
D O I
10.1016/j.ajo.2013.04.035
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To compare the keratometric (K) readings from the Pentacam-HR (High Resolution) unit with the automated K values from the IOLMaster keratometer (K-IOLM), and to evaluate them in the commonly used intraocular lens (IOL) power calculation formulas for routine cataract surgery. DESIGN: Prospective, comparative observational study. METHODS: SETTING: Private practice, Lynwood, California. STUDY POPULATION: Fifty cataractous eyes scheduled for surgery between July and August 2012. OBSERVATION PROCEDURE: The K readings from the Pentacam-HR unit taken at the 2-, 3-, 4-, and 5-mm zones and the 2-, 3-, 4-, and 5-mm rings, respectively, from 3 different maps: sagittal corneal front (K-F), true net power (K-TNP), and total refractive power (K-RP) are compared with K-IOLM. IOL power calculations were performed with each of the 25 sets of K readings. MAIN OUTCOME MEASURES: The IOL prediction median absolute error (MedAE) obtained with each measurement. RESULTS: K-F averaged 0.03-0.13 diopter (D) higher than K-IOLM (P > .05), K-TNP averaged 1.16-1.21 D lower than K-IOLM (P > .001), and K-RP averaged 0.23-0.72 D lower than K-IOLM (P > .001), with large variations in the measurements. The MedAE obtained with the different Pentacam K readings ranged from 0.44-0.64 D vs 0.52 D obtained with K-IOLM (P >.05). MedAE was lower in all categories when the pupil was 3 mm or smaller. CONCLUSION: The Pentacam K-F values were the closest to K-IOLM and the K-F readings from the 2-mm ring yielded the best results for IOL power calculation. ((C) (C) 2013 by Elsevier Inc. All rights reserved.)
引用
收藏
页码:460 / 467
页数:8
相关论文
共 17 条
[1]  
Dupont WD, 2009, PS POWER SAMPLE SIZE
[3]   Corneal power after refractive surgery for myopia: Contact lens method [J].
Haigis, W .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (07) :1397-1411
[4]   Comparison of spherical aberration and small-pupil profiles in improving depth of focus for presbyopic corrections [J].
Hickenbotham, Adam ;
Tiruveedhula, Pavan ;
Roorda, Austin .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (12) :2071-2079
[5]  
Hoffer KJ, 2007, J CATARACT REFR SURG, V33, P2, DOI 10.1016/j.jcrs.2006.08.054
[6]   A 3-PART SYSTEM FOR REFINING INTRAOCULAR-LENS POWER CALCULATIONS [J].
HOLLADAY, JT ;
PRAGER, TC ;
CHANDLER, TY ;
MUSGROVE, KH ;
LEWIS, JW ;
RUIZ, RS .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1988, 14 (01) :17-24
[7]   Repeatability, reproducibility, and agreement characteristics of rotating Scheimpflug photography and scanning-slit corneal topography for corneal power measurement [J].
Kawamorita, Takushi ;
Uozato, Hiroshi ;
Kamiya, Kazutaka ;
Bax, Leon ;
Tsutsui, Kenta ;
Aizawa, Daisuke ;
Shimizu, Kimiya .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (01) :127-133
[8]   Improved accuracy of intraocular lens power calculation with the Zeiss IOLMaster [J].
Olsen, Thomas .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2007, 85 (01) :84-87
[9]   Ray-tracing analysis of intraocular lens power in situ [J].
Olsen, Thomas ;
Funding, Mikkel .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (04) :641-647
[10]   DEVELOPMENT OF THE SRK/T INTRAOCULAR-LENS IMPLANT POWER CALCULATION FORMULA [J].
RETZLAFF, JA ;
SANDERS, DR ;
KRAFF, MC .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1990, 16 (03) :333-340