A broadband variable-temperature test system for complex permittivity measurements of solid and powder materials

被引:32
|
作者
Zhang, Yunpeng [1 ]
Li, En [1 ,2 ]
Zhang, Jing [1 ]
Yu, Chengyong [1 ]
Zheng, Hu [1 ]
Guo, Gaofeng [1 ]
机构
[1] Univ Elect Sci & Technol China, Sch Elect Engn, Chengdu 611731, Sichuan, Peoples R China
[2] Univ Elect Sci & Technol China, Natl Engn Res Ctr Electromagnet Radiat Control, Chengdu 610054, Sichuan, Peoples R China
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2018年 / 89卷 / 02期
基金
中国国家自然科学基金;
关键词
DIELECTRIC-PROPERTIES; MICROWAVE-FREQUENCIES; RAPID MEASUREMENTS; CAVITY; 1400-DEGREES-C; RF;
D O I
10.1063/1.4993507
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A microwave test system to measure the complex permittivity of solid and powder materials as a function of temperature has been developed. The system is based on a TM0n0 multi-mode cylindrical cavity with a slotting structure, which provides purer test modes compared to a traditional cavity. To ensure the safety, effectiveness, and longevity, heating and testing are carried out separately and the sample can move between two functional areas through an Alundum tube. Induction heating and a pneumatic platform are employed to, respectively, shorten the heating and cooling time of the sample. The single trigger function of the vector network analyzer is added to test software to suppress the drift of the resonance peak during testing. Complex permittivity is calculated by the rigorous field theoretical solution considering multilayer media loading. The variation of the cavity equivalent radius caused by the sample insertion holes is discussed in detail, and its influence to the test result is analyzed. The calibration method for the complex permittivity of the Alundum tube and quartz vial (for loading powder sample), which vary with the temperature, is given. The feasibility of the system has been verified by measuring different samples in a wide range of relative permittivity and loss tangent, and variable-temperature test results of fused quartz and SiO2 powder up to 1500 degrees C are compared with published data. The results indicate that the presented system is reliable and accurate. The stability of the system is verified by repeated and long-term tests, and error analysis is presented to estimate the error incurred due to the uncertainties in different error sources. Published by AIP Publishing.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Broadband Complex Permittivity Measurements of Microwave Interconnects and Bonding Materials
    Rahman, Nahid
    Korolev, Konstantin A.
    Afsar, Mohammed N.
    Cheung, Rudy
    Aghion, Maurice
    2008 33RD INTERNATIONAL CONFERENCE ON INFRARED, MILLIMETER AND TERAHERTZ WAVES, VOLS 1 AND 2, 2008, : 723 - +
  • [2] Broadband Measurements of the Complex Permittivity of Carbonaceous Asteroid Regolith Analog Materials
    Boivin, A. L.
    Hickson, D.
    Tsai, C.
    Cunje, A.
    Ghent, R. R.
    Daly, M. G.
    JOURNAL OF GEOPHYSICAL RESEARCH-PLANETS, 2018, 123 (12) : 3088 - 3104
  • [3] Broadband Measurements of Soil Complex Permittivity
    Stellini, Justin
    Farrugia, Lourdes
    Farhat, Iman
    Bonello, Julian
    Persico, Raffaele
    Sacco, Anthony
    Spiteri, Kyle
    Sammut, Charles V.
    SENSORS, 2023, 23 (11)
  • [4] A Compact Variable-Temperature Broadband Series-Resistor Calibration
    Orloff, Nathan D.
    Mateu, Jordi
    Lewandowski, Arkadiusz
    Rocas, Eduard
    King, Josh
    Gu, Dazhen
    Lu, Xiaoli
    Collado, Carlos
    Takeuchi, Ichiro
    Booth, James C.
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2011, 59 (01) : 188 - 195
  • [5] Energetic materials:: variable-temperature crystal structure of β-NTO
    Bolotina, NB
    Zhurova, EA
    Pinkerton, AA
    JOURNAL OF APPLIED CRYSTALLOGRAPHY, 2003, 36 : 280 - 285
  • [6] A test system for complex permittivity measurements of low-loss materials at high temperatures up to 2000°C
    Guo, Gaofeng
    Li, En
    Li, Zhongping
    Zhang, Qishao
    He, Fengmei
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2011, 22 (04)
  • [7] Analysis of broadband microwave conductivity and permittivity measurements of semiconducting materials
    Ritz, Elvira
    Dressel, Martin
    JOURNAL OF APPLIED PHYSICS, 2008, 103 (08)
  • [8] A VARIABLE-TEMPERATURE SAMPLE CONTAINER FOR LOW-TEMPERATURE NEUTRON POWDER DIFFRACTION
    ARZI, E
    SANDOR, E
    JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1983, 16 (OCT) : 449 - 452
  • [9] Resolving ambiguity in broadband waveguide permittivity measurements on moist materials
    Ball, JAR
    Horsfield, B
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1998, 47 (02) : 390 - 392
  • [10] Broadband coaxial cavity resonator for complex permittivity measurements of liquids
    Raveendranath, U
    Bijukumar, S
    Mathew, KT
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2000, 49 (06) : 1305 - 1312