A computational approach to fretting wear prediction at the head-stem taper junction of total hip replacements

被引:49
作者
English, Russell [1 ]
Ashkanfar, Ariyan [1 ]
Rothwell, Glynn [1 ]
机构
[1] Liverpool John Moores Univ, Sch Engn Technol & Maritime Operat, Liverpool L3 3AF, Merseyside, England
关键词
Wear modelling; Finite element modelling; Fretting; Hip joint prosthesis; Total hip replacements; Taper junction; POLYETHYLENE WEAR; METAL; SIMULATION; FAILURE; RETRIEVAL; CORROSION; TRUNNION; SURFACE; STRESS; ALLOY;
D O I
10.1016/j.wear.2015.06.016
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Wear is one of the main reasons for failure of modular total hip replacements. Recent evidence suggests that fretting wear occurs at the taper junction which provides fixation between the prosthesis femoral head and stem components. The fine metallic wear debris that is released can lead to adverse soft-tissue reactions which can necessitate a revision surgery. The present study proposes a computational methodology utilising an energy wear law and a 3D finite element model to predict fretting wear at the taper junction. The method is novel in that it simulates the weakening of the initial taper 'fixation' (created at impaction of the head onto the stem in surgery) due to the wearing process. The taper fixation is modelled using a contact analysis with overlapped meshes at the taper junction. The reduction in fixation is modelled by progressive removal of the overlap between components based on calculated wear. The fretting wear analysis approach has been shown to model the evolution of wear effectively; however, it has been shown that accurate, quantitative values for wear are critically dependant on mesh refinement, wear scaling factor and fraction, wear coefficient used and knowledge of the device loading history. The method has been implemented with a 3D finite element model of the taper junction of a commercial total hip replacement. This has been used to determine taper wear patterns, wear damage and wear rates which have been shown to be consistent with those found from observation and measurement of retrieved prostheses. The numerical method could be used to consider the effect of design changes and clinical technique on subsequent fretting wear in modular prosthetic devices. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:210 / 220
页数:11
相关论文
共 35 条
[1]   HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[2]   Wear patterns of taper connections in retrieved large diameter metal-on-metal bearings [J].
Bishop, Nicholas ;
Witt, Florian ;
Pourzal, Robin ;
Fischer, Alfons ;
Ruetschi, Marcel ;
Michel, Markus ;
Morlock, Michael .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (07) :1116-1122
[3]   High failure rates with a large-diameter hybrid metal-on-metal total hip replacement CLINICAL, RADIOLOGICAL AND RETRIEVAL ANALYSIS [J].
Bolland, B. J. R. F. ;
Culliford, D. J. ;
Langton, D. J. ;
Millington, J. P. S. ;
Arden, N. K. ;
Latham, J. M. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05) :608-615
[4]   Determining material loss from the femoral stem trunnion in hip arthroplasty using a coordinate measuring machine [J].
Bone, Martin C. ;
Sidaginamale, Raghavendra P. ;
Lord, James K. ;
Scholes, Susan C. ;
Joyce, Thomas J. ;
Nargol, Anthony V. F. ;
Langton, David J. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2015, 229 (01) :69-76
[5]  
Cohen D., 2012, BMJ BR MED J, V344
[6]   Environmental effect on fretting of metallic materials for orthopaedic implants [J].
Duisabeau, L ;
Combrade, P ;
Forest, B .
WEAR, 2004, 256 (7-8) :805-816
[7]   Stability and Trunnion Wear Potential in Large-diameter Metal-on-Metal Total Hips: A Finite Element Analysis [J].
Elkins, Jacob M. ;
Callaghan, John J. ;
Brown, Thomas D. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2014, 472 (02) :529-542
[8]   Hard-on-Hard Total Hip Impingement Causes Extreme Contact Stress Concentrations [J].
Elkins, Jacob M. ;
O'Brien, Megan K. ;
Stroud, Nicholas J. ;
Pedersen, Douglas R. ;
Callaghan, John J. ;
Brown, Thomas D. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (02) :454-463
[9]  
Fessler H, 1989, Proc Inst Mech Eng H, V203, P1, DOI 10.1243/PIME_PROC_1989_203_003_01
[10]   Computational hip joint simulator for wear and heat generation [J].
Fialho, Jorge C. ;
Fernandes, Paulo R. ;
Eca, Luis ;
Folgado, Joao .
JOURNAL OF BIOMECHANICS, 2007, 40 (11) :2358-2366