Phase coherent averaging in magnetic resonance Spectroscopy using interleaved navigator scans: Compensation of motion artifacts and magnetic field instabilities

被引:61
|
作者
Thiel, T [1 ]
Czisch, M [1 ]
Elbel, GK [1 ]
Hennig, F [1 ]
机构
[1] Univ Hosp, Dept Diagnost Radiol, Sect Med Phys, Freiburg, Germany
关键词
1H MRS; phase coherent averaging; motion correction; navigator acquisition;
D O I
10.1002/mrm.10174
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The quality of spectra in H-1 magnetic resonance spectroscopy (MRS) is strongly affected by temporal signal instabilities during the acquisition. One reason for these instabilities are hardware imperfections, e.g., drifts of the main magnetic field in superconducting magnets. This is of special concern in high-field systems where the specification of the field stability is close to the spectral linewidth. A second major potential source of artifacts, particularly in clinical MRS, is patient motion. Using standard acquisition schemes of phase-cycled averaging of the individual acquisitions, long-term effects (field drifts) as well as changes on a shorter time scale (motion) can severely reduce spectral quality. The new technique for volume-selective MRS presented here is based on the additional interleaved acquisition of a navigator signal during the recovery time of the metabolite acquisition. It corrects for temporal signal instabilities by means of a deconvolution of the metabolite and the navigator signal. This leads to phase-corrected individual metabolite scans and upon summation to a phase-coherent averaging scheme. The interleaved navigator acquisition does not require any user interaction or supervision, while sequence efficiency is maintained.
引用
收藏
页码:1077 / 1082
页数:6
相关论文
共 50 条
  • [1] Theory of coherent averaging in magnetic resonance using effective Hamiltonians
    Garg, Rajat
    Ramachandran, Ramesh
    JOURNAL OF CHEMICAL PHYSICS, 2020, 153 (03):
  • [2] Reduction of rigid-motion artifacts in interleaved projection reconstruction magnetic resonance imaging
    Van de Walle, R
    Lemahieu, I
    Raskin, G
    Lateur, G
    PROCEEDINGS OF THE 20TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 20, PTS 1-6: BIOMEDICAL ENGINEERING TOWARDS THE YEAR 2000 AND BEYOND, 1998, 20 : 648 - 651
  • [3] A Novel Phase and Frequency Navigator for Proton Magnetic Resonance Spectroscopy Using Water-Suppression Cycling
    Ernst, Thomas
    Li, Jikun
    MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (01) : 13 - 17
  • [4] Reduction of motion artifacts in magnetic resonance imaging of the neck and cervical spine by long-term averaging
    Seitz, J
    Strotzer, M
    Völk, M
    Held, P
    Djavidani, B
    Nitz, WR
    Feuerbach, S
    INVESTIGATIVE RADIOLOGY, 2000, 35 (06) : 380 - 384
  • [5] Automatic correction of motion artifacts in magnetic resonance images using an entropy focus criterion
    Atkinson, D
    Hill, DLG
    Stoyle, PNR
    Summers, PE
    Keevil, SF
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (06) : 903 - 910
  • [6] Improving spectral quality in fetal brain magnetic resonance spectroscopy using constructive averaging
    Shetty, Anil N.
    Gabr, Refaat E.
    Rendon, David A.
    Cassady, Christopher I.
    Mehollin-Ray, Amy R.
    Lee, Wesley
    PRENATAL DIAGNOSIS, 2015, 35 (13) : 1294 - 1300
  • [7] RETROSPECTIVE CORRECTION OF MOTION INDUCED ARTIFACTS IN 1H MAGNETIC RESONANCE SPECTROSCOPY OF THE FETAL BRAIN
    Evangelou, Iordanis E.
    Noeske, Ralph
    Limperopoulos, Catherine
    2015 IEEE 12TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI), 2015, : 853 - 857
  • [8] Concept of Phase Cycling in Pulsed Magnetic Resonance Using Sinusoidal Magnetic Field Modulation
    Tseytlin, Mark
    ZEITSCHRIFT FUR PHYSIKALISCHE CHEMIE-INTERNATIONAL JOURNAL OF RESEARCH IN PHYSICAL CHEMISTRY & CHEMICAL PHYSICS, 2017, 231 (03): : 689 - 703
  • [9] Artifacts reduction in strain maps of tagged magnetic resonance imaging using harmonic phase
    Wang, Daolei
    Fu, YaBo
    Ashraf, Muhammad Aqeel
    OPEN MEDICINE, 2015, 10 (01): : 425 - 433
  • [10] Using nuclear magnetic resonance spectroscopy for measuring ternary phase diagrams
    Woodworth, Jennifer K.
    Terrance, Jacob C.
    Hoffmann, Markus M.
    JOURNAL OF CHEMICAL EDUCATION, 2006, 83 (07) : 1065 - 1066