Fully discharging phases - A new approach to biphasic waveforms for external defibrillation

被引:8
作者
Yamanouchi, Y
Brewer, JE
Olson, KF
Mowrey, KA
Mazgalev, TN
Wilkoff, BL
Tchou, PJ
机构
[1] Cleveland Clin Fdn, Dept Cardiol F15, Clin Cardiac Electrophysiol, Cleveland, OH 44195 USA
[2] SurVivaLink Corp, Minneapolis, MN USA
关键词
defibrillation; death; sudden; ventricles;
D O I
10.1161/01.CIR.100.8.826
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Phase-2 voltage and maximum pulse width are dependent on phase-1 pulse characteristics in a single-capacitor biphasic waveform. The use of 2 separate output capacitors avoids these limitations and may allow waveforms with lower defibrillation thresholds. A previous report also suggested that the optimal tilt may be >70%. This study was designed to determine an optimal biphasic waveform by use of a combination of 2 separate and fully (95% tilt) discharging capacitors. Methods and Results-We performed 2 external defibrillation studies in a pig ventricular fibrillation model. In group I, 9 waveforms from a combination of 3 phase-1 capacitor values (30, 60, and 120 mu F) and 3 phase-2 capacitor values (0=monophasic, 1/3, and 1.0 times the phase-1 capacitor) were tested. Biphasic waveforms with phase-2 capacitors of 1/3 times that of phase 1 provided the highest defibrillation efficacy (stored energy and voltage) compared with corresponding monophasic and biphasic waveforms with the same capacitors in both phases except for waveforms with. a 30-mu F phase-1 capacitor. In group 2, 10 biphasic waveforms from a combination of 2 phase-1 capacitor values (30 and 60 mu F) and 5 phase-2 capacitor values (10, 20, 30, 40, and 50 mu F) were tested. In this range, phase-2 capacitor size was more critical for the 30-mu F phase-1 than for the 60-mu F phase-1 capacitor. The optimal combinations of fully discharging capacitors for defibrillation were 60/20 and 60/30 mu F Conclusions-Phase-2 capacitor size plays an important role in reducing defibrillation energy in biphasic waveforms when 2 separate and fully discharging capacitors are used.
引用
收藏
页码:826 / 831
页数:6
相关论文
共 25 条
  • [21] Clinical evaluation of defibrillation efficacy with a new single-capacitor biphasic waveform in patients undergoing implantation of an implantable cardioverter defibrillator
    Brugada, J
    Herse, B
    Sandsted, B
    Michel, U
    Schubert, BD
    Hahn, SJ
    EUROPACE, 2001, 3 (04): : 278 - 284
  • [22] A NEW APPROACH TOWARDS DEFIBRILLATION ELECTRODES - HIGHLY CONDUCTIVE ISOTROPIC CARBON-FIBERS
    ALT, E
    THERES, H
    HEINZ, M
    ALBRECHT, K
    GEORG, H
    BLOEMER, H
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11): : 1923 - 1928
  • [23] ILCOR’s revised Covid-19 defibrillation recommendation requires a new approach to training
    John A. Stewart
    Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine, 28
  • [24] ILCOR's revised Covid-19 defibrillation recommendation requires a new approach to training
    Stewart, John A.
    SCANDINAVIAN JOURNAL OF TRAUMA RESUSCITATION & EMERGENCY MEDICINE, 2020, 28 (01)
  • [25] The fully automatic external cardioverter defibrillator:: Reality of a new meaningful scenario for in-hospital cardiac arrests
    Martínez-Rubio, A
    Gusi, G
    Guillaumet, E
    Cazorla, M
    Galán, S
    Bagà, R
    Guilera, E
    Bonastre, M
    Gumà, JR
    Anguera, I
    Ibars, S
    Ochagavia, A
    Mestre, J
    Font, J
    Saura, P
    Dalmases, M
    Blanch, L
    Artigas, A
    EXPERT REVIEW OF MEDICAL DEVICES, 2005, 2 (01) : 33 - 39