Establishing a standard calibration methodology for MOSFET detectors in computed tomography dosimetry

被引:18
作者
Brady, S. L. [1 ]
Kaufman, R. A. [1 ]
机构
[1] St Jude Childrens Res Hosp, Dept Radiol Sci, Memphis, TN 38105 USA
关键词
CT; MOSFET; radiation; EFFECT TRANSISTOR TECHNOLOGY; CORONARY CT ANGIOGRAPHY; PHYSICAL PHANTOM; SENSITIVITY; ENERGY; KV; PROTOCOL; BREAST; ORGANS; RANGE;
D O I
10.1118/1.4712221
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The use of metal-oxide-semiconductor field-effect transistor (MOSFET) detectors for patient dosimetry has increased by -25% since 2005. Despite this increase, no standard calibration methodology has been identified nor calibration uncertainty quantified for the use of MOSFET dosimetry in CT. This work compares three MOSFET calibration methodologies proposed in the literature, and additionally investigates questions relating to optimal time for signal equilibration and exposure levels for maximum calibration precision. Methods: The calibration methodologies tested were (1) free in-air (FIA) with radiographic x-ray tube, (2) FIA with stationary CT x-ray tube, and (3) within scatter phantom with rotational CT x-ray tube. Each calibration was performed at absorbed dose levels of 10, 23, and 35 mGy. Times of 0 min or 5 min were investigated for signal equilibration before or after signal read out. Results: Calibration precision was measured to be better than 5%-7%, 3%-5%, and 2%-4% for the 10, 23, and 35 mGy respective dose levels, and independent of calibration methodology. No correlation was demonstrated for precision and signal equilibration time when allowing 5 min before or after signal read out. Differences in average calibration coefficients were demonstrated between the FIA with CT calibration methodology 26.761.1 mV cGy -1 versus the CT scatter phantom 29.261.0 mV cGy -1 and FIA with x-ray 29.961.1 mV cGy -1 methodologies. A decrease in MOSFET sensitivity was seen at an average change in read out voltage of -3000 mV. Conclusions: The best measured calibration precision was obtained by exposing the MOSFET detectors to 23 mGy. No signal equilibration time is necessary to improve calibration precision. A significant difference between calibration outcomes was demonstrated for FIA with CT compared to the other two methodologies. If the FIA with a CT calibration methodology was used to create calibration coefficients for the eventual use for phantom dosimetry, a measurement error -12% will be reflected in the dosimetry results. The calibration process must emulate the eventual CT dosimetry process by matching or excluding scatter when calibrating the MOSFETs. Finally, the authors recommend that the MOSFETs are energy calibrated approximately every 2500-3000 mV. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4712221]
引用
收藏
页码:3031 / 3040
页数:10
相关论文
共 26 条
[1]   Direct Quantification of Breast Dose During Coronary CT Angiography and Evaluation of Dose Reduction Strategies [J].
Abadi, Sobhi ;
Mehrez, Hatem ;
Ursani, Ali ;
Parker, Mark ;
Paul, Narinder .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (02) :W152-W158
[2]   Characterizing Metal Oxide Semiconductor Field Effect Transistor MOSFET) Radiation Detectors In a Scatter Medium Using CT Radiation Beam Delivery System [J].
Bercha, Ishtiaq H. ;
Maghsoodpour, A. ;
Keyes, G. ;
Kaufman, R. .
MEDICAL PHYSICS, 2009, 36 (06)
[3]   Implementation of radiochromic film dosimetry protocol for volumetric dose assessments to various organs during diagnostic CT procedures [J].
Brady, Samuel ;
Yoshizumi, Terry ;
Toncheva, Greta ;
Frush, Donald .
MEDICAL PHYSICS, 2010, 37 (09) :4782-4792
[4]   Characterization of high-sensitivity metal oxide semiconductor field effect transistor dosimeters system and LiF:Mg,Cu,P thermoluminescence dosimeters for use in diagnostic radiology [J].
Dong, SL ;
Chu, TC ;
Lan, GY ;
Wu, TH ;
Lin, YC ;
Lee, JS .
APPLIED RADIATION AND ISOTOPES, 2002, 57 (06) :883-891
[5]   Application of commercial MOSFET detectors for in vivo dosimetry in the therapeutic x-ray range from 80 kV to 250 W [J].
Ehringfeld, C ;
Schmid, S ;
Poljanc, K ;
Kirisits, C ;
Aiginger, H ;
Georg, D .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (02) :289-303
[6]   Radiation dose to the female breast from 16-MDCT body protocols [J].
Hurwitz, LM ;
Yoshizumi, T ;
Reiman, RE ;
Paulson, EK ;
Frush, DP ;
Nguyen, GT ;
Toncheva, GI ;
Goodman, PC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (06) :1718-1722
[7]   Radiation dose from contemporary cardiothoracic multidetector CT Protocols with an anthropomorphic female phantom: Implications for cancer induction [J].
Hurwitz, Lynne M. ;
Reiman, Robert E. ;
Yoshizumi, Terry T. ;
Goodman, Philip C. ;
Toncheva, Greta ;
Nguyen, Giao ;
Lowry, Carolyn .
RADIOLOGY, 2007, 245 (03) :742-750
[8]   Effective dose determination using an anthropomorphic phantom and metal oxide semiconductor field effect transistor technology for clinical adult body multidetector array computed tomography protocols [J].
Hurwitz, Lynne M. ;
Yoshizumi, Terry T. ;
Goodman, Philip C. ;
Frush, Donald P. ;
Nguyen, Giao ;
Toncheva, Greta ;
Lowry, Carolyn .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2007, 31 (04) :544-549
[9]   Free in air characterization of metal oxide semiconductor field effect transistor (MOSFET) dosimeters using computed tomography radiation beam delivery system [J].
Hussain, I. ;
Anderson, S. ;
Yee, B. ;
Kaufman, R. .
MEDICAL PHYSICS, 2007, 34 (06) :2348-2348
[10]   THE BUILDUP OF BACKSCATTER IN THE ENERGY-RANGE 1 MM AL TO 8 MM AL HVT [J].
KLEVENHAGEN, SC .
PHYSICS IN MEDICINE AND BIOLOGY, 1982, 27 (08) :1035-1043