Using the slope-only technique and estimated glomerular filtration rate for checking the reliability of slope-intercept measurement of glomerular filtration rate

被引:8
|
作者
Bird, Nicholas J. [1 ]
Peters, Christina [1 ]
Michell, A. Robert [2 ]
Peters, A. Michael [1 ]
机构
[1] Addenbrookes Hosp, Dept Nucl Med, Cambridge CB2 2QQ, England
[2] St Bartholomews Med Sch, William Harvey Res Inst, Dept Biochem Pharmacol, London, England
关键词
Cr-51-ethylenediaminetetraacetic acid; glomerular filtration rate; iohexol; single sample; validation;
D O I
10.1097/MNM.0b013e328310b378
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aim To develop a strategy for checking the reliability of slope intercept measurement of glomerular filtration rate (GFR). Methods Six blood samples, obtained bilaterally 20-240 min after the injection of Cr-51-ethylenediaminetetraacetic acid (EDTA) and iohexol into opposite arms, were assayed for indicator injected contralaterally. GFR6(iohexol), the reference value, was measured for all six samples and GFR3(EDTA) (slope-intercept) for the last three samples (per 1.73 m(2)). GFR was measured from the half-time [GFR/extracellular fluid volume (ECV)3(EDTA)], for the single samples at 2-4h (Christensen-Groth; GFR1) and for creatinine [estimated GFR (eGFR)]. Results In six of the 97 studies, the correlation coefficient W of the fit to the last three sample points was less than 0.99. In the remaining 91 studies, GFR3(EDTA) disagreed with GFR/ECVEDTA by more than 15% in 19 studies (group A) and by less than 15% in 72 studies (group 13). GFR3(EDTA) disagreed with GFR6(iohexol) by 12.6% in group A but only by 6.3% in group B (P < 0.001). No such discrimination was displayed by eGFR. Although GFR3(EDTA) was within 15% of GFR6(iohexol) in 14 group A studies, eGFR was within 15% of GFR3(EDTA) in only eight studies and disagreed by more than 15% in six studies. Conversely, in the five of the 19 studies in which GFR3(EDTA) disagreed with GFR6(iohexol) by more than 15%, the agreement between eGFR and GFR3(EDTA) Was less than 15% in two studies and more than 15% in three studies. GFR3(EDTA) was within 15% of GFR/ECV3(EDTA) in all six studies in which r was less than 0.99, and GFR3(EDTA) disagreed with GFR6(iohexol) by less than 15%. Confidence in GFR3(EDTA) was not improved by GFR1 or eGFR. Conclusion Slope intercept GFR is reliable if within 15% of GFR/ECV3. Estimated GFR was ineffective as a second checkpoint. GFR1 did not help when the fit was poor. Nucl Med Commun 29:1086-1092 (c) 2008 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:1086 / 1092
页数:7
相关论文
共 50 条
  • [31] Glomerular filtration rate in cows estimated by a prediction formula
    Murayama, Isao
    Miyano, Anna
    Sato, Tsubasa
    Iwama, Ryosuke
    Satoh, Hiroshi
    Ichijyo, Toshihiro
    Sato, Shigeru
    Furuhama, Kazuhisa
    ANIMAL SCIENCE JOURNAL, 2014, 85 (12) : 1001 - 1004
  • [32] Estimated glomerular filtration rate in people living with HIV
    Pontes, Priscila Silva
    Melo, Elizabete Santos
    Braz Costa, Christefany Regia
    Antonini, Marcela
    Milanes Sousa, Laelson Rochelle
    Gir, Elucir
    Reis, Renata Karina
    ACTA PAULISTA DE ENFERMAGEM, 2019, 32 (05) : 493 - 499
  • [33] Bias and precision of estimated glomerular filtration rate in children
    Gretz, Norbert
    Schock, Daniel
    Sadick, Maliha
    Pill, Johannes
    PEDIATRIC NEPHROLOGY, 2007, 22 (02) : 167 - 169
  • [34] Glomerular filtration rate measurement and prediction equations
    Soares, Ariana Aguiar
    Eyff, Tatiana Falcao
    Campani, Raquel Barth
    Ritter, Luciana
    Camargo, Joiza Lins
    Silveiro, Sandra Pinho
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2009, 47 (09) : 1023 - 1032
  • [35] Glomerular Filtration Rate Measurement and Chemotherapy Dosing
    Rosner, Mitchell H.
    Sprangers, Ben
    Sandhu, Geeta
    Malyszko, Jolanta
    SEMINARS IN NEPHROLOGY, 2022, 42 (06)
  • [36] Measured glomerular filtration rate predicts liver related deaths better than estimated glomerular filtration rate in advanced chronic liver disease
    Gonzalez-Alayon, Carlos
    Hernandez-Guerra, M.
    Luis-Lima, Sergio
    Lima, Coriolano Cruz Perera
    Santana-Delgado, Andrea
    Diaz-Mesa, Carlos
    Morant-Dominguez, Andrea
    Martin, Laura Diaz
    Gonzalez-Rinne, Federico
    Hernandez-Bustabad, Alberto
    Moreno, Miguel
    Gaspari, Flavio
    Porrini, Esteban
    DIGESTIVE AND LIVER DISEASE, 2025, 51 (02) : 477 - 484
  • [37] Transdermal Measurement of Glomerular Filtration Rate in Mice
    Scarfe, Lauren
    Schock-Kusch, Daniel
    Ressel, Lorenzo
    Friedemann, Jochen
    Shulhevich, Yury
    Murray, Patricia
    Wilm, Bettina
    de Caestecker, Mark
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2018, (140):
  • [38] Clarifying the Fog of Natural and Manmade Renal Function Tests Creatinine, Clearances, Glomerular Filtration Rate, and Estimated Glomerular Filtration Rate
    Toffaletti, John G.
    POINT OF CARE, 2011, 10 (01): : 45 - 50
  • [39] Comparison of Measured Glomerular Filtration Rate Versus Estimated Glomerular Filtration Rate in Indian Cirrhotic Patients: Report of a Pilot Study
    Divyaveer, Smita
    Kashyap, Madhuri
    Kekan, Kushal
    Yadav, Ashok K.
    Kaur, Jaskiran
    Premkumar, Madhumita
    Gandotra, Akash
    Prajapati, Kanchan
    De, Arka
    Duseja, Ajay K.
    Verma, Nipun
    Aggarwal, Radhika
    Venkatasubramanian, Vaishnavi
    Tiwari, Vaibhav
    Bagur, Vishnuvardhan
    Patil, Amol N.
    Safiq, Nusrat
    Kohli, Harbir S.
    JOURNAL OF CLINICAL AND EXPERIMENTAL HEPATOLOGY, 2025, 15 (02)
  • [40] Suitability of a simplified technique based on iohexol for decentralized measurement of glomerular filtration rate
    Bird, Nicholas J.
    Peters, Christina
    Michell, A. Robert
    Peters, A. Michael
    SCANDINAVIAN JOURNAL OF UROLOGY AND NEPHROLOGY, 2008, 42 (05): : 472 - 480