Use of the photoelastic method and finite element analysis in the assessment of wall strain in abdominal aortic aneurysm models

被引:26
作者
Doyle, Barry J. [1 ,2 ,3 ]
Killion, John [1 ,2 ,4 ]
Callanan, Anthony [1 ,2 ]
机构
[1] Univ Limerick, Ctr Appl Biomed Engn Res CABER, Dept Mech Aeronaut & Biomed Engn, Limerick, Ireland
[2] Univ Limerick, Mat & Surface Sci Inst, Limerick, Ireland
[3] Univ Edinburgh, Ctr Cardiovasc Sci, Edinburgh, Midlothian, Scotland
[4] Athlone Inst Technol, Mat Res Inst, Athlone, Ireland
关键词
Abdominal aortic aneurysm; Photoelastic method; FEA; Biomechanics; Strain; Validation; RUPTURE LOCATIONS; STRESS; IDENTIFICATION; DIAMETER; RISK;
D O I
10.1016/j.jbiomech.2012.05.004
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Abdominal aortic aneurysm (AAA) is a significant health problem. Current clinical rupture-risk relies primarily on the maximum diameter of the AAA and also growth rate. However, AAAs are a patient-specific problem and recently, numerical tools have been employed to estimate rupture-potential. Alternatively, experimental assessment of AAA biomechanics receives less attention, yet, rigorous validation of numerical tools is required prior to clinical acceptance. This paper examines the use of the photoelastic method to assess wall strain and its validation using finite element analysis (FEA) in a small number of patient-specific AAA models. Experimental models were manufactured in-house using the injection-moulding procedure together with a commercially available photoelastic material. The material was mechanically characterised prior to testing, with models examined under three loading regimes (80, 120 and 160 mmHg). Each experimental model was imaged using computed tomography (CT) and reconstructed in three dimensions (3D) for numerical analyses. Experimental wall strain was measured and numerical wall strain calculated with finite element analysis (FEA). Results were qualitatively and quantitatively compared. There was good qualitative agreement between the experimental and numerical methods, with similar trends apparent throughout all models at all pressures. Overall, acceptable percentage errors between the techniques were observed for all models. Median errors of -6.5%, -0.4% and 3.9% for the models at 80, 120 and 160 mmHg pressures, respectively, were determined. The photoelastic method is a very useful experimental tool that provides instant, easy to interpret, information regarding wall strain. The technique is useful for validation of numerical AAA studies. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1759 / 1768
页数:10
相关论文
共 34 条
[1]   Strain estimation in abdominal aortic aneurysms from 2-D ultrasound [J].
Brekken, R ;
Bang, J ;
Odegard, A ;
Aasland, J ;
Hernes, TAN ;
Myhre, HO .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2006, 32 (01) :33-42
[2]  
Callanan A., 2011, COMPUTER METHODS BIO, V1, P1
[3]   Engineering Silicone Rubbers for In Vitro Studies: Creating AAA Models and ILT Analogues With Physiological Properties [J].
Corbett, T. J. ;
Doyle, B. J. ;
Callanan, A. ;
Walsh, M. T. ;
McGloughlin, T. M. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2010, 132 (01)
[4]  
da Silva ES, 2000, CARDIOVASC SURG, V8, P526
[5]  
DARLING RC, 1977, CIRCULATION, V56, P161
[6]   3D reconstruction and manufacture of real abdominal aortic aneurysms: From CT scan to silicone model [J].
Doyle, B. J. ;
Morris, L. G. ;
Callanan, A. ;
Kelly, P. ;
Vorp, D. A. ;
McGloughlin, T. M. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2008, 130 (03)
[7]  
Doyle B.J., 2011, Biomechanics and Mechanobiology of Aneurysms, P119, DOI 10.1007/8415_2011_70
[8]   Identification of rupture locations in patient-specific abdominal aortic aneurysms using experimental and computational techniques [J].
Doyle, Barry J. ;
Cloonan, Aidan J. ;
Walsh, Michael T. ;
Vorp, David A. ;
McGloughlin, Timothy M. .
JOURNAL OF BIOMECHANICS, 2010, 43 (07) :1408-1416
[9]   Experimental modelling of aortic aneurysms: Novel applications of silicone rubbers [J].
Doyle, Barry J. ;
Corbett, Timothy J. ;
Cloonan, Aidan J. ;
O'Donnell, Michael R. ;
Walsh, Michael T. ;
Vorp, David A. ;
McGloughlin, Timothy M. .
MEDICAL ENGINEERING & PHYSICS, 2009, 31 (08) :1002-1012
[10]   An Experimental and Numerical Comparison of the Rupture Locations of an Abdominal Aortic Aneurysm [J].
Doyle, Barry J. ;
Corbett, Timothy J. ;
Callanan, Anthony ;
Walsh, Michael T. ;
Vorp, David A. ;
McGloughlin, Timothy M. .
JOURNAL OF ENDOVASCULAR THERAPY, 2009, 16 (03) :322-335