Calorimetric determination of the absorbed dose to water for medium-energy x-rays with generating voltages from 70 to 280 kV

被引:34
作者
Krauss, A. [1 ]
Bueermann, L. [1 ]
Kramer, H-M [1 ]
Selbach, H-J [1 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
关键词
RADIOTHERAPY; DOSIMETRY;
D O I
10.1088/0031-9155/57/19/6245
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For medium energy x-rays produced with tube voltages from 70 to 280 kV, the absorbed dose to water, D-w, has been determined by means of water calorimetry with relative standard uncertainties ranging from 0.45% to 0.98% at 280 and 70 kV. The results were confirmed by Monte Carlo calculations, in which the ratios of D-w at 5 cm depth in a reference water phantom to the air kerma free in air, K-a, at the same point in space were compared to the corresponding ratios determined experimentally. The general agreement between measurement and calculation was better than 1%. These results confirm earlier investigations in which the absorbed dose to graphite was determined by means of a graphite extrapolation chamber. For the Monte Carlo calculations, an attempt was made to present a complete uncertainty budget, taking into account type B contributions also.
引用
收藏
页码:6245 / 6268
页数:24
相关论文
共 46 条
  • [1] Andreo P., 2000, IAEA TRS-398
  • [2] [Anonymous], DOS16 PTB
  • [3] [Anonymous], J RES NATL I STAND T
  • [4] [Anonymous], 2006, METROLOGIA, DOI [10.1088/0026-1394/43/1A/06005, DOI 10.1088/0026-1394/43/1A/06005]
  • [5] [Anonymous], P NRC WORKSH WAT CAL
  • [6] [Anonymous], 1992, 4812 NISTIR
  • [7] [Anonymous], 2006, PIRS701 NRCC
  • [8] [Anonymous], DOS35 PTB
  • [9] [Anonymous], PHYS MED BIOL UNPUB
  • [10] [Anonymous], P NRC WORKSH WAT CAL