Development of a Respiratory Inductive Plethysmography Module Supporting Multiple Sensors for Wearable Systems

被引:61
作者
Zhang, Zhengbo [1 ]
Zheng, Jiewen [2 ]
Wu, Hao [1 ]
Wang, Weidong [1 ]
Wang, Buqing [1 ]
Liu, Hongyun [1 ]
机构
[1] Chinese PLA Peoples Liberat Army Gen Hosp, Dept Biomed Engn, Beijing 100853, Peoples R China
[2] Quartermaster Res Inst, Gen Logist Dept, Beijing 100010, Peoples R China
基金
北京市自然科学基金;
关键词
respiratory inductive plethysmography; pulse amplitude modulation; time-division multiplexing scheme; LC oscillator; sensing unit; multiple sensors; wearable system; CARDIORESPIRATORY MEASUREMENTS; ACCURACY; VENTILATION; EXERCISE; CALIBRATION;
D O I
10.3390/s121013167
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this paper, we present an RIP module with the features of supporting multiple inductive sensors, no variable frequency LC oscillator, low power consumption, and automatic gain adjustment for each channel. Based on the method of inductance measurement without using a variable frequency LC oscillator, we further integrate pulse amplitude modulation and time division multiplexing scheme into a module to support multiple RIP sensors. All inductive sensors are excited by a high-frequency electric current periodically and momentarily, and the inductance of each sensor is measured during the time when the electric current is fed to it. To improve the amplitude response of the RIP sensors, we optimize the sensing unit with a matching capacitor parallel with each RIP sensor forming a frequency selection filter. Performance tests on the linearity of the output with cross-sectional area and the accuracy of respiratory volume estimation demonstrate good linearity and accurate lung volume estimation. Power consumption of this new RIP module with two sensors is very low. The performance of respiration measurement during movement is also evaluated. This RIP module is especially desirable for wearable systems with multiple RIP sensors for long-term respiration monitoring.
引用
收藏
页码:13167 / 13184
页数:18
相关论文
共 28 条
[1]   Cardiorespiratory Measurements During Field Tests in CF: Use of an Ambulatory Monitoring System [J].
Bradley, Judy M. ;
Kent, Lisa ;
O'Neill, Brenda ;
Nevill, Alan ;
Boyle, Lesley ;
Elborn, J. Stuart .
PEDIATRIC PULMONOLOGY, 2011, 46 (03) :253-260
[2]   Respiratory Monitoring by Inductive Plethysmography in Unrestrained Subjects Using Position Sensor-Adjusted Calibration [J].
Bruellmann, Gregor ;
Fritsch, Karsten ;
Thurnheer, Robert ;
Bloch, Konrad E. .
RESPIRATION, 2010, 79 (02) :112-120
[3]   Signal processing technique for non-invasive real-time estimation of cardiac output by inductance cardiography (thoracocardiography) [J].
Bucklar, GB ;
Kaplan, V ;
Bloch, KE .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2003, 41 (03) :302-309
[4]   ACCURACY OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY DURING WAKEFULNESS AND SLEEP IN PATIENTS WITH OBSTRUCTIVE SLEEP-APNEA [J].
CANTINEAU, JP ;
ESCOURROU, P ;
SARTENE, R ;
GAULTIER, C ;
GOLDMAN, M .
CHEST, 1992, 102 (04) :1145-1151
[5]   Evaluation of respiratory inductive plethysmography - Accuracy for analysis of respiratory waveforms [J].
Carry, PY ;
Baconnier, P ;
Eberhard, A ;
Cotte, P ;
Benchetrit, G .
CHEST, 1997, 111 (04) :910-915
[6]  
CHADHA TS, 1982, AM REV RESPIR DIS, V125, P644
[7]   Monitoring of ventilation during exercise by a portable respiratory inductive plethysmograph [J].
Clarenbach, CF ;
Senn, O ;
Brack, T ;
Kohler, M ;
Bloch, KE .
CHEST, 2005, 128 (03) :1282-1290
[8]   DESIGN OF AN INDUCTIVE PLETHYSMOGRAPH FOR VENTILATION MEASUREMENT [J].
COHEN, KP ;
PANESCU, D ;
BOOSKE, JH ;
WEBSTER, JG ;
TOMPKINS, WJ .
PHYSIOLOGICAL MEASUREMENT, 1994, 15 (02) :217-229
[9]  
COHN MA, 1982, B EUR PHYSIOPATH RES, V18, P643
[10]   Respiratory inductive plethysmography to assess respiratory variability and complexity in humans [J].
Fiamma, Marie-Noelle ;
Samara, Ziyad ;
Baconnier, Pierre ;
Similowski, Thomas ;
Straus, Christian .
RESPIRATORY PHYSIOLOGY & NEUROBIOLOGY, 2007, 156 (02) :234-239