Comparison of Intraocular Pressure Measurements between Icare Pro Rebound Tonometer and Tono-Pen XL Tonometer in Supine and Lateral Decubitus Body Positions

被引:9
作者
Lee, Tae-Eun [1 ]
Yoo, Chungkwon [1 ]
Hwang, Jin-Young [1 ]
Lin, Shan [2 ]
Kim, Yong Yeon [1 ]
机构
[1] Korea Univ, Coll Med, Dept Ophthalmol, Seoul 136705, South Korea
[2] Univ Calif San Francisco, Dept Ophthalmol, San Francisco, CA USA
关键词
Body position; Icare tonometer; intraocular pressure; lateral decubitus; rebound tonometer; supine; tono-pen tonometer; CENTRAL CORNEAL THICKNESS; CURVATURE;
D O I
10.3109/02713683.2014.964416
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare intraocular pressure (IOP) measurements obtained using the Icare Pro rebound tonometer and Tono-Pen XL tonometer in supine and lateral decubitus body positions. Methods: One-hundred eyes of 50 subjects (normal volunteers or glaucoma suspects) were enrolled in this prospective observational study. IOP was measured in both eyes using the Icare Pro and Tono-Pen XL in the sitting position and the recumbent positions including supine, right lateral decubitus and left lateral decubitus. IOP was measured five minutes after assuming each of the recumbent postures in a randomized sequence. The eye on the lower side in the lateral decubitus position was termed as the dependent eye. Agreement of IOP readings between the Icare Pro and Tono-Pen was assessed in all recumbent positions. Differences of IOP readings (Delta IOP) between the two tonometers and their correlations with ocular parameters were also assessed in all positions. Results: The IOP readings obtained using Icare Pro and Tono-Pen showed good correlations in supine and lateral decubitus positions (all r>0.7, p<0.005), although Icare Pro readings were higher than Tono-Pen readings (all p<0.001) in all positions. The DIOP showed a weakly positive correlation with central corneal thickness in both eyes, whereas such a positive correlation was found only in the dependent eye in the lateral decubitus positions (r=0.307-0.531, all p<0.005). Both the spherical equivalents and axial lengths were not correlated with Delta IOP in all positions. Conclusion: IOP readings obtained with Tono-Pen and Icare Pro tonometers showed good agreement in supine as well as in lateral decubitus positions, although Icare readings were higher than Tono-Pen readings in all positions. Such differences in IOP readings between the different tonometers need to be considered when measuring IOPs in various body positions.
引用
收藏
页码:923 / 929
页数:7
相关论文
共 41 条
  • [31] The Impact of Corneal Edema on Intraocular Pressure Measurements Using Goldmann Applanation Tonometry, Tono-Pen XL, iCare, and ORA: An In Vitro Model
    Neuburger, Matthias
    Maier, Philip
    Boehringer, Daniel
    Reinhard, Thomas
    Jordan, Jens F.
    [J]. JOURNAL OF GLAUCOMA, 2013, 22 (07) : 584 - 590
  • [32] Comparison of Scleral Tono-Pen Intraocular Pressure Measurements with Goldmann Applanation Tonometry
    Badakere, Swathi Vallabh
    Choudhari, Nikhil S.
    Rao, Harsha L.
    Chary, C. Raghava
    Garudadri, Chandra Sekhar
    Senthil, Sirisha
    [J]. OPTOMETRY AND VISION SCIENCE, 2018, 95 (02) : 129 - 135
  • [33] Comparison of Disposable Goldmann Applanation Tonometer, ICare ic100, and Tonopen XL to Standards of Care Goldmann Nondisposable Applanation Tonometer for Measuring Intraocular Pressure
    Wong, Benny
    Parikh, Deep
    Rosen, Lisa
    Gorski, Matthew
    Angelilli, Allison
    Shih, Carolyn
    [J]. JOURNAL OF GLAUCOMA, 2018, 27 (12) : 1119 - 1124
  • [34] Effect of mild to moderate myopic correction by laser-assisted in situ keratomileusis on intraocular pressure measurements with Goldmann applanation tonometer, Tono-Pen, and pneumatonometer
    Vakili, R
    Choudhri, SA
    Tauber, S
    Shields, MB
    [J]. JOURNAL OF GLAUCOMA, 2002, 11 (06) : 493 - 496
  • [35] Comparison of intraocular pressure measurements using Goldmann tonometer, I-care pro, Tonopen XL, and Schiotz tonometer in patients after Descemet stripping endothelial keratoplasty
    Ohana, Oded
    Varssano, David
    Shemesh, Gabi
    [J]. INDIAN JOURNAL OF OPHTHALMOLOGY, 2017, 65 (07) : 579 - 583
  • [36] Comparison of simultaneous readings of intraocular pressure in rabbits using Perkins handheld, Tono-Pen XL, and TonoVet tonometers
    Kalesnykas, Giedrius
    Uusitalo, Hannu
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2007, 245 (05) : 761 - 762
  • [37] Comparison of simultaneous readings of intraocular pressure in rabbits using Perkins handheld, Tono-Pen XL, and TonoVet tonometers
    Giedrius Kalesnykas
    Hannu Uusitalo
    [J]. Graefe's Archive for Clinical and Experimental Ophthalmology, 2007, 245 : 761 - 762
  • [38] Comparisons of the Tono-Pen® and Goldmann Applanation Tonometer in the Measurement of Intraocular Pressure of Primary Open Angle Glaucoma Patients in a Hospital Population in Southwest Nigeria
    Onochie, Chizoba
    Okoye, Obiekwe
    Ogunro, Adunola
    Aribaba, Tade
    Hassan, Kunle
    Onakoya, Adeola
    [J]. MEDICAL PRINCIPLES AND PRACTICE, 2016, 25 (06) : 566 - 571
  • [39] Comparison of Intraocular Pressure Measurements Between iCARE Tonovet and Tonopen XL
    Villamarin, A.
    Roy, S.
    Stergiopulos, N.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [40] Intraocular Pressure Curves of Untreated Glaucoma Suspects and Glaucoma Patients in Sitting and Lateral Decubitus Positions Using the Goldmann Applanation Tonometer
    Piven, Ilia
    Glovinsky, Yoseph
    [J]. JOURNAL OF GLAUCOMA, 2014, 23 (08) : 541 - 546