Comparison of Intraocular Pressure Measurements between Icare Pro Rebound Tonometer and Tono-Pen XL Tonometer in Supine and Lateral Decubitus Body Positions

被引:9
|
作者
Lee, Tae-Eun [1 ]
Yoo, Chungkwon [1 ]
Hwang, Jin-Young [1 ]
Lin, Shan [2 ]
Kim, Yong Yeon [1 ]
机构
[1] Korea Univ, Coll Med, Dept Ophthalmol, Seoul 136705, South Korea
[2] Univ Calif San Francisco, Dept Ophthalmol, San Francisco, CA USA
关键词
Body position; Icare tonometer; intraocular pressure; lateral decubitus; rebound tonometer; supine; tono-pen tonometer; CENTRAL CORNEAL THICKNESS; CURVATURE;
D O I
10.3109/02713683.2014.964416
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare intraocular pressure (IOP) measurements obtained using the Icare Pro rebound tonometer and Tono-Pen XL tonometer in supine and lateral decubitus body positions. Methods: One-hundred eyes of 50 subjects (normal volunteers or glaucoma suspects) were enrolled in this prospective observational study. IOP was measured in both eyes using the Icare Pro and Tono-Pen XL in the sitting position and the recumbent positions including supine, right lateral decubitus and left lateral decubitus. IOP was measured five minutes after assuming each of the recumbent postures in a randomized sequence. The eye on the lower side in the lateral decubitus position was termed as the dependent eye. Agreement of IOP readings between the Icare Pro and Tono-Pen was assessed in all recumbent positions. Differences of IOP readings (Delta IOP) between the two tonometers and their correlations with ocular parameters were also assessed in all positions. Results: The IOP readings obtained using Icare Pro and Tono-Pen showed good correlations in supine and lateral decubitus positions (all r>0.7, p<0.005), although Icare Pro readings were higher than Tono-Pen readings (all p<0.001) in all positions. The DIOP showed a weakly positive correlation with central corneal thickness in both eyes, whereas such a positive correlation was found only in the dependent eye in the lateral decubitus positions (r=0.307-0.531, all p<0.005). Both the spherical equivalents and axial lengths were not correlated with Delta IOP in all positions. Conclusion: IOP readings obtained with Tono-Pen and Icare Pro tonometers showed good agreement in supine as well as in lateral decubitus positions, although Icare readings were higher than Tono-Pen readings in all positions. Such differences in IOP readings between the different tonometers need to be considered when measuring IOPs in various body positions.
引用
收藏
页码:923 / 929
页数:7
相关论文
共 41 条
  • [1] Comparison of intraocular pressure measurements with the tono-pen, goldmann applanation tonometer, and noncontact tonometer in nonglaucomatous pseudophakic children
    Yalcin, Sibel Oskan
    Kaplan, Aysin Tuba
    INTERNATIONAL OPHTHALMOLOGY, 2024, 44 (01)
  • [2] Comparison of intraocular pressure measurement with Schiotz tonometer and Tono-Pen Vet tonometer in healthy dogs
    Wrzesniewska, Karolina
    Madany, Jacek
    Winiarczyk, Dagmara
    JOURNAL OF VETERINARY RESEARCH, 2018, 62 (02) : 243 - 247
  • [3] Intraocular Pressure Measurements in Standing, Sitting, and Supine Position: Comparison between Tono-Pen Avia and Icare Pro Tonometers
    De Bernardo, Maddalena
    Abbinante, Giulia
    Borrelli, Maria
    Di Stasi, Margherita
    Cione, Ferdinando
    Rosa, Nicola
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (21)
  • [4] Variations of Intraocular Pressure Measured by Goldmann Applanation Tonometer, Tono-Pen, iCare Rebound Tonometer, and Pascal Dynamic Contour Tonometer in Patients With Corneal Edema After Phacoemulsification
    Kiddee, Weerawat
    Tanjana, Arunee
    JOURNAL OF GLAUCOMA, 2021, 30 (04) : 317 - 324
  • [5] Measurement of intraocular pressure in clinically normal Turkish Shepherd Dogs with the rebound tonometer (TonoVet®) and the applanation tonometer (Tono-Pen Vet®)
    Kulualp, Kadri
    Yurdakul, Ibrahim
    Erol, Hanifi
    Atalan, Gultekin
    Kilic, Servet
    MEDYCYNA WETERYNARYJNA-VETERINARY MEDICINE-SCIENCE AND PRACTICE, 2018, 74 (09): : 568 - 573
  • [6] Repeatability of intraocular pressure measurements with Icare PRO rebound, Tono-Pen AVIA, and Goldmann tonometers in sitting and reclining positions
    Caterina Schweier
    James VM Hanson
    Jens Funk
    Marc Töteberg-Harms
    BMC Ophthalmology, 13
  • [7] Comparison of Intraocular Pressure Measurements Obtained by Icare PRO Rebound Tonometer, Tomey FT-1000 Noncontact Tonometer, and Goldmann Applanation Tonometer in Healthy Subjects
    Guler, Mete
    Bilak, Semsettin
    Bilgin, Burak
    Simsek, Ali
    Capkin, Musa
    Reyhan, Ali Hakim
    JOURNAL OF GLAUCOMA, 2015, 24 (08) : 613 - 618
  • [8] Comparison of the intraocular pressure measurements with the tono-pen and the goldman applanation tonometer and the effect of central corneal thickness on measurements
    Dervisogullari, Mehmet Serdar
    Akarsu, Cengiz
    Ergin, Ahmet
    ANNALS OF CLINICAL AND ANALYTICAL MEDICINE, 2019, 10 (04): : 479 - 484
  • [9] Comparison of the Tono-Pen and Goldmann tonometer for measuring intraocular pressure in patients with glaucoma
    Horowitz, GS
    Byles, J
    Lee, JL
    D'Este, C
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2004, 32 (06) : 584 - 589
  • [10] The influence of central corneal thickness and corneal curvature on Intraocular pressure measured by Tono-Pen and rebound tonometer in children
    Sahin, Afsun
    Basmak, Hikmet
    Yidirim, Nigun
    JOURNAL OF GLAUCOMA, 2008, 17 (01) : 57 - 61