Microwave TDR for Real-Time Control of Intravenous Drip Infusions

被引:38
作者
Cataldo, Andrea [1 ]
Cannazza, Giuseppe [1 ]
Giaquinto, Nicola [2 ]
Trotta, Amerigo [2 ]
Andria, Gregorio [2 ]
机构
[1] Univ Salento, Dept Engn Innovat, I-73100 Lecce, Italy
[2] Polytech Bari, Dept Elect & Elect, I-70125 Bari, Italy
关键词
Biomedical monitoring; medical services; microwave measurements; microwave reflectometry; time-domain reflectometry (TDR); REFLECTOMETRY; PERMITTIVITIES;
D O I
10.1109/TIM.2012.2192346
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper explores the use of a microwave-reflectometry-based system for the automatic control and real-time monitoring of the flow and of the liquid level in intravenous (IV) medical infusions. In medical and hospital contexts, other kinds of devices, mainly based on the optical detection and counting of the infusion drops, are used. Nevertheless, the proposed system is aimed at circumventing some typical drawbacks deriving from the adoption of these traditional methods, thus allowing an efficient alternative for automatically monitoring the instantaneous flow of IV medical solutions. To this purpose, the proposed system combines microwave time-domain reflectometry (TDR) measurements with a noninvasive sensing element (i.e., strip electrodes directly attached to the external surface of the infusion bottle). Experimental results confirm that, by using low-cost portable TDR devices, the solution flow process can be controlled with acceptable accuracy. Therefore, the proposed method can be regarded as a promising control tool for in-hospital patient management as well as for telemedicine programs.
引用
收藏
页码:1866 / 1873
页数:8
相关论文
共 13 条
[11]  
Voss G. I., 2000, BIOMEDICAL ENG HDB
[12]   Design of Medical Infusion Monitor and Protection System Based On Wireless Communication Technology [J].
Wen, Xinling .
2008 INTERNATIONAL SYMPOSIUM ON INTELLIGENT INFORMATION TECHNOLOGY APPLICATION, VOL II, PROCEEDINGS, 2008, :755-759
[13]  
Yang YS, 1998, P ANN INT IEEE EMBS, V20, P1795, DOI 10.1109/IEMBS.1998.746937