Asymmetric quadrature split birdcage coil for hyperpolarized 3He lung MRI at 1.5T

被引:37
作者
De Zanche, Nicola [2 ,3 ]
Chhina, Navjeet
Teh, Kevin [1 ]
Randell, Chris
Pruessmann, Klaas P. [2 ,3 ]
Wild, Jim M. [1 ]
机构
[1] Univ Sheffield, Unit Acad Radiol, Sheffield S10 2JF, S Yorkshire, England
[2] Univ Zurich, Inst Biomed Engn, Zurich, Switzerland
[3] Swiss Fed Inst Technol, Zurich, Switzerland
基金
英国工程与自然科学研究理事会;
关键词
asymmetric birdcage coil; hyperpolarized gas imaging; helium-3; lung imaging; radio frequency coil;
D O I
10.1002/mrm.21664
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An asymmetric quadrature birdcage body coil for hyperpolarized (Hp) He-3 lung imaging at 1.5T is presented. The coil is designed to rest on top of the patient support and be used as a temporary insert in a clinical system. A two-part construction facilitates patient access and the asymmetric design makes maximal use of available bore space to ensure comfort. H g homogeneous, circularly polarized RF magnetic fields are produced at 48.5 MHz using a conformal mapping method for the geometrical design, combined with an algebraic method to calculate the individual capacitance values on the birdcage coil's ladder network. Efficiency and isolation from the system's proton body coil are ensured by an integrated RF screen. The design methodology is readily applicable to other field strengths or nuclei. Improvements over existing 3He coils were found in terms of sensitivity and transmit field homogeneity, an important feature in HP MRI.
引用
收藏
页码:431 / 438
页数:8
相关论文
共 24 条
  • [1] Barberi EA, 2000, MAGNET RESON MED, V43, P284, DOI 10.1002/(SICI)1522-2594(200002)43:2<284::AID-MRM16>3.0.CO
  • [2] 2-C
  • [3] Burke G.J., 1981, NUMERICAL ELECTROMAG
  • [4] Davis P.J., 1994, CIRCULANT MATRICES, Vsecond
  • [5] Analytical approach to noncircular section birdcage coil design: Verification with a Cassinian oval coil
    De Zanche, N
    Yahya, A
    Vermeulen, FE
    Allen, PS
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (01) : 201 - 211
  • [6] DEZANCHE N, 2006, P 23 ANN M EUR SOC M
  • [7] DUTOIT CF, 1995, J APPL COMPUT ELECTR, V10, P31
  • [8] DETERMINING CURRENT DISTRIBUTIONS FOR RF RESONATORS IN MAGNETIC-RESONANCE-IMAGING
    FORBES, LK
    CROZIER, S
    DODDRELL, DM
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 1995, 6 (03) : 284 - 292
  • [9] Hoult DI, 2000, CONCEPT MAGNETIC RES, V12, P173, DOI 10.1002/1099-0534(2000)12:4<173::AID-CMR1>3.0.CO
  • [10] 2-Q