Validation of the A&D UA-1020 upper-arm blood pressure monitor for home blood pressure monitoring according to the British Hypertension Society Protocol

被引:2
|
作者
Zeng, Wei-Fang [1 ]
Kang, Yuan-Yuan [1 ]
Liu, Ming [1 ]
Li, Yan [1 ]
Wang, Ji-Guang [1 ]
机构
[1] Shanghai Jiao Tong Univ, Ruijin Hosp,Sch Med, Shanghai Inst Hypertens, Ctr Epidemiol Studies & Clin Trials, Shanghai 200025, Peoples R China
关键词
blood pressure determination; blood pressure monitor; British Hypertension Society Protocol; validation studies; EUROPEAN-SOCIETY;
D O I
10.1097/MBP.0b013e32836141ca
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective The present study aimed to evaluate the accuracy of the automated oscillometric upper-arm blood pressure (BP) monitor A&D UA-1020 with two different-shaped cuffs for home BP monitoring according to the British Hypertension Society (BHS) Protocol. Methods We recruited individuals for each of the two cuffs (D-ring and cylindrical) until there were 85 eligible participants (255 pairs of comparisons) and their BP could meet the BP distribution requirements specified by the BHS Protocol. For each participant, we sequentially measured the systolic and diastolic BP using a mercury sphygmomanometer (two observers) and the UA-1020 device (one supervisor). Results For the D-ring cuff, the device achieved grade A. The percentage of BP differences within 5, 10, and 15 mmHg was 67, 87, and 96%, respectively, for systolic BP, and 70, 90, and 99%, respectively, for diastolic BP. The average (+/- SD) of the device-observer differences was -0.2 +/- 7.3 mmHg (P = 0.64) and 1.7 +/- 5.8 mmHg (P < 0.0001) for systolic and diastolic BP, respectively. For the cylindrical cuff, the device also achieved grade A. The percentage of BP differences within 5, 10, and 15 mmHg was 67, 88, and 97%, respectively, for systolic BP and 64, 89, and 98%, respectively, for diastolic BP. The average of the device-observer differences was - 0.1 +/- 7.0mmHg (P = 0.89) and 2.0 +/- 6.3mmHg (P < 0.0001) for systolic and diastolic BP, respectively. Conclusion The UA-1020 device has passed the requirements of the BHS Protocol with both the D-ring and the cylindrical cuffs, and hence can be recommended for home use in adults. (c) 2013 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:177 / 181
页数:5
相关论文
共 50 条
  • [1] Validation of the A&D UA-1020 upper-arm blood pressure monitor with six different-shaped or different-sized cuffs according to the British Hypertension Society protocol
    Zeng, Wei-Fang
    Kang, Yuan-Yuan
    Liu, Ming
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2013, 18 (05) : 272 - 277
  • [2] Validation of A&D TM-2430 upper-arm blood pressure monitor for ambulatory blood pressure monitoring in children and adolescents, according to the British Hypertension Society protocol
    Yip, Gabriel Wai-Kwok
    So, Hung-Kwan
    Li, Albert Martin
    Tomlinson, Brian
    Wong, Sik-Nin
    Sung, Rita Yn-Tz
    BLOOD PRESSURE MONITORING, 2012, 17 (02) : 76 - 79
  • [3] Validation of the SEJOY BP-1307 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Lei, Lei
    Chen, Yi
    Chen, Qi
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2017, 22 (06) : 371 - 374
  • [4] Validation of the Rossmax CF175 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Zhang, Lu
    Kang, Yuan-Yuan
    Zeng, Wei-Fang
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2015, 20 (02) : 104 - 107
  • [5] Validation of the BPUMP BF1112 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Chen, Qi
    Kang, Yuan-Yuan
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2017, 22 (02) : 105 - 108
  • [6] Validation of the AVITA BPM64 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Liu, Chang-Yuan
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2018, 23 (01) : 45 - 48
  • [7] Validation of the Andon KD-5965 upper-arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Huang, Jinhua
    Li, Zhijie
    Li, Guimei
    Liu, Zhaoying
    BLOOD PRESSURE MONITORING, 2015, 20 (05) : 283 - 285
  • [8] Validation of the Rossmax AC1000f upper-arm blood pressure monitor in adults according to the British Hypertension Society Protocol
    Huang, Jian-Feng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2020, 25 (04) : 231 - 235
  • [9] Validation of the fully automated A&D TM-2656 blood pressure monitor according to the British Hypertension Society Protocol
    Zeng, Wei-Fang
    Liu, Ming
    Kang, Yuan-Yuan
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2013, 18 (04) : 223 - 226
  • [10] Validation of the A&D UA-705 device for self-measurement of blood pressure according to the British Hypertension Society protocol
    Verdecchia, Paolo
    Angeli, Fabio
    Gattobigio, Roberto
    Angeli, Enrica
    Pede, Silvia
    Pittavini, Loretta
    BLOOD PRESSURE MONITORING, 2006, 11 (04) : 223 - 227