Validation of the A&D UA-1020 upper-arm blood pressure monitor for home blood pressure monitoring according to the British Hypertension Society Protocol
被引:2
|作者:
Zeng, Wei-Fang
[1
]
Kang, Yuan-Yuan
[1
]
Liu, Ming
[1
]
Li, Yan
[1
]
Wang, Ji-Guang
[1
]
机构:
[1] Shanghai Jiao Tong Univ, Ruijin Hosp,Sch Med, Shanghai Inst Hypertens, Ctr Epidemiol Studies & Clin Trials, Shanghai 200025, Peoples R China
关键词:
blood pressure determination;
blood pressure monitor;
British Hypertension Society Protocol;
validation studies;
EUROPEAN-SOCIETY;
D O I:
10.1097/MBP.0b013e32836141ca
中图分类号:
R6 [外科学];
学科分类号:
1002 ;
100210 ;
摘要:
Objective The present study aimed to evaluate the accuracy of the automated oscillometric upper-arm blood pressure (BP) monitor A&D UA-1020 with two different-shaped cuffs for home BP monitoring according to the British Hypertension Society (BHS) Protocol. Methods We recruited individuals for each of the two cuffs (D-ring and cylindrical) until there were 85 eligible participants (255 pairs of comparisons) and their BP could meet the BP distribution requirements specified by the BHS Protocol. For each participant, we sequentially measured the systolic and diastolic BP using a mercury sphygmomanometer (two observers) and the UA-1020 device (one supervisor). Results For the D-ring cuff, the device achieved grade A. The percentage of BP differences within 5, 10, and 15 mmHg was 67, 87, and 96%, respectively, for systolic BP, and 70, 90, and 99%, respectively, for diastolic BP. The average (+/- SD) of the device-observer differences was -0.2 +/- 7.3 mmHg (P = 0.64) and 1.7 +/- 5.8 mmHg (P < 0.0001) for systolic and diastolic BP, respectively. For the cylindrical cuff, the device also achieved grade A. The percentage of BP differences within 5, 10, and 15 mmHg was 67, 88, and 97%, respectively, for systolic BP and 64, 89, and 98%, respectively, for diastolic BP. The average of the device-observer differences was - 0.1 +/- 7.0mmHg (P = 0.89) and 2.0 +/- 6.3mmHg (P < 0.0001) for systolic and diastolic BP, respectively. Conclusion The UA-1020 device has passed the requirements of the BHS Protocol with both the D-ring and the cylindrical cuffs, and hence can be recommended for home use in adults. (c) 2013 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:177 / 181