Background: The effect of the atrial septal defect (ASD) shape on ASD size measurements remains unclear. We assessed the relationships between the stretched balloon diameter (SBD) and the diameters measured using two-dimensional and three-dimensional (3D) transesophageal echocardiography (TEE) and evaluated the effect of ASD shape on these relationships in patients with secundum ASD. Methods: We prospectively enrolled 107 consecutive patients who underwent transcatheter closure of ASD. The SBDs and ASD diameters on two-dimensional and 3D-TEE were measured. The circular index of the ASD was defined as the ratio of the maximal diameter to the minimal diameter on the 3D-TEE image. All patients were clinically followed up for a median of 1.3 months after the procedure. Results: The correlations and agreements between measurements were significantly better in patients with a round (circular index less than 1.5) than with an oval ASD. The differences between the SBDs and maximal diameter on 3D-TEE images were significantly smaller in patients with a large oval ASD than in the other patients. The differences between the size of the devices finally implanted and maximal diameter on 3D-TEE images were smaller in patients with a large oval than with a round ASD. On multivariate linear regression analysis, a formula of the relationship between the finally selected device size and 3D-TEE parameters was constructed: device size = 0.964 x maximal diameter on 3D-TEE image - 2.622 x circular index + 7.084. No major complications, including device embolization, developed during follow-up. Conclusions: The relationships between the SBDs and maximal diameters measured on two-dimensional and 3D-TEE images are influenced by the ASD shape and size. Therefore, the ASD shape, as well as the size, should be considered when the device size is determined without SBD measurement. (JAm Soc Echocardiogr 2012;25:1031-40.)
引用
收藏
页码:1031 / 1040
页数:10
相关论文
共 28 条
  • [1] Assessment of atrial septal defect size with 3D-transesophageal echocardiography:: Comparison with balloon method
    Abdel-Massih, T
    Dulac, Y
    Taktak, A
    Aggoun, Y
    Massabuau, P
    Elbaz, M
    Carrié, D
    Acar, P
    [J]. ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2005, 22 (02): : 121 - 127
  • [2] Influence of atrial septal defect anatomy in patient selection and assessment of closure with the Cardioseal device - A three-dimensional transoesophageal echocardiographic reconstruction
    Acar, P
    Saliba, Z
    Bonhoeffer, P
    Aggoun, Y
    Bonnet, D
    Sidi, D
    Kachaner, J
    [J]. EUROPEAN HEART JOURNAL, 2000, 21 (07) : 573 - 581
  • [3] Erosion of amplatzer septal occluder device after closure of secundum atrial septal defects: Review of registry of complications and recommendations to minimize future risk
    Amin, Z
    Hijazi, ZM
    Bass, JL
    Cheatham, JP
    Hellenbrand, WE
    Kleinman, CS
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 63 (04) : 496 - 502
  • [4] Transcatheter closure of multiple atrial septal defects - Initial results and value of two- and three-dimensional transoesophageal echocardiography
    Cao, QL
    Radtke, W
    Berger, F
    Zhu, W
    Hijazi, ZM
    [J]. EUROPEAN HEART JOURNAL, 2000, 21 (11) : 941 - 947
  • [5] Transcatheter closure of atrial septal defect and interatrial communications with a new self expanding nitinol double disc device (Amplatzer septal occluder): multicentre UK experience
    Chan, KC
    Godman, MJ
    Walsh, K
    Wilson, N
    Redington, A
    Gibbs, JL
    [J]. HEART, 1999, 82 (03) : 300 - 306
  • [6] Secundum atrial septal defect is a dynamic three-dimensional entity
    Dall'Agata, A
    McGhie, J
    Taams, MA
    Cromme-Dijkhuis, AH
    Spitaels, SEC
    Breburda, CSM
    Roelandt, JRTC
    Bogers, AJJC
    [J]. AMERICAN HEART JOURNAL, 1999, 137 (06) : 1075 - 1081
  • [7] Comparison of transcatheter closure of secundum atrial septal defect using the Amplatzer septal occluder associated with deficient versus sufficient rims
    Du, ZD
    Koenig, P
    Cao, DL
    Waight, D
    Heitschmidt, M
    Hijazi, ZM
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2002, 90 (08) : 865 - 869
  • [8] Franke A, 1997, CIRCULATION, V96, P323
  • [9] En Face View of Atrial Septal Defect by Two-Dimensional Transthoracic Echocardiography: Comparison to Real-Time Three-Dimensional Transesophageal Echocardiography
    Huang, Xinsheng
    Shen, Junjun
    Huang, Yigao
    Zheng, Zhichao
    Fei, Hongwen
    Hou, Yueshuang
    Xu, Yan
    Huang, Tao
    Huang, Wenhui
    Zhang, Caojin
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2010, 23 (07) : 714 - 721
  • [10] BIPLANE TRANSESOPHAGEAL ECHO-DOPPLER STUDIES OF ATRIAL SEPTAL-DEFECTS - QUANTITATIVE-EVALUATION AND MONITORING FOR TRANSCATHETER CLOSURE
    ISHII, M
    KATO, H
    INOUE, O
    TAKAGI, J
    MAENO, Y
    SUGIMURA, T
    MIYAKE, T
    KUMATE, M
    KOSUGA, K
    OHISHI, K
    [J]. AMERICAN HEART JOURNAL, 1993, 125 (05) : 1363 - 1368