Ultrasonic Scanner for In Vivo Measurement of Cancellous Bone Properties From Backscattered Data

被引:18
作者
Litniewski, Jerzy [1 ]
Cieslik, Lucyna [1 ]
Lewandowski, Marcin [1 ]
Tymkiewicz, Ryszard [1 ]
Zienkiewicz, Boguslaw [1 ]
Nowicki, Andrzej [1 ]
机构
[1] Polish Acad Sci, Inst Fundamental Technol Res, Ultrason Dept, Warsaw, Poland
关键词
HUMAN PROXIMAL FEMUR; QUANTITATIVE ULTRASOUND; TRABECULAR BONE; FREQUENCY-DEPENDENCE; ELDERLY-WOMEN; TRANSMISSION; OSTEOPOROSIS; DENSITY; ATTENUATION; CALCANEUS;
D O I
10.1109/TUFFC.2012.2347
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A dedicated ultrasonic scanner for acquiring RF echoes backscattered from the trabecular bone was developed. The design of device is based on the goal of minimizing of custom electronics and computations executed solely on the main computer processor and the graphics card. The electronic encoder-digitizer module executing all of the transmission and reception functions is based on a single low-cost field programmable gate array (FPGA). The scanner is equipped with a mechanical sector-scan probe with a concave transducer with 50 mm focal length, center frequency of 1.5 MHz and 60% bandwidth at -6 dB. The example of femoral neck bone examination shows that the scanner can provide ultrasonic data from deeply located bones with the ultrasound penetrating the trabecular bone up to a depth of 20 mm. It is also shown that the RF echo data acquired with the scanner allow for the estimation of attenuation coefficient and frequency dependence of backscattering coefficient of trabecular bone. The values of the calculated parameters are in the range of corresponding in vitro data from the literature but their variation is relatively high.
引用
收藏
页码:1470 / 1477
页数:8
相关论文
共 29 条
[1]   A device for in vivo measurements of quantitative ultrasound variables at the human proximal femur [J].
Barkmann, Reinhard ;
Laugier, Pascal ;
Moser, Urs ;
Dencks, Stefanie ;
Klausner, Michael ;
Padilla, Frederic ;
Haiat, Guilleaume ;
Glueer, Claus-C. .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2008, 55 (06) :1197-1204
[2]   In vivo ultrasonic attenuation slope estimates for detecting cervical ripening in rats:: Preliminary results [J].
Bigelow, Timothy A. ;
McFarlin, Barbara L. ;
O'Brien, William D., Jr. ;
Oelze, Michael L. .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2008, 123 (03) :1794-1800
[3]   Frequency dependence of ultrasonic backscattering in cancellous bone:: Autocorrelation model and experimental results [J].
Chaffaï, S ;
Roberjot, V ;
Peyrin, F ;
Berger, G ;
Laugier, P .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 108 (05) :2403-2411
[4]   Ultrasonic characterization of human cancellous bone using transmission and backscatter measurements:: Relationships to density and microstructure [J].
Chaffaî, S ;
Peyrin, F ;
Nuzzo, S ;
Porcher, R ;
Berger, G ;
Laugier, P .
BONE, 2002, 30 (01) :229-237
[5]   Ultrasonographic heel measurements to predict hip fracture in elderly women: The EPIDOS prospective study [J].
Hans, D ;
DargentMolina, P ;
Schott, AM ;
Sebert, JL ;
Cormier, C ;
Kotzki, PO ;
Delmas, PD ;
Pouilles, JM ;
Breart, G ;
Meunier, PJ .
LANCET, 1996, 348 (9026) :511-514
[6]   In vitro ultrasonic characterization of human cancellous femoral bone using transmission and backscatter measurements:: Relationships to bone mineral density [J].
Jenson, F ;
Padilla, F ;
Bousson, V ;
Bergot, C ;
Laredo, JD ;
Laugier, P .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2006, 119 (01) :654-663
[7]   Multi-site bone ultrasound measurements in elderly women with and without previous hip fractures [J].
Karjalainen, J. P. ;
Riekkinen, O. ;
Toyras, J. ;
Hakulinen, M. ;
Kroger, H. ;
Rikkonen, T. ;
Salovaara, K. ;
Jurvelin, J. S. .
OSTEOPOROSIS INTERNATIONAL, 2012, 23 (04) :1287-1295
[8]   ULTRASOUND BACKSCATTER IMAGING PROVIDES FREQUENCY-DEPENDENT INFORMATION ON STRUCTURE, COMPOSITION AND MECHANICAL PROPERTIES OF HUMAN TRABECULAR BONE [J].
Karjalainen, Janne P. ;
Toyras, Juha ;
Riekkinen, Ossi ;
Hakulinen, Mikko ;
Jurvelin, Jukka S. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2009, 35 (08) :1376-1384
[9]  
LANGTON CM, 1994, CLIN RHEUMATOL, V13, P13
[10]  
Laugier P, 2000, ULTRASON, P1341, DOI 10.1109/ULTSYM.2000.921571