Study of the factors affecting the performance of microextraction by packed sorbent (MEPS) using liquid scintillation counter and liquid chromatography-tandem mass spectrometry

被引:81
作者
Altun, Zeki [1 ]
Abdel-Rehim, Mohamed [1 ,2 ,3 ]
机构
[1] Karlstad Univ, Fac Sci & Technol, SE-65188 Karlstad, Sweden
[2] AstraZeneca R&D Sodertalje, Clin Pharmacol, SE-15185 Sodertalje, Sweden
[3] AstraZeneca R&D Sodertalje, DMPK, SE-15185 Sodertalje, Sweden
关键词
Microextraction by packed sorbent; Sample preparation; Liquid scintillation counter; Mass spectrometry; Radioactively labelled bupivacaine;
D O I
10.1016/j.aca.2008.09.067
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Microextraction by packed sorbent (MEPS) is a new technique for sample preparation that can be connected on-line with LC or GC. In MEPS, approximately 1-2 mg of the solid packing material is inserted into a syringe (100-250 mu L) as a plug. Sample preparation takes place on the packed bed. The bed can be packed or coated to provide selective and suitable sampling conditions. The new method is very promising for extraction of drugs and metabolites from biological samples. In this paper, some factors affecting the performance of MEPS such as recovery, carry-over, leakage, washing volume and elution volume were studied using C18 and hydroxylated polystyrene-divinylbenzene copolymer (ENV+) as sorbents. Radioactively labelled bupivacaine in plasma samples was used as test analyte. For the extraction of this drug, using methanol/water 95:5 (v/v) (0.25% ammonium hydroxide) was used as elution solvent. The analyte response increased with increasing the elution volume and it was linear upp up to 100 mu L utilizing liquid scintillation counter. Further, for concentrating the sample, we found that MEPS may be used such that the sample can be drawn through the needle, up and down, several times. The analyte leakage increases as the volume washing increases, though higher washing volumes may also result in cleaner extracts. To eliminate analyte carry-over, the sorbents were washed first with 3 x 250 mu L elution solution and then with 3 x 250 mu L, washing solution. In addition, the reproducibility measurements show relatively good relative standard deviation (RSD) % values concerning analyte recovery and analyte leakage. The present study provides an understanding of basic aspects when optimizing methods for MEPS. In this study, MEPS was used off-line with liquid scintillation counter and on-line with LC-MS/MS. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:116 / 123
页数:8
相关论文
共 38 条
[1]   Microextraction in packed syringe online with liquid chromatography-tandem mass spectrometry: Molecularly imprinted polymer as packing material for MEPS in selective extraction of ropivacaine from plasma [J].
Abdel-Rehim, M. ;
Andersson, L. I. ;
Altun, Z. ;
Blomberg, L. G. .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2006, 29 (12) :1725-1736
[2]   Microextraction in packed syringe/liquid chromatography/electrospray tandem mass spectrometry for quantification of olomoucine in human plasma samples [J].
Abdel-Rehim, M ;
Skansen, P ;
Vita, M ;
Hassan, Z ;
Blomberg, L ;
Hassan, M .
ANALYTICA CHIMICA ACTA, 2005, 539 (1-2) :35-39
[3]   Microextraction in packed syringe (MEPS) for liquid and gas chromatographic applications. Part II - Determination of ropivacaine and its metabolites in human plasma samples using MEPS with liquid chromatography/tandem mass spectrometry [J].
Abdel-Rehim, M ;
Altun, Z ;
Blomberg, L .
JOURNAL OF MASS SPECTROMETRY, 2004, 39 (12) :1488-1493
[4]   New trend in sample preparation: on-line microextraction in packed syringe for liquid and gas chromatography applications - I. Determination of local anaesthetics in human plasma samples using gas chromatography-mass spectrometry [J].
Abdel-Rehim, M .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 801 (02) :317-321
[5]   Simultaneous determination of busulphan in plasma samples by liquid chromatography-electrospray ionization mass spectrometry utilizing microextraction in packed syringe (MEPS) as on-line sample preparation method [J].
Abdel-Rehim, Mohamed ;
Hassan, Zuzana ;
Skansem, Patrik ;
Hassan, Moustapha .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2007, 30 (20) :3029-3041
[6]   Microextraction in packed syringe (MEPS) utilizing methylcyanopropyl-silarylene as coating polymer for extraction of drugs in biological samples [J].
Abdel-Rehim, Mohamed ;
Dahlgren, Marie ;
Blomberg, Lars ;
Claude, Saturnin ;
Tabacchi, Raphael .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2006, 29 (17) :2537-2544
[7]   Quantification of 4-OH-2,6-xylidine and its conjugates in human urine samples utilising microextraction in packed syringe on-line with liquid chromatography and electrospray tandem mass spectrometry (MEPS-LC-MS/MS) [J].
Abdel-Rehim, Mohamed ;
Askemark, Yvonne ;
Norsten-Hoog, Carina ;
Pettersson, Karl-Johan ;
Halldin, Magnus .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2006, 29 (16) :2413-2424
[8]   Quantification of ropivacaine and its major metabolites in human urine samples utilizing microextraction in a packed syringe automated with liquid chromatography-tandem mass spectrometry (MEPS-LC-MS/MS) [J].
Abdel-Rehim, Mohamed ;
Dahlgren, Marie ;
Blomberg, Lars .
JOURNAL OF SEPARATION SCIENCE, 2006, 29 (11) :1658-1661
[9]  
ABDELREHIM M, J CHROMATOG IN PRESS
[10]   Drug screening using microextraction in a packed syringe (MEPS)/mass spectrometry utilizing monolithic-, polymer-, and silica-based sorbents [J].
Altun, Z ;
Blomberg, LG ;
Jagerdeo, E ;
Abdel-Rehim, M .
JOURNAL OF LIQUID CHROMATOGRAPHY & RELATED TECHNOLOGIES, 2006, 29 (06) :829-840