Treatment planning for prostate focal laser ablation in the face of needle placement uncertainty

被引:15
作者
Cepek, Jeremy [1 ,2 ]
Lindner, Uri [3 ]
Davidson, Sean R. H. [4 ]
Haider, Masoom A. [5 ,6 ]
Ghai, Sangeet [7 ]
Trachtenberg, John [8 ]
Fenster, Aaron [1 ,2 ]
机构
[1] John P Robarts Res Inst, London, ON N6A 5K8, Canada
[2] Univ Western Ontario, London, ON N6A 5B9, Canada
[3] Univ Hlth Network, Div Urol, Dept Surg Oncol, Toronto, ON M5G 2C4, Canada
[4] Univ Hlth Network, Ontario Canc Inst, Toronto, ON M5G 2M9, Canada
[5] Sunnybrook Hlth Sci Ctr, Dept Med Imaging, Toronto, ON M4N 3M5, Canada
[6] Univ Toronto, Dept Med Imaging, Toronto, ON M5S 2J7, Canada
[7] Univ Hlth Network, Dept Med Imaging, Toronto, ON M5G 2M9, Canada
[8] Univ Hlth Network, Div Urol, Dept Surg Oncol, Toronto, ON M5G 2C4, Canada
基金
加拿大自然科学与工程研究理事会; 加拿大健康研究院;
关键词
prostate cancer; focal therapy; treatment planning; needle guidance; RADIOFREQUENCY ABLATION; THERMAL THERAPY; MULTIPARAMETRIC MRI; SEED PLACEMENT; CANCER; BRACHYTHERAPY; IMPLANT; TUMORS; BIOPSY; THERMOTHERAPY;
D O I
10.1118/1.4842535
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To study the effect of needle placement uncertainty on the expected probability of achieving complete focal target destruction in focal laser ablation (FLA) of prostate cancer. Methods: Using a simplified model of prostate cancer focal target, and focal laser ablation region shapes, Monte Carlo simulations of needle placement error were performed to estimate the probability of completely ablating a region of target tissue. Results: Graphs of the probability of complete focal target ablation are presented over clinically relevant ranges of focal target sizes and shapes, ablation region sizes, and levels of needle placement uncertainty. In addition, a table is provided for estimating the maximum target size that is treatable. The results predict that targets whose length is at least 5 mm smaller than the diameter of each ablation region can be confidently ablated using, at most, four laser fibers if the standard deviation in each component of needle placement error is less than 3 mm. However, targets larger than this (i.e., near to or exceeding the diameter of each ablation region) require more careful planning. This process is facilitated by using the table provided. Conclusions: The probability of completely ablating a focal target using FLA is sensitive to the level of needle placement uncertainty, especially as the target length approaches and becomes greater than the diameter of ablated tissue that each individual laser fiber can achieve. The results of this work can be used to help determine individual patient eligibility for prostate FLA, to guide the planning of prostate FLA, and to quantify the clinical benefit of using advanced systems for accurate needle delivery for this treatment modality. (C) 2014 Author(s). All article content, except where otherwise noted, is licensed under a Creative Commons Attribution 3.0 Unported License.
引用
收藏
页数:14
相关论文
共 52 条
[1]   Focal Therapy in Prostate Cancer: Determinants of Success and Failure [J].
Ahmed, Hashim Uddin ;
Moore, Caroline ;
Lecornet, Emilie ;
Emberton, Mark .
JOURNAL OF ENDOUROLOGY, 2010, 24 (05) :819-825
[2]  
Alterovitz R., 2007, P ROB SCI SYST ATL G
[3]   Motion Planning Under Uncertainty for Image-guided Medical Needle Steering [J].
Alterovitz, Ron ;
Branicky, Michael ;
Goldberg, Ken .
INTERNATIONAL JOURNAL OF ROBOTICS RESEARCH, 2008, 27 (11-12) :1361-1374
[4]   INTERSTITIAL LASER PHOTOCOAGULATION FOR THE TREATMENT OF PROSTATIC-CANCER [J].
AMIN, Z ;
LEES, WR ;
BOWN, SG .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (791) :1044-1047
[5]  
Atri M, 2009, CUAJ-CAN UROL ASSOC, V3, P125
[6]   Radiofrequency Ablation of Lung Tumors in Swine Assisted by a Navigation Device with Preprocedural Volumetric Planning [J].
Banovac, Filip ;
Cheng, Patrick ;
Campos-Nanez, Enrique ;
Kallakury, Bhaskar ;
Popa, Teo ;
Wilson, Emmanuel ;
Abeledo, Hernan ;
Cleary, Kevin .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2010, 21 (01) :122-129
[7]   The robustness of dose distributions to displacement and migration of 125I permanent seed implants over a wide range of seed number, activity, and designs [J].
Beaulieu, L ;
Archambault, L ;
Aubin, S ;
Oral, E ;
Taschereau, R ;
Pouliot, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 58 (04) :1298-1308
[8]   Transperineal prostate biopsy under magnetic resonance image guidance: A needle placement accuracy study [J].
Blumenfeld, Philip ;
Hata, Nobuhiko ;
DiMaio, Simon ;
Zou, Kelly ;
Haker, Steven ;
Fichtinger, Gabor ;
Tempany, Clare M. C. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 26 (03) :688-694
[9]   Prostate tissue analysis immediately following magnetic resonance imaging guided transurethral ultrasound thermal therapy [J].
Boyes, Aaron ;
Tang, Kee ;
Yaffe, Martin ;
Sugar, Linda ;
Chopra, Rajiv ;
Bronskill, Michael .
JOURNAL OF UROLOGY, 2007, 178 (03) :1080-1085
[10]   Focal therapy of prostate cancer: energies and procedures [J].
Bozzini, G. ;
Colin, P. ;
Nevoux, P. ;
Villers, A. ;
Mordon, S. ;
Betrouni, N. .
UROLOGIC ONCOLOGY-SEMINARS AND ORIGINAL INVESTIGATIONS, 2013, 31 (02) :155-167