Simultaneous determination of iodine, bromine, selenium and arsenic in geological samples by inductively coupled plasma mass spectrometry

被引:31
作者
Li, B [1 ]
He, HL [1 ]
Shi, SY [1 ]
Ma, XR [1 ]
Wen, HL [1 ]
Lu, CF [1 ]
机构
[1] Natl Res Ctr Geoanal, Beijing 100037, Peoples R China
关键词
D O I
10.1039/b107161k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The feasibility of determining trace iodine. bromine, selenium and arsenic simultaneously in geological samples by inductively coupled plasma mass spectrometry (ICP-MS) was investigated. The samples were mixed with flux (sodium carbonate and zinc oxide), sintered at 650 degreesC for 40 min in a muffle furnace and then dissolved in hot water. The cations were separated from the sample solution by using a cation-exchange resin. The analyte anions were determined by ICP-MS directly. The signal responses for the different anion species in different solution media were assessed. The sintering temperature, holding time, and amounts of flux and cation-exchange resin were optimised, The ion signal intensity of I- in nitric acid was much higher than that OF IO3- and IO4- In addition, the observed I- signal in nitric acid was not stable. Weak acid solution or dilute ammonia solution was found to be a suitable medium for determination of these elements. Ammonia solution is also effective in eliminating the memory effect of iodine and bromine. The method was applied to the determination of trace Br, I, Se and As in a series of Chinese geological reference materials (GRM, soils and stream sediments). Most or the results were found to be in reasonable agreement, with the certified values. The limits of quantitation for Br, I, Se and As were 0.15, 0.03, 0.004 and 0.04 mug g(-1) respectively. Precision was between 0.8 and 2.8% RSD. It was shown that the method has great potential for the routine determination of total iodine, bromine, selenium and arsenic in geological samples.
引用
收藏
页码:371 / 376
页数:6
相关论文
共 15 条
[1]   DETERMINATION OF IODINE AND BROMINE IN PLASMA AND URINE BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ALLAIN, P ;
MAURAS, Y ;
DOUGE, C ;
JAUNAULT, L ;
DELAPORTE, T ;
BEAUGRAND, C .
ANALYST, 1990, 115 (06) :813-815
[2]   RAPID AND SENSITIVE DETERMINATION OF IODINE IN FRESH MILK AND MILK POWDER BY INDUCTIVELY COUPLED PLASMA - MASS-SPECTROMETRY (ICP-MS) [J].
BAUMANN, H .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 338 (07) :809-812
[3]  
CHENG S, 1957, IMPORTANT INORGANIC
[4]   DETERMINATION OF I-129 IN VEGETABLE SAMPLES BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
COX, RJ ;
PICKFORD, CJ ;
THOMPSON, M .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (04) :635-640
[5]   APPLICATION OF INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY TO THE SIMULTANEOUS DETERMINATION OF CHLORINE, BROMINE AND IODINE IN THE NATIONAL-BUREAU-OF-STANDARDS STANDARD REFERENCE MATERIAL-1648 URBAN PARTICULATE [J].
DATE, AR ;
STUART, ME .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (05) :659-665
[6]  
FANG R, 1988, GEOLOGICAL LAB, V4, P16
[7]  
FANG R, 1993, CHINESE J ANAL CHEM, V21, P364
[8]   SIMULTANEOUS DETERMINATION OF FLUORINE AND CHLORINE IN SILICATE ROCKS BY A RAPID SPECTROPHOTOMETRIC METHOD [J].
HUANG, WH ;
JOHNS, WD .
ANALYTICA CHIMICA ACTA, 1967, 37 (04) :508-&
[9]  
Jarvis K. E., 1992, HDB ICP MS
[10]   Determination of iodine in food-related certified reference materials using wet ashing and detection by inductively coupled plasma mass spectrometry [J].
Larsen, EH ;
Ludwigsen, MB .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (04) :435-439