Image distortion in MRI-based polymer gel dosimetry of Gamma Knife stereotactic radiosurgery systems

被引:32
作者
Watanabe, Y
Perera, GM
Mooij, RB
机构
[1] Columbia Univ, Coll Phys & Surg, Dept Radiat Oncol, New York, NY 10032 USA
[2] Columbia Univ, Coll Phys & Surg, Dept Radiol, New York, NY 10032 USA
关键词
BANG-3 polymer gel; absorbed dose; Gamma Knife; MRI; magnetic susceptibility;
D O I
10.1118/1.1470204
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We have studied effects of MR (magnetic resonance) image distortion on polymer gel dosimetry of Gamma Knife stereotactic radiosurgery systems. MR images of BANG polymer gel phantoms were acquired by using a Hahn spin-echo sequence and a fast 3D imaging GRASS sequence. Image artifacts were studied by varying the directions of frequency encoding and the receiver bandwidth. The phantoms were also CT (computed tomography) scanned. The studies showed that the measured dose distributions are shifted by 1.8+/-0.5 mm (2 pixels) in the frequency encoding direction. The magnitude of the shift is inversely proportional to the receiver bandwidth in agreement with theory. Comparison of MRI with CT showed the same image shift. We concluded that the discrepancy is caused by MR image distortion due to a difference in susceptibility effects between the phantom and the fiducial markers of the Leksell localization box. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:797 / 802
页数:6
相关论文
共 18 条
[1]   MAGNETIC-RESONANCE IMAGE-DIRECTED STEREOTAXIC NEUROSURGERY - USE OF IMAGE FUSION WITH COMPUTERIZED-TOMOGRAPHY TO ENHANCE SPATIAL ACCURACY [J].
ALEXANDER, E ;
KOOY, HM ;
VANHERK, M ;
SCHWARTZ, M ;
BARNES, PD ;
TARBELL, N ;
MULKERN, RV ;
HOLUPKA, EJ ;
LOEFFLER, JS .
JOURNAL OF NEUROSURGERY, 1995, 83 (02) :271-276
[2]   ANALYSIS OF MACHINE-DEPENDENT AND OBJECT-INDUCED GEOMETRIC DISTORTION IN 2DFT MR IMAGING [J].
BAKKER, CJG ;
MOERLAND, MA ;
BHAGWANDIEN, R ;
BEERSMA, R .
MAGNETIC RESONANCE IMAGING, 1992, 10 (04) :597-608
[3]   MRI distortion and stereotactic neurosurgery using the Cosman-Roberts-Wells and Leksell frames [J].
Burchiel, KJ ;
Nguyen, TT ;
Coombs, BD ;
Szumoski, J .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1996, 66 (1-3) :123-136
[4]   High-resolution dose profile studies based on MR Imaging with polymer BANG™ gels in stereotactic radiation techniques [J].
Ertl, A ;
Berg, A ;
Zehetmayer, M ;
Frigo, P .
MAGNETIC RESONANCE IMAGING, 2000, 18 (03) :343-349
[5]   Quality assurance for the Leksell gamma unit: Considering magnetic resonance image-distortion and delineation failure in the targeting of the internal auditory canal [J].
Ertl, A ;
Saringer, W ;
Heimberger, K ;
Kindl, P .
MEDICAL PHYSICS, 1999, 26 (02) :166-170
[6]   New dosimetric approach for multidimensional dose evaluation in gamma knife radiosurgery - Technical note [J].
Foroni, R ;
Gambraini, G ;
Danesi, U ;
Mauri, M ;
Pompilio, E ;
Pirola, L ;
Nicolato, A ;
Ferraresi, P ;
Gerosa, M .
JOURNAL OF NEUROSURGERY, 2000, 93 :239-242
[7]   An evaluation of the accuracy of magnetic-resonance-guided gamma knife surgery [J].
Guo, WY ;
Chu, WC ;
Wu, MC ;
Chung, WY ;
Gwan, WP ;
Lee, YL ;
Pan, HC ;
Chang, CY .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1996, 66 :85-92
[8]  
Haake E, 1999, MAGNETIC RESONANCE I
[9]   Three-dimensional visualization and measurement of conformal dose distributions using magnetic resonance imaging of BANG polymer gel dosimeters [J].
Ibbott, GS ;
Maryanski, MJ ;
Eastman, P ;
Holcomb, SD ;
Zhang, YS ;
Avison, RG ;
Sanders, M ;
Gore, JC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (05) :1097-1103
[10]   A COMPARISON BETWEEN MAGNETIC-RESONANCE-IMAGING AND COMPUTED-TOMOGRAPHY FOR STEREOTAXIC COORDINATE DETERMINATION [J].
KONDZIOLKA, D ;
DEMPSEY, PK ;
LUNSFORD, D ;
HEILBRUN, MP ;
KESTLE, JRW ;
DOLAN, EJ ;
KANAL, E ;
TASKER, RR .
NEUROSURGERY, 1992, 30 (03) :402-407