Influence of Implant Connection, Abutment Design and Screw Insertion Torque on Implant-Abutment Misfit

被引:27
作者
Velez, Jorge [1 ]
Pelaez, Jesus [1 ]
Lopez-Suarez, Carlos [1 ]
Agustin-Panadero, Ruben [2 ]
Tobar, Celia [1 ]
Suarez, Maria J. [1 ]
机构
[1] Univ Complutense Madrid, Dept Conservat Dent & Bucofacial Prosthesis, Fac Odontol, Madrid 28040, Spain
[2] Univ Valencia, Dept Dent Med, Fac Med & Dent, Valencia 46010, Spain
关键词
external connection; internal connection; abutment; implant-abutment interface; misfit; IN-VITRO EVALUATION; MARGINAL FIT; ZIRCONIA ABUTMENTS; SEALING CAPABILITY; TITANIUM ABUTMENTS; BACTERIAL LEAKAGE; EXTERNAL CONNECTION; MICROBIAL LEAKAGE; INTERFACE; MICROLEAKAGE;
D O I
10.3390/jcm9082365
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: An accurate fit at the implant-abutment interface is an important factor to avoid biological and mechanical complications. The aim of this study was to evaluate the marginal misfit at the implant-abutment interface on external and Morse taper connection, with straight and angulated abutments under different insertion torque loads. Materials and Methods: A total of 120 implants were used, 60 with external connection (EC) and 60 with Morse taper connection (IC). Straight (SA) (n= 60) and angulated abutments (AA) (n= 60) were randomly screwed to each connection at different torque levels (n= 10 each): 10, 20 and 30 Ncm. All specimens were subjected to thermal and cyclic loading and the misfit was measured by scanning electron microscopy. Data were analyzed with one-way ANOVA,t-test and Kruskal-Wallis test. Results: Significant differences (p< 0.001) were found between connections and abutments regardless of the torque applied. Morse taper connections with straight and angulated abutments showed the lowest misfit values (0.6 mu m). Misfit values decreased as torque increased. Conclusions: The misfit was affected by the type of connection. The type of abutment did not influence the fit in the Morse taper connection. The higher the tightening torque applied the increase in the fit of the implant-abutment interface.
引用
收藏
页码:1 / 13
页数:13
相关论文
共 59 条
[1]   OSSEOINTEGRATED TITANIUM IMPLANTS - REQUIREMENTS FOR ENSURING A LONG-LASTING, DIRECT BONE-TO-IMPLANT ANCHORAGE IN MAN [J].
ALBREKTSSON, T ;
BRANEMARK, PI ;
HANSSON, HA ;
LINDSTROM, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (02) :155-170
[2]   Effects of different torque levels on the implant-abutment interface in a conical internal connection [J].
Alexandre Gehrke, Sergio ;
Shibli, Jamil Awad ;
Aramburu Junior, Jaime Sarda ;
Mate Sanchez de Val, Jose Eduardo ;
Luis Calvo-Girardo, Jose ;
Dedavid, Berenice Anina .
BRAZILIAN ORAL RESEARCH, 2016, 30 (01)
[3]   A Comparison of Precision of Fit, Rotational Freedom, and Torque Loss with Copy-Milled Zirconia and Prefabricated Titanium Abutments [J].
Alikhasi, Marzieh ;
Monzavi, Abbas ;
Bassir, Seyed Hossein ;
Naini, Roshanak Baghai ;
Khosronedjad, Nasim ;
Keshavarz, Sareh .
INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2013, 28 (04) :996-1002
[4]   Original vs. non-original abutments for screw-retained single implant crowns: An in vitro evaluation of internal fit, mechanical behaviour and screw loosening [J].
Alonso-Perez, Raquel ;
Bartolome, Jose F. ;
Ferreiroa, Alberto ;
Salido, Maria P. ;
Pradies, Guillermo .
CLINICAL ORAL IMPLANTS RESEARCH, 2018, 29 (12) :1230-1238
[5]  
[Anonymous], 2013, 148012007 ISO
[6]  
Baixe S, 2010, INT J ORAL MAX IMPL, V25, P455
[7]   Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study [J].
Benjaboonyazit, Kanyarin ;
Chaijareenont, Pisaisit ;
Khongkhunthian, Pathawee .
INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2019, 5 (1)
[8]  
Binon P P, 1996, Implant Dent, V5, P235, DOI 10.1097/00008505-199600540-00002
[9]  
Branemark P I, 1977, Scand J Plast Reconstr Surg Suppl, V16, P1
[10]  
Breeding L C, 1993, Int J Prosthodont, V6, P435