Initial Experience with Surpass Evolve Flow Diverter in the Treatment of Intracranial Aneurysms

被引:19
作者
Maus, V. [1 ]
Weber, W. [1 ]
Berlis, A. [2 ]
Maurer, C. [2 ]
Fischer, S. [1 ]
机构
[1] Ruhr Univ, Knappschaftskrankenhaus Bochum, Dept Diagnost & Intervent Neuroradiol & Nucl Med, Schornau 23-25, D-44892 Bochum, Germany
[2] Univ Hosp Augsburg, Dept Diagnost & Intervent Neuroradiol, Augsburg, Germany
关键词
Endovascular treatment; Intracranial stenting; Flow diversion; Interventional neuroradiology; Aneurysm occlusion; PIPELINE EMBOLIZATION DEVICE;
D O I
10.1007/s00062-020-00972-5
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background and Purpose The principle of flow diversion has revolutionized the treatment of brain aneurysms. In this study, we report our experience of the new Surpass Evolve (SE) flow diverter in the treatment of intracranial aneurysms. Material and Methods Patients were treated with the SE as first-line therapy between May 2019 and June 2020 at 2 experienced institutions. Inclusion criteria were wide-necked, blister-like, or fusiform/dissecting aneurysms in the anterior and posterior circulation. Primary endpoint was technical success defined as favorable navigation to the target vessel and successful deployment of the SE. Secondary endpoints were favorable aneurysm occlusion defined as O'Kelly Marotta (OKM) scale C1-3 + D on follow-up, procedure-related complications and retreatment. Results A total of 46 aneurysms in 42 patients were treated with 57 SE flow diverters. Median aneurysm size was 6.6 mm (IQR 4.0-12.2 mm) with a median neck width of 4 mm (IQR 2.2-5.4 mm). On admission, 6 (13%) aneurysms were ruptured and 41 (89%) were located in the anterior circulation. The primary endpoint was reached in 96%. Median follow-up was 116 days (IQR 92-134 days) and available for 36/46 (78%) aneurysms. Favorable aneurysm occlusion was seen in 31/36 (86%) aneurysms and 27/36 (75%) were occluded completely. Parent artery occlusion appeared in 3 (3%) patients on follow-up and 2 aneurysms (6%) required additional treatment due to insufficient closure. Conclusion The new SE flow diverter is safe and seems to be effective with promising occlusion rates at short-term follow-up.
引用
收藏
页码:681 / 689
页数:9
相关论文
共 17 条
[1]   Long-Term Clinical and Angiographic Outcomes Following Pipeline Embolization Device Treatment of Complex Internal Carotid Artery Aneurysms: Five-Year Results of the Pipeline for Uncoilable or Failed Aneurysms Trial [J].
Becske, Tibor ;
Brinjikji, Waleed ;
Potts, Matthew B. ;
Kallmes, David F. ;
Shapiro, Maksim ;
Moran, Christopher J. ;
Levy, Elad I. ;
McDougall, Cameron G. ;
Szikora, Istvan ;
Lanzino, Giuseppe ;
Woo, Henry H. ;
Lopes, Demetrius K. ;
Siddiqui, Adnan H. ;
Albuquerque, Felipe C. ;
Fiorella, David J. ;
Saatci, Isil ;
Cekirge, Saruhan H. ;
Berez, Aaron L. ;
Cher, Daniel J. ;
Berentei, Zsolt ;
Marosfoi, Miklos ;
Nelson, Peter K. .
NEUROSURGERY, 2017, 80 (01) :40-48
[2]   Comparison of intra-aneurysmal flow modification using optical flow imaging to evaluate the performance of Evolve and Pipeline flow diverting stents [J].
Cancelliere, Nicole Mariantonia ;
Nicholson, Patrick ;
Radovanovic, Ivan ;
Mendes, Karla Mirella ;
Orru, Emanuele ;
Krings, Timo ;
Pereira, Vitor M. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2020, 12 (08) :814-817
[3]   Prospective study on embolization of intracranial aneurysms with the pipeline device: the PREMIER study 1 year results [J].
Hanel, Ricardo A. ;
Kallmes, David F. ;
Lopes, Demetrius Klee ;
Nelson, Peter Kim ;
Siddiqui, Adnan ;
Jabbour, Pascal ;
Pereira, Vitor M. ;
Istvan, Istvan Szikora ;
Zaidat, Osama O. ;
Bettegowda, Chetan ;
Colby, Geoffrey P. ;
Mokin, Maxim ;
Schirmer, Clemens ;
Hellinger, Frank R. ;
Given, Curtis, II ;
Krings, Timo ;
Taussky, Philipp ;
Toth, Gabor ;
Fraser, Justin F. ;
Chen, Michael ;
Priest, Ryan ;
Kan, Peter ;
Fiorella, David ;
Frei, Don ;
Aagaard-Kienitz, Beverly ;
Diaz, Orlando ;
Malek, Adel M. ;
Cawley, C. Michael ;
Puri, Ajit S. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2020, 12 (01) :62-66
[4]   A new endoluminal, flow-disrupting device for treatment of saccular aneurysms [J].
Kallmes, David F. ;
Ding, Yong Hong ;
Dai, Daying ;
Kadirvel, Ramanathan ;
Lewis, Debra A. ;
Cloft, Harry J. .
STROKE, 2007, 38 (08) :2346-2352
[5]   Meta-analysis of treatment outcomes of posterior circulation non-saccular aneurysms by flow diverters [J].
Kiyofuji, Satoshi ;
Graffeo, Christopher S. ;
Perry, Avital ;
Murad, Mohammad Hassan ;
Flemming, Kelly D. ;
Lanzino, Giuseppe ;
Rangel-Castilla, Leonardo ;
Brinjikji, Waleed .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2018, 10 (05) :500-506
[6]   The Use of Flow Diverter in Ruptured, Dissecting Intracranial Aneurysms of the Posterior Circulation [J].
Maus, Volker ;
Mpotsaris, Anastasios ;
Dorn, Franziska ;
Moehlenbruch, Markus ;
Borggrefe, Jan ;
Stavrinou, Pantelis ;
Abdullayev, Nuran ;
Barnikol, Utako Birgit ;
Liebig, Thomas ;
Kabbasch, Christoph .
WORLD NEUROSURGERY, 2018, 111 :E424-E433
[7]   SCENT Trial: One-Year Outcomes [J].
Meyers, Philip M. ;
Coon, Alexander L. ;
Kan, Peter T. ;
Wakhloo, Ajay K. ;
Hanel, Ricardo A. .
STROKE, 2019, 50 (06) :1473-1479
[8]   The Pipeline Embolization Device for the Intracranial Treatment of Aneurysms Trial [J].
Nelson, P. K. ;
Lylyk, P. ;
Szikora, I. ;
Wetzel, S. G. ;
Wanke, I. ;
Fiorella, D. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2011, 32 (01) :34-40
[9]   A Novel Grading Scale for the Angiographic Assessment of Intracranial Aneurysms Treated Using Flow Diverting Stents [J].
O'Kelly, C. J. ;
Krings, T. ;
Fiorella, D. ;
Marotta, T. R. .
INTERVENTIONAL NEURORADIOLOGY, 2010, 16 (02) :133-137
[10]   First clinical experience with the new Surpass Evolve flow diverter: technical and clinical considerations [J].
Orru, Emanuele ;
Rice, Henry ;
De Villiers, Laetitia ;
Klostranec, Jesse M. ;
Wakhloo, Ajay K. ;
Coon, Alexander L. ;
Radovanovic, Ivan ;
Kortman, Hans ;
Bhatia, Kartik Dev ;
Krings, Timo ;
Pereira, Vitor M. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2020, 12 (10) :974-980