A new use for an old Holter monitor: An ambulatory cough meter

被引:47
作者
Chang, AB
Newman, RG
Phelan, PD
Robertson, CF
机构
[1] UNIV MELBOURNE,ROYAL CHILDRENS HOSP,DEPT BIOMED ENGN,MELBOURNE,VIC 3052,AUSTRALIA
[2] UNIV MELBOURNE,ROYAL CHILDRENS HOSP,DEPT PAEDIAT,MELBOURNE,VIC 3052,AUSTRALIA
关键词
children; cough; cough meter;
D O I
10.1183/09031936.97.10071637
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Cough is commonly used as an outcome measure in clinical studies, although the subjective reporting of cough is unreliable when compared to objective measures, We describe an inexpensive new ambulatory cough meter that is based on a disused Bolter monitor, The cough meter consists of a Holter monitor and a cough processor, designed on a computer to select the most appropriate filters. The cough meter was then validated against the overnight tape recorder on 21 occasions in 18 children (aged 6-15 yrs). The agreement between the cough meter and the tape recorder was good (mean difference of -0.3 coughs.h(-1); limits of agreement -2.2 to 1.7 coughs.h(-1)). We conclude that our newly described ambulatory cough meter provides a valid and inexpensive method of objectively monitoring cough for up to 24 h.
引用
收藏
页码:1637 / 1639
页数:3
相关论文
共 15 条
[1]   NIGHT COUGH COUNTS AND DIARY CARD SCORES IN ASTHMA [J].
ARCHER, LNJ ;
SIMPSON, H .
ARCHIVES OF DISEASE IN CHILDHOOD, 1985, 60 (05) :473-474
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   AN ELECTROMYOGRAPHIC METHOD OF OBJECTIVELY ASSESSING COUGH INTENSITY AND USE OF THE METHOD TO ASSESS EFFECTS OF CODEINE ON THE DOSE-RESPONSE CURVE TO CITRIC-ACID [J].
COX, ID ;
WALLIS, PJW ;
APPS, MCP ;
HUGHES, DTD ;
EMPEY, DW ;
OSMAN, RCA ;
BURKE, CA .
BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1984, 18 (03) :377-382
[4]   THE INFLUENCE OF PSYCHOLOGICAL STATUS ON RESPIRATORY SYMPTOM REPORTING [J].
DALES, RE ;
SPITZER, WO ;
SCHECHTER, MT ;
SUISSA, S .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1989, 139 (06) :1459-1463
[5]   POOR AGREEMENT BETWEEN REPORTED AND RECORDED NOCTURNAL COUGH IN ASTHMA [J].
FALCONER, A ;
OLDMAN, C ;
HELMS, P .
PEDIATRIC PULMONOLOGY, 1993, 15 (04) :209-211
[6]   ELECTROMYOGRAM PATTERN OF DIAPHRAGMATIC FATIGUE [J].
GROSS, D ;
GRASSINO, A ;
ROSS, WRD ;
MACKLEM, PT .
JOURNAL OF APPLIED PHYSIOLOGY, 1979, 46 (01) :1-7
[7]   COUGHING FREQUENCY IN PATIENTS WITH PERSISTENT COUGH - ASSESSMENT USING A 24-HOUR AMBULATORY RECORDER [J].
HSU, JY ;
STONE, RA ;
LOGANSINCLAIR, RB ;
WORSDELL, M ;
BUSST, CM ;
CHUNG, KF .
EUROPEAN RESPIRATORY JOURNAL, 1994, 7 (07) :1246-1253
[8]   EPIDEMIOLOGIC-STUDY OF WHEEZE, DOCTOR DIAGNOSED ASTHMA, AND COUGH IN PRESCHOOL-CHILDREN IN LEICESTERSHIRE [J].
LUYT, DK ;
BURTON, PR ;
SIMPSON, H .
BMJ-BRITISH MEDICAL JOURNAL, 1993, 306 (6889) :1386-1390
[9]  
MAUTONE AJ, 1989, PULMONARY PHYSL, P198
[10]   A NEW DEVICE FOR AMBULATORY COUGH RECORDING [J].
MUNYARD, P ;
BUSST, C ;
LOGANSINCLAIR, R ;
BUSH, A .
PEDIATRIC PULMONOLOGY, 1994, 18 (03) :178-186