Determination of traces of uranium and thorium in titanium and copper used for the construction of the Russian Emission Detector 100 by inductively coupled plasma mass spectrometry

被引:0
|
作者
Poteshin, Sergey S. [1 ]
Sysoev, Alexey A. [1 ,2 ]
Lagunov, Sergey S. [1 ]
Sereda, Andrei [3 ]
Sosnovtsev, Valery V. [1 ]
Bolozdynya, Alexander I. [1 ]
Efremenko, Yuriy B. [1 ,4 ]
机构
[1] Natl Res Nucl Univ MEPhI, Moscow 115409, Russia
[2] Linantec Ltd, Moscow 115409, Russia
[3] Minist Def Russian Federat, Fed State Inst, Main Fed Ctr Med & Forens Examinat 111, Moscow, Russia
[4] Univ Tennessee, Knoxville, TN USA
关键词
LA-ICP-MS; titanium; thorium; uranium; ULTRATRACE;
D O I
10.1255/ejms.1314
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
The Russian Emission Detector 100 (RED-100) under construction at the National Research Nuclear University MEPhI (Moscow Engineering Physics Institute) is designed to detect the presently undiscovered effect of coherent neutrino scattering. One of the factors limiting the sensitivity of the detector is the spontaneous decay of uranium and thorium in the detector materials. Radioactive impurities in detector materials at levels of parts per billion can significantly affect the sensitivity. Five random samples of titanium and one of copper-from materials used in the construction of the detector were selected for assay. The concentrations of Th-232 and U-238 were measured by inductively coupled plasma mass spectrometry (ICP-MS) in solid titanium using both solutions in acids and direct sampling by laser ablation (LA-ICP-MS). The LA-ICP-MS method allowed us to determine U-238 and Th-232 at subnanogram per gram levels. This method is much faster than ICP-MS with liquid injection. The titanium samples studied have impurities in the range between 1 ng g(-1) and 21 ng g(-1) for U-238 and 3 ng g(-1) and 31 ng g(-1) for Th-232. In copper we set upper limits of 0.4 ng g(-1) for U-238 and 1 ng g(-1) for Th-232. The total activity of the cryostat constructed from the materials studied was estimated to be 43 Bq.
引用
收藏
页码:335 / 340
页数:6
相关论文
共 50 条
  • [41] Determination of copper in coal fly ash in the presence of excess titanium by dynamic reaction cell inductively coupled plasma mass spectrometry
    Hui-Tao Liu
    Shiuh-Jen Jiang
    Analytical and Bioanalytical Chemistry, 2003, 375 : 306 - 309
  • [42] Determination of Uranium Isotope Ratios in Uranium Particles by Laser Ablation Multiple Collector Inductively Coupled Plasma Mass Spectrometry
    Wang Wei
    Li Zhi-Ming
    Xu Jiang
    Zhou Guo-Qing
    Shen Xiao-Pan
    Zhai Li-Hua
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2015, 43 (05) : 703 - 708
  • [43] Determination of impurities in uranium oxide by inductively coupled plasma mass spectrometry (ICPMS) by the matrix matching method
    O. P. Oliveira Junior
    J. E. S. Sarkis
    Journal of Radioanalytical and Nuclear Chemistry, 2002, 254 : 519 - 526
  • [44] Determination of phosphorus and sulfur in uranium ore concentrates by triple quadrupole inductively coupled plasma mass spectrometry
    Fletcher, Nathaniel D.
    Manard, Benjamin T.
    Bostick, Debra A.
    Bostick, William D.
    Metzger, Shalina C.
    Ticknor, Brian W.
    Rogers, Kayron T.
    Hexel, Cole R.
    TALANTA, 2021, 221
  • [45] Determination of uranium in urine - measurement of isotope ratios and quantification by use of inductively coupled plasma mass spectrometry
    Krystek, P
    Ritsema, R
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2002, 374 (02) : 226 - 229
  • [46] APPLICATION OF ISOTOPE-DILUTION FOR THE DETERMINATION OF THORIUM IN BIOLOGICAL SAMPLES BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY
    IGARASHI, Y
    SHIRAISHI, K
    TAKAKU, Y
    MASUDA, K
    SEKI, R
    YAMAMOTO, M
    ANALYTICAL SCIENCES, 1992, 8 (04) : 475 - 479
  • [47] Rapid determination of thorium in urine by quadrupole inductively coupled plasma mass spectrometry (Q-ICP-MS)
    Yongzhong Liu
    Ge Xiao
    Robert L. Jones
    Journal of Radioanalytical and Nuclear Chemistry, 2022, 331 : 3957 - 3964
  • [48] Optimization of the operating conditions using factorial designs for determination of uranium by inductively coupled plasma optical emission spectrometry
    Santos, Juracir S.
    Teixeira, Leonardo S. G.
    Araujo, Rennan G. O.
    Fernandes, Andrea P.
    Korn, Maria Gracas A.
    Ferreira, Sergio L. C.
    MICROCHEMICAL JOURNAL, 2011, 97 (02) : 113 - 117
  • [49] Determination of titanium and vanadium in water samples by inductively coupled plasma mass spectrometry with on-line preconcentration
    Yang, KL
    Jiang, SJ
    Hwang, TJ
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (02) : 139 - 143
  • [50] Determination of Trace Fission Products Associated with Thorium and HCl Matrix by Inductively Coupled Plasma Mass Spectrometry after Anion Exchange Separation
    He, Shuhua
    Chen, Mumei
    Li, Zheng
    Cong, Haixia
    Zhang, Lan
    Li, Qinnuan
    ANALYTICAL SCIENCES, 2017, 33 (11) : 1265 - 1270