Determination of traces of uranium and thorium in titanium and copper used for the construction of the Russian Emission Detector 100 by inductively coupled plasma mass spectrometry

被引:0
|
作者
Poteshin, Sergey S. [1 ]
Sysoev, Alexey A. [1 ,2 ]
Lagunov, Sergey S. [1 ]
Sereda, Andrei [3 ]
Sosnovtsev, Valery V. [1 ]
Bolozdynya, Alexander I. [1 ]
Efremenko, Yuriy B. [1 ,4 ]
机构
[1] Natl Res Nucl Univ MEPhI, Moscow 115409, Russia
[2] Linantec Ltd, Moscow 115409, Russia
[3] Minist Def Russian Federat, Fed State Inst, Main Fed Ctr Med & Forens Examinat 111, Moscow, Russia
[4] Univ Tennessee, Knoxville, TN USA
关键词
LA-ICP-MS; titanium; thorium; uranium; ULTRATRACE;
D O I
10.1255/ejms.1314
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
The Russian Emission Detector 100 (RED-100) under construction at the National Research Nuclear University MEPhI (Moscow Engineering Physics Institute) is designed to detect the presently undiscovered effect of coherent neutrino scattering. One of the factors limiting the sensitivity of the detector is the spontaneous decay of uranium and thorium in the detector materials. Radioactive impurities in detector materials at levels of parts per billion can significantly affect the sensitivity. Five random samples of titanium and one of copper-from materials used in the construction of the detector were selected for assay. The concentrations of Th-232 and U-238 were measured by inductively coupled plasma mass spectrometry (ICP-MS) in solid titanium using both solutions in acids and direct sampling by laser ablation (LA-ICP-MS). The LA-ICP-MS method allowed us to determine U-238 and Th-232 at subnanogram per gram levels. This method is much faster than ICP-MS with liquid injection. The titanium samples studied have impurities in the range between 1 ng g(-1) and 21 ng g(-1) for U-238 and 3 ng g(-1) and 31 ng g(-1) for Th-232. In copper we set upper limits of 0.4 ng g(-1) for U-238 and 1 ng g(-1) for Th-232. The total activity of the cryostat constructed from the materials studied was estimated to be 43 Bq.
引用
收藏
页码:335 / 340
页数:6
相关论文
共 50 条
  • [1] Determination of Uranium and Thorium in Aluminum by Inductively Coupled Plasma Mass Spectrometry
    Shoji Kozuka
    Yuji Yamada
    Miyuki Takenaka
    Masaru Hayashi
    Hideki Matsunaga
    Analytical Sciences, 1997, 13 : 1017 - 1019
  • [2] Determination of uranium and thorium in aluminum by inductively coupled plasma mass spectrometry
    Kozuka, S
    Yamada, Y
    Takenaka, M
    Hayashi, M
    Matsunaga, H
    ANALYTICAL SCIENCES, 1997, 13 (06) : 1017 - 1019
  • [3] Determination of thorium and uranium in urine with inductively coupled argon plasma mass spectrometry
    Ting, BG
    Paschal, DC
    Caldwell, KL
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (05) : 339 - 342
  • [4] Extraction Preconcentration of Uranium and Thorium Traces in the Analysis of Bottom Sediments by Inductively Coupled Plasma Mass Spectrometry
    V. G. Torgov
    M. G. Demidova
    A. I. Saprykin
    I. V. Nikolaeva
    T. V. Us
    E. P. Chebykin
    Journal of Analytical Chemistry, 2002, 57 : 303 - 310
  • [5] Determination of Uranium and Thorium Isotope Ratios by Inductively Coupled Plasma Mass Spectrometry
    M. G. Demidova
    A. I. Saprykin
    Journal of Analytical Chemistry, 2004, 59 : 45 - 49
  • [6] Simultaneous Determination of Thorium and Uranium in Mineral Fertilizers by Inductively Coupled Plasma Optical Emission Spectrometry
    Souza, Sidnei O.
    Avila, Dayara V. L.
    Monteiro, Adnivia S. C.
    Garcia, Carlos Alexandre B.
    Alves, Jose P. H.
    Maranhao, Tatiane A.
    Araujo, Rennan G. O.
    JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 2016, 27 (04) : 799 - 806
  • [7] Determination of traces of uranium and thorium in (Ba, Sr) TiO3 ferroelectrics by inductively coupled plasma mass spectrometry
    M. Fukuda
    Yasumasa Sayama
    Fresenius' Journal of Analytical Chemistry, 1997, 357 : 647 - 651
  • [8] Bioimaging of uranium and thorium in mice organs by laser ablation inductively coupled plasma mass spectrometry
    Li, Xiaoyun
    Wang, Xiaomei
    Zhou, Jing
    Li, Zi-Jian
    Zhang, Linjuan
    Wang, Jian-Qiang
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2023, 332 (07) : 2559 - 2569
  • [9] DETERMINATION OF URANIUM AND THORIUM IN BASALTS AND URANIUM IN AQUEOUS-SOLUTION BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY
    BAILEY, EH
    KEMP, AJ
    RAGNARSDOTTIR, KV
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (04) : 551 - 556
  • [10] Flow-injection technique for determination of uranium and thorium isotopes in urine by inductively coupled plasma mass spectrometry
    Karima Benkhedda
    Vladimir N. Epov
    R. Douglas Evans
    Analytical and Bioanalytical Chemistry, 2005, 381 : 1596 - 1603