The fundamental radiation properties of normoxic polymer gel dosimeters: a comparison between a methacrylic acid based gel and acrylamide based gels

被引:218
作者
De Deene, Y
Vergote, K
Claeys, C
De Wagter, C
机构
[1] Univ Ghent, B-9000 Ghent, Belgium
[2] State Univ Ghent Hosp, MR Dept, B-9000 Ghent, Belgium
关键词
D O I
10.1088/0031-9155/51/3/012
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Polymer gel dosimeters offer a wide range of applications in the three-dimensional verification of complex dose distributions such as in intensity-modulated radiotherapy. One of the major difficulties with polymer gel dosimeters is their sensitivity to oxygen, as oxygen inhibits the radiation-induced polymerization reaction. For several years, oxygen was removed from the gels by bubbling the sol with inert gases for several hours during the gel fabrication. Also, the gel had to be poured in containers with low oxygen permeability and Solubility. Recently, it was found that these technical difficulties can easily be solved by adding an antioxidant to the gel. These gels are called 'normoxic' gels as they can be produced under normal atmospheric conditions. In this study several properties of polymer gel dosimeters have been investigated: the dose sensitivity, the temporal and spatial stability of the gel, the sensitivity of the dose response to temperature during irradiation and during MR imaging, the energy dependence and the dose-rate dependence. This study reveals that the normoxic polymer gel dosimeter based on methacrylic acid (nMAG) studied in this work has inferior radiation properties as compared to the polyacrylamide gelatine (PAG) gel dosimeters. It is shown that from the three different gel dosimeters investigated in this study, the nPAG gel dosimeter results in a less sensitive gel dosimeter but with superior radiation properties as compared to the nMAG gel dosimeter. The importance of investigating relevant radiation properties of gel dosimeters apart from the radiation sensitivity-prior to their use for dosimetric validation experiments-is illustrated and emphasized throughout this study. Other combinations of monomer and gelling agent may result in more reliable normoxic polymer gel dosimeters.
引用
收藏
页码:653 / 673
页数:21
相关论文
共 25 条
[1]  
[Anonymous], 2010, XCOM PHOTON CROSS SE
[2]   VERIFICATION OF WATER EQUIVALENCE OF FEMRI GELS USING MONTE-CARLO SIMULATION [J].
CHAN, MF ;
AYYANGAR, K .
MEDICAL PHYSICS, 1995, 22 (04) :475-478
[3]   Three-dimensional dosimetry using polymer gel and magnetic resonance imaging applied to the verification of conformal radiation therapy in head-and-neck cancer [J].
De Deene, Y ;
De Wagter, C ;
Van Duyse, B ;
Derycke, S ;
De Neve, W ;
Achten, E .
RADIOTHERAPY AND ONCOLOGY, 1998, 48 (03) :283-291
[4]  
De Deene Y, 2002, PHYS MED BIOL, V47, P2459, DOI 10.1088/0031-9155/47/14/307
[5]  
De Deene Y, 2002, PHYS MED BIOL, V47, P3441, DOI 10.1088/0031-9155/47/19/301
[6]  
De Deene Y, 2001, PHYS MED BIOL, V46, P2697, DOI 10.1088/0031-9155/46/10/312
[7]  
De Deene Y, 2000, PHYS MED BIOL, V45, P859, DOI 10.1088/0031-9155/45/4/304
[8]  
De Deene YH, 2001, PHYS MED BIOL, V46, P2801, DOI 10.1088/0031-9155/46/11/304
[9]  
DEDEENE Y, 1996, THESIS U GENT
[10]   Polymer gels for magnetic resonance imaging of radiation dose distributions at normal room atmosphere [J].
Fong, PM ;
Keil, DC ;
Does, MD ;
Gore, JC .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (12) :3105-3113