Optimization of Screw Positioning in Mandible during Bilateral Sagittal Split Osteotomy Using Finite Element Method

被引:0
作者
Najafi, A. Raeisi [1 ]
Pashaei, A. [2 ]
Maid, S. [3 ]
Oskui, I. Zoljanahi [4 ]
Bohluli, B. [5 ]
机构
[1] Univ Illinois, Dept Ind & Enterprise Syst Engn, Urbana, IL 61801 USA
[2] Univ Pompeu Fabra, Dept Technol, CISTIB, Barcelona, Spain
[3] Engn & Sci Res Associates, Olney, MD 20832 USA
[4] Amirkabir Univ Technol, Dept Biomed Engn, Tehran, Iran
[5] Azad Sch Dent, Dept Oral & Maxillofacial Surg, Tehran, Iran
来源
26TH SOUTHERN BIOMEDICAL ENGINEERING CONFERENCE: SBEC 2010 | 2010年 / 32卷
关键词
Dentistry; Bilateral Sagittal Split Osteotomy Finite Element Method and surgery; FIXATION;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Stress analysis of five fixation methods used in Bilateral Sagital Split Osteotomy are performed using finite element method. Geometrical model is built using CT imaging. Mandible is considered isotropic and static stress analysis is performed using linear elasticity. Results are compared to identify the configuration that minimizes the von-Mises Stress values at the entire mode, screws and bone-screw interface to reduce the transferred load to different elements of bone screw model. Results suggest that among the studied configurations Model 3 (Horizontal T), with maximum stress values of 64.59 MPa, 5.29 MPa and 2.67 MPa at the entire model, screws and bone-screw interface, respectively is the optimum positioning configuration to fix the two bone segments.
引用
收藏
页码:516 / +
页数:2