Resolution and noise measurements of five CRT and LCD medical displays

被引:41
作者
Saunders, RS [1 ]
Samei, E
机构
[1] Duke Univ, Ctr Med, Dept Radiol, Duke Adv Imaging Labs, Durham, NC 27710 USA
[2] Duke Univ, Dept Phys, Durham, NC 27710 USA
[3] Duke Univ, Dept Biomed Engn, Durham, NC 27710 USA
关键词
image quality; medical display; modulation transfer function; normalized noise power spectrum; liquid crystal display; cathode ray tube;
D O I
10.1118/1.2150777
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The performance of soft-copy displays plays a significant role in the overall image quality of a digital radiographic system. In this work, we discuss methods to characterize the resolution and noise of both cathode ray tube (CRT) and liquid crystal display (LCD) devices. We measured the image quality of five different commercial display devices, representing both CRT and LCD technologies, using a high-quality charge-coupled device (CCD) camera. The modulation transfer function (MTF) was calculated using the line technique, correcting for the MTF of the CCD camera and the display pixel size. The normalized noise power spectrum (NPS) was computed from two-dimensional Fourier analysis of uniform images. To separate the effects of pixel structure from interpixel luminance variations, we created structure-free images by eliminating the pixel structures of the display device. The NPS was then computed from these structure-free images to isolate interpixel luminance variations. We found that the MTF of LCDs remained close to the theoretical limit dictated by their inherent pixel size (0.85 +/- 0.08 at Nyquist frequency), in contrast to the MTF for the two CRT displays, which dropped to 0.15 +/- 0.08 at the Nyquist frequency. However, the NPS of LCDs showed significant peaks due to the subpixel structure, while the NPS of CRT displays exhibited a nearly flat power spectrum. After removing the pixel structure, the structured noise peaks for LCDs were eliminated and the overall noise magnitude was significantly reduced. The average total noise-to-signal ratio for CRT displays was 6.55% +/- 0.59%, of which 6.03% +/- 0.24% was due to interpixel luminance variations, while LCD displays had total noise to signal ratios of 46.1% +/- 5.1% of which 1.50% +/- 0.41% were due to interpixel luminance variations. Depending on the extent of the blurring and prewhitening processes of the human visual system, the magnitude of the display noise (including pixel structure) potentially perceived by the observer was reduced to 0.43% +/- 0.01% (accounting for blurring only) and 0.40 +/- 0.01% (accounting for blurring and prewhitening) for CRTs, and 1.02% +/- 0.22% (accounting for blurring only) and 0.36% +/- 0.08% (accounting for blurring and prewhitening) for LCDs. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:308 / 319
页数:12
相关论文
共 25 条
[1]   Noise in flat-panel displays with subpixel structure [J].
Badano, A ;
Gagne, RM ;
Jennings, RJ ;
Drilling, SE ;
Imhoff, BR ;
Muka, E .
MEDICAL PHYSICS, 2004, 31 (04) :715-723
[2]   Luminance effects on display resolution and noise [J].
Badano, A ;
Hipper, SJ ;
Jennings, RJ .
MEDICAL IMAGING 2002: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2002, 4681 :305-313
[3]   Characterization of high-resolution liquid-crystal displays for medical images [J].
Blume, H ;
Steven, PM ;
Cobb, M ;
Ho, AM ;
Stevens, F ;
Muller, S ;
Roehrig, H ;
Fan, JH .
MEDICAL IMAGING 2002: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2002, 4681 :271-292
[4]   Use of a slit camera for MTF measurements [J].
Bradford, CD ;
Peppler, WW ;
Waidelich, JM .
MEDICAL PHYSICS, 1999, 26 (11) :2286-2294
[5]   Factors affecting CRT display performance: specifying what works [J].
Compton, KD .
MEDICAL IMAGING 2000: IMAGE DISPLAY AND VISUALIZATION, 2000, 3976 :412-423
[6]   Automated computer evaluation and optimization of image compression of x-ray coronary angiograms for signal known exactly detection tasks [J].
Eckstein, MP ;
Bartroff, JL ;
Abbey, CK ;
Whiting, JS ;
Bochud, FO .
OPTICS EXPRESS, 2003, 11 (05) :460-475
[7]   Spatial noise of high resolution liquid crystal displays for medical imaging: quantitative analysis, estimation and compensation [J].
Fan, JH ;
Dallas, WJ ;
Roehrig, H ;
Krupinski, E ;
Gandhi, K ;
Sundareshan, MK .
MEDICAL IMAGING 2004: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2004, 5367 :433-443
[8]   Improving visualization of digital mammograms on a CRT display system [J].
Fan, JH ;
Dallas, WJ ;
Roehrig, H ;
Krupinski, E .
MEDICAL IMAGING 2003: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2003, 5029 :746-753
[9]   Experimental comparison of noise and resolution for 2k and 4k storage phosphor radiography systems [J].
Flynn, MJ ;
Samei, E .
MEDICAL PHYSICS, 1999, 26 (08) :1612-1623
[10]   INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .6. MTFS OF II-TV DIGITAL IMAGING-SYSTEMS [J].
FUJITA, H ;
DOI, K ;
GIGER, ML .
MEDICAL PHYSICS, 1985, 12 (06) :713-720