Cloud point extraction as a preconcentration step prior to capillary electrophoresis

被引:92
作者
Carabias-Martínez, R [1 ]
Rodríguez-Gonzalo, E [1 ]
Domínguez-Alvarez, J [1 ]
Hernández-Méndez, J [1 ]
机构
[1] Univ Salamanca, Fac Quim, Dept Quim Analit Nutric & Bromatol, E-37008 Salamanca, Spain
关键词
D O I
10.1021/ac9812227
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Cloud point extraction was applied as a preconcentration step prior to capillary electrophoresis. The behavior of a surfactant-rich micellar phase injected into a capillary electrophoresis system was studied using different separation modes: micellar electrokinetic capillary chromatography and capillary zone electrophoresis (CZE). A problem that appeared on introducing a surfactant-rich phase into a bare fused silica capillary was that the surfactant was adsorbed onto the wall of the capillary, leading to a marked loss of efficiency and reproducibility both in the migration times and in the areas of the electrophoretic peaks. The use of cetyltrimethylammonium bromide dynamically coated capillaries afforded reproducible results, although the half-life of the capillary was short. The most satisfactory results were obtained by using nonaqueous media in the CZE mode, thus avoiding surfactant adsorption. Other parameters related to the composition of the injection medium were also studied to optimize the electrophoretic behavior of the analytes and the sensitivity of the determination. The optimized procedure was applied to the determination of triazines in tap and river Rater samples.
引用
收藏
页码:2468 / 2474
页数:7
相关论文
共 27 条
[1]   ELECTRICALLY DRIVEN OPEN-TUBULAR LIQUID-CHROMATOGRAPHY [J].
BRUIN, GJM ;
TOCK, PPH ;
KRAAK, JC ;
POPPE, H .
JOURNAL OF CHROMATOGRAPHY, 1990, 517 :557-572
[2]   ONLINE PRECONCENTRATION OF TRIAZINE HERBICIDES WITH TANDEM OCTADECYL CAPILLARIES-CAPILLARY ZONE ELECTROPHORESIS [J].
CAI, J ;
ELRASSI, Z .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1992, 15 (6-7) :1179-1192
[3]   MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY OF NEUTRAL SOLUTES WITH MICELLES OF ADJUSTABLE SURFACE-CHARGE DENSITY [J].
CAI, J ;
ELRASSI, Z .
JOURNAL OF CHROMATOGRAPHY, 1992, 608 (1-2) :31-45
[4]   CAPILLARY ZONE ELECTROPHORESIS OF 2 CATIONIC HERBICIDES, PARAQUAT AND DIQUAT [J].
CAI, JY ;
ELRASSI, Z .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1992, 15 (6-7) :1193-1200
[5]   Capillary zone electrophoresis in nonaqueous solvents in the presence of ionic additives [J].
CarabiasMartinez, R ;
RodriguezGonzalo, E ;
DominguezAlvarez, J ;
HernandezMendez, J .
ANALYTICAL CHEMISTRY, 1997, 69 (21) :4437-4444
[6]   CLOUD POINT METHODOLOGY - A NEW APPROACH FOR PRECONCENTRATION AND SEPARATION IN HYDRODYNAMIC SYSTEMS OF ANALYSIS [J].
CORDERO, BM ;
PAVON, JLP ;
PINTO, CG ;
LAESPADA, EF .
TALANTA, 1993, 40 (11) :1703-1710
[7]   ATRAZINE AND SIMAZINE DETERMINATION IN RIVER WATER SAMPLES BY MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY [J].
DESIDERIO, C ;
FANALI, S .
ELECTROPHORESIS, 1992, 13 (9-10) :698-700
[8]   USE OF CAPILLARY ELECTROPHORESIS FOR DETECTION OF METSULFURON AND CHLORSULFURON IN TAP WATER [J].
DINELLI, G ;
VICARI, A ;
CATIZONE, P .
JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 1993, 41 (05) :742-746
[9]   SEPARATION AND DETECTION OF HERBICIDES IN WATER BY MICELLAR ELECTROKINETIC CAPILLARY CHROMATOGRAPHY [J].
DINELLI, G ;
BONETTI, A ;
CATIZONE, P ;
GALLETTI, GC .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1994, 656 (01) :275-280
[10]  
DOMINGUEZALVARE.J, COMMUNICATION