Imaging the Parasinus Region with a Third-Generation Dual-Source CT and the Effect of Tin Filtration on Image Quality and Radiation Dose

被引:53
作者
Lell, M. M. [1 ,4 ]
May, M. S. [1 ]
Brand, M. [1 ]
Eller, A. [1 ]
Buder, T. [3 ]
Hofmann, E. [2 ]
Uder, M. [1 ,4 ]
Wuest, W. [1 ]
机构
[1] Univ Erlangen Nurnberg, Dept Radiol, D-91054 Erlangen, Germany
[2] Univ Erlangen Nurnberg, Dept Orthodont & Orofacial Orthoped, D-91054 Erlangen, Germany
[3] Univ Erlangen Nurnberg, Dept 1, Inst Anat, D-91054 Erlangen, Germany
[4] Univ Erlangen Nurnberg, Imaging Sci Inst, D-91054 Erlangen, Germany
关键词
BEAM COMPUTED-TOMOGRAPHY; PARANASAL SINUS; CONE-BEAM; MULTIDETECTOR CT; SELECTION; EXPOSURE; DISEASE;
D O I
10.3174/ajnr.A4270
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: CT is the imaging technique of choice in the evaluation of midface trauma or inflammatory disease. We performed a systematic evaluation of scan protocols to optimize image quality and radiation exposure on third-generation dual-source CT. MATERIALS AND METHODS: CT protocols with different tube voltage (70-150 kV), current (25-300 reference mAs), prefiltration, pitch value, and rotation time were systematically evaluated. All images were reconstructed with iterative reconstruction (Advanced Modeled Iterative Reconstruction, level 2). To individually compare results with otherwise identical factors, we obtained all scans on a frozen human head. Conebeam CT was performed for image quality and dose comparison with multidetector row CT. Delineation of important anatomic structures and incidental pathologic conditions in the cadaver head was evaluated. RESULTS: One hundred kilovolts with tin prefiltration demonstrated the best compromise between dose and image quality. The most dose-effective combination for trauma imaging was Sn100 kV/250 mAs (volume CT dose index, 2.02 mGy), and for preoperative sinus surgery planning, Sn100 kV/150 mAs (volume CT dose index, 1.22 mGy). "Sn" indicates an additional prefiltration of the x-ray beam with a tin filter to constrict the energy spectrum. Exclusion of sinonasal disease was possible with even a lower dose by using Sn100 kV/25 mAs (volume CT dose index, 0.2 mGy). CONCLUSIONS: High image quality at very low dose levels can be achieved by using a Sn100-kV protocol with iterative reconstruction. The effective dose is comparable with that of conventional radiography, and the high image quality at even lower radiation exposure favors multidetector row CT over conebeam CT.
引用
收藏
页码:1225 / 1230
页数:6
相关论文
共 27 条
[1]   Reliability of High-Pitch Ultra-Low-Dose Paranasal Sinus Computed Tomography for Evaluating Paranasal Sinus Anatomy and Sinus Disease [J].
Aksoy, Elif Ayanoglu ;
Ozden, Sila Ulus ;
Karaarslan, Ercan ;
Unal, Omer Faruk ;
Tanyeri, Hasan .
JOURNAL OF CRANIOFACIAL SURGERY, 2014, 25 (05) :1801-1804
[2]   Multidetector CT of the paranasal sinus: Potential for radiation dose reduction [J].
Brem, Matthias H. ;
Zamani, Amir A. ;
Riva, Roberto ;
Zou, Kelly H. ;
Rumboldt, Zoran ;
Hennig, Friedrich F. ;
Kikinis, Ron ;
Norbash, Alexander M. ;
Schoepf, U. Joseph .
RADIOLOGY, 2007, 243 (03) :847-852
[3]   Letter to the editor re: low-dose computed tomography of the paranasal sinus and facial skull using a high-pitch dual-source system-First clinical results [J].
Bulla, Stefan ;
Blanke, Philipp ;
Langer, Mathias ;
Pache, Gregor .
EUROPEAN RADIOLOGY, 2011, 21 (07) :1447-1448
[4]   An automated robotic approach with redundant navigation for minimal invasive extended transsphenoidal skull base surgery [J].
Bumm, K ;
Wurm, J ;
Rachinger, J ;
Dannenmann, T ;
Bohr, C ;
Fahlbusch, R ;
Iro, H ;
Nimsky, C .
MINIMALLY INVASIVE NEUROSURGERY, 2005, 48 (03) :159-164
[5]   Current knowledge on tumour induction by computed tomography should be carefully used [J].
Candela-Juan, Cristian ;
Montoro, Alegria ;
Ruiz-Martinez, Enrique ;
Ignacio Villaescusa, Juan ;
Marti-Bonmati, Luis .
EUROPEAN RADIOLOGY, 2014, 24 (03) :649-656
[6]   Multisection CT Protocols: Sex- and Age-specific Conversion Factors Used to Determine Effective Dose from Dose-Length Product [J].
Deak, Paul D. ;
Smal, Yulia ;
Kalender, Willi A. .
RADIOLOGY, 2010, 257 (01) :158-166
[7]   LOW-DOSE CT AND INFLAMMATORY DISEASE OF THE PARANASAL SINUSES [J].
DUVOISIN, B ;
LANDRY, M ;
CHAPUIS, L ;
KRAYENBUHL, M ;
SCHNYDER, P .
NEURORADIOLOGY, 1991, 33 (05) :403-406
[8]   Carotid CTA: Radiation Exposure and Image Quality with the Use of Attenuation-Based, Automated Kilovolt Selection [J].
Eller, A. ;
Wuest, W. ;
Kramer, M. ;
May, M. ;
Schmid, A. ;
Uder, M. ;
Lell, M. M. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (02) :237-241
[9]   Attenuation-based automatic kilovolt (kV)-selection in computed tomography of the chest: Effects on radiation exposure and image quality [J].
Eller, Achim ;
Wuest, Wolfgang ;
Scharf, Michael ;
Brand, Michael ;
Achenbach, Stephan ;
Uder, Michael ;
Lell, Michael M. .
EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (12) :2386-2391
[10]   Attenuation-Based Automatic Kilovolt Selection in Abdominal Computed Tomography Effects on Radiation Exposure and Image Quality [J].
Eller, Achim ;
May, Matthias S. ;
Scharf, Michael ;
Schmid, Axel ;
Kuefner, Michael ;
Uder, Michael ;
Lell, Michael M. .
INVESTIGATIVE RADIOLOGY, 2012, 47 (10) :559-565