A pixel chamber to monitor the beam performances in hadron therapy

被引:15
作者
Bonin, R
Boriano, A
Bourhaleb, F
Cirio, R
Donetti, M
Garelli, E
Giordanengo, S
Marchetto, F
Peroni, C
Freire, CJS
Simonetti, L
机构
[1] Univ Turin, Dipartimento Fis Sperimentale, I-10125 Turin, Italy
[2] Ist Nazl Fis Nucl, I-10125 Turin, Italy
[3] Fdn TERA, I-28100 Novara, Italy
[4] ASP, I-10125 Turin, Italy
关键词
radiotherapy; beam monitor; ionization chamber; dosimetry; carbon ion;
D O I
10.1016/j.nima.2003.11.104
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this paper we describe the design, construction, and tests of a parallel plate ionization chamber with the anode segmented in (32 x 32) square pixels. The performance of the read out and data acquisition systems is also discussed. The design of the chamber has been finalized to be used as a beam monitor for therapeutical treatments. Position and flux resolution obtained with a carbon ion beam are presented. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:674 / 686
页数:13
相关论文
共 8 条