Coherence scanning and phase imaging optical interference microscopy at the lateral resolution limit

被引:36
作者
Lehmann, Peter [1 ]
Xie, Weichang [1 ]
Allendorf, Benedikt [1 ]
Tereschenko, Stanislav [1 ]
机构
[1] Univ Kassel, Dept Elect Engn & Comp Sci, Measurement Technol, Wilhelmshoher Allee 71, D-34121 Kassel, Germany
关键词
NUMERICAL APERTURE; CALIBRATION; FIELD; MIRAU; SYSTEMS;
D O I
10.1364/OE.26.007376
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
To get physical insight into the 3D transfer characteristics of interference microscopy at high numerical apertures we study reflecting rectangular grating structures. In general, the height obtained from phase information seems to be reduced, whereas height values resulting from coherence scanning sometimes seem to be systematically overestimated. Increasing the numerical aperture of an interference microscope broadens the spectra of the resulting interference signals, thus offering a broad variety of wavelength contributions to be analyzed. If phase analysis of a measured far-field interference wavefront is performed at very short wavelengths the periodical profiles obtained from coherence scanning and phase shifting analysis differ only by the measured amplitude. However, at longer wavelength there is a 180 degrees phase shift of the measured profiles obtained from phase analysis compared to coherence peak analysis. Increasing the evaluation wavelength improves the lateral resolution since the long wavelength contributions are related to electromagnetic waves of high angles of incidence. This behavior is to the best of our knowledge not documented in literature so far. It was first observed experimentally and could be confirmed by simulation results obtained from either Kirchhoff diffraction theory or an extended Richards-Wolf model developed in our group. Compared to original input profiles used for the simulation the profiles obtained from phase evaluation correspond quite well at longer wavelength, whereas the results obtained from coherence peak analysis are typically inverted with respect to height. (C) 2018 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
引用
收藏
页码:7376 / 7389
页数:14
相关论文
共 40 条
[1]   Microsphere-assisted super-resolved Mirau digital holographic microscopy for cell identification [J].
Aakhte, Mostafa ;
Abbasian, Vahid ;
Akhlaghi, Ehsan Ahadi ;
Moradi, Ali-Reza ;
Anand, Arun ;
Javidi, Bahram .
APPLIED OPTICS, 2017, 56 (09) :D8-D13
[2]  
Abbe E., 1873, ARCH MIKROSK ANAT, V9, P413, DOI [DOI 10.1007/BF02956173, 10.1007/BF02956173]
[3]   Spatial and temporal coherence effects in interference microscopy and full-field optical coherence tomography [J].
Abdulhalim, Ibrahim .
ANNALEN DER PHYSIK, 2012, 524 (12) :787-804
[4]   ON THE LIMITATIONS OF THE CONFOCAL SCANNING OPTICAL MICROSCOPE AS A PROFILOMETER [J].
AGUILAR, JF ;
MENDEZ, ER .
JOURNAL OF MODERN OPTICS, 1995, 42 (09) :1785-1794
[5]  
[Anonymous], 2013, 25178604 ISO
[6]  
Beckmann P., 1987, Scattering of Electromagnetic Waves from Rough Surfaces
[7]   CALIBRATION REQUIREMENTS FOR MIRAU AND LINNIK MICROSCOPE INTERFEROMETERS [J].
BIEGEN, JF .
APPLIED OPTICS, 1989, 28 (11) :1972-1974
[8]   Isotropic image in structured illumination microscopy patterned with a spatial light modulator [J].
Chang, Bo-Jui ;
Chou, Li-Jun ;
Chang, Yun-Ching ;
Chiang, Su-Yu .
OPTICS EXPRESS, 2009, 17 (17) :14710-14721
[9]  
Corle T.R., 1996, Confocal scanning optical microscopy and related imaging systems
[10]   CALIBRATION OF NUMERICAL APERTURE EFFECTS IN INTERFEROMETRIC MICROSCOPE OBJECTIVES [J].
CREATH, K .
APPLIED OPTICS, 1989, 28 (16) :3333-3338