Sensitivity analysis of physics and planning SmartArc parameters for single and partial arc VMAT planning

被引:15
作者
Yang, Kai [1 ]
Yan, Di [1 ]
Tyagi, Neelam [1 ]
机构
[1] William Beaumont Hosp, Dept Radiat Oncol, Royal Oak, MI 48073 USA
关键词
SmartArc; leaf speed; control points; VMAT; VOLUMETRIC MODULATED ARC; DELIVERY; RAPIDARC; THERAPY; VERIFICATION; QUALITY; SYSTEM; IMRT;
D O I
10.1120/jacmp.v13i6.3760
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We investigate the sensitivity of various physics and planning SmartArc parameters to generate single and partial arc VMAT plans with equivalent or better plan quality as IMRT. Patients previously treated with step-and-shoot IMRT for several treatment sites were replanned using SmartArc. These treatment sites included head and neck, prostate, lung, and spine. Effect of various physics and planning SmartArc parameters, such as continuous vs. binned dose rate, dynamic leaf gap, leaf speed, maximum delivery time, number of arcs, and control point spacing, were investigated for Elekta Axesse and Synergy linacs. Absolute dose distribution was measured by using the ArcCHECK 3D cylindrical diode array. For all cases investigated, plan metrics such as conformity indices and dose homogeneity indices increased, while plan QA decreased with increasing leaf speed. Leaf speed had a significant impact on the segment size for low dose per fractionation cases. Constraining leaf motion to a lower speed not only avoids tiny large leaf travel and low-dose rate value, but also achieves better PTV coverage (defined as the volume receiving prescription dose) with less total MUs. Maximum delivery time, the number of arcs, and the spacing of control points all had similar effects as the leaf motion constraint on dose rate and segment size. The maximum delivery time had a significant effect on the optimization, acting as a hard constraint. Increasing the control point spacing from 2 to 6 degrees increased the PTV coverage, but reduced the absolute dose gamma passing rate. Plans generated using continuous and binned dose rate modes did not show any difference in the quality and the delivery for the Elekta machines. Dosimetric analysis with a 3D cylindrical QA phantom resulted in 93.6%-99.3% of detectors with a gamma index (3%/2 mm) < 1 for all cases.
引用
收藏
页码:34 / 45
页数:12
相关论文
共 12 条
[1]   COMMISSIONING OF VOLUMETRIC MODULATED ARC THERAPY (VMAT) [J].
Bedford, James L. ;
Warrington, Alan P. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (02) :537-545
[2]   Development and evaluation of an efficient approach to volumetric arc therapy planning [J].
Bzdusek, Karl ;
Friberger, Henrik ;
Eriksson, Kjell ;
Hardemark, Bjorn ;
Robinson, David ;
Kaus, Michael .
MEDICAL PHYSICS, 2009, 36 (06) :2328-2339
[3]   Evaluation of a new VMAT QA device, or the "X" and "O" array geometries [J].
Feygelman, Vladimir ;
Zhang, Geoffrey ;
Stevens, Craig ;
Nelms, Benjamin E. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02) :146-168
[4]   Initial dosimetric evaluation of SmartArc - a novel VMAT treatment planning module implemented in a multi-vendor delivery chain [J].
Feygelman, Vladimir ;
Zhang, Geoffrey ;
Stevens, Craig .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2010, 11 (01) :99-116
[5]   Measurement Comparison and Monte Carlo Simulations for RapidArc Delivery Verification using ArcCheck 4D Dosimetry System [J].
Koren, S. ;
Lin, M. ;
Fan, J. ;
Wang, L. ;
Cherian, G. ;
Price, R. A. ;
Ma, C. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03) :S737-S738
[6]   Dosimetric verification of RapidArc treatment delivery [J].
Korreman, Stine ;
Medin, Joakim ;
Kjaer-Kristoffersen, Flemming .
ACTA ONCOLOGICA, 2009, 48 (02) :185-191
[7]   Commissioning and quality assurance of RapidArc radiotherapy delivery system [J].
Ling, C. Clifton ;
Zhang, Pengpeng ;
Archambault, Yves ;
Bocanek, Jim ;
Tang, Grace ;
LoSasso, Thomas .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (02) :575-581
[8]   A technique for the quantitative evaluation of dose distributions [J].
Low, DA ;
Harms, WB ;
Mutic, S ;
Purdy, JA .
MEDICAL PHYSICS, 1998, 25 (05) :656-661
[9]   CLINICAL APPLICATIONS OF VOLUMETRIC MODULATED ARC THERAPY [J].
Matuszak, Martha M. ;
Yan, Di ;
Grills, Inga ;
Martinez, Alvaro .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 77 (02) :608-616
[10]   Comparing planning time, delivery time and plan quality for IMRT, RapidArc and tomotherapy [J].
Oliver, Mike ;
Ansbacher, Will ;
Beckham, Wayne A. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (04) :117-131