New Functionalities of SINDBAD Simulation Software for Spectral X-Ray Imaging Using Counting Detectors with Energy Discrimination

被引:0
作者
Rebuffel, Veronique [1 ]
Tabary, Joachim [1 ]
Hugonnard, Patrick [1 ]
Popa, Emil [1 ]
Brambilla, Andrea [1 ]
Montemont, Guillaume [1 ]
Verger, Loick [1 ]
机构
[1] CEA Leti, F-38054 Grenoble, France
来源
2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC) | 2012年
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The computer simulation of realistic radiographs has been proved to be an efficient tool for the design of X-ray systems, the specification of their components, the development of dedicated image processing methods and the final validation. The radiographic simulation software Sindbad has been developed in our laboratory for that purpose. It allows the modeling of any radiographic set-up, any object to be examined, and takes into account the various X-ray/matter interactions. Until now, Sindbad proposes various models of detectors; but all of them provide a single value, corresponding to the deposited energy integrated over the energy range. Recently emerged semiconductor based X-ray detectors (CdTe or CZT) offer new capabilities in counting mode with energy discrimination. They can potentially decrease image noise at constant dose, enable material identification or accurate quantification. The purpose of this study is to describe recent developments of Sindbad allowing taking into account such energy sensitive detectors. These ones have been modeled, considering both physical and electronics phenomena. The integration into Sindbad offers spectral imaging simulation functionalities in both radiography and tomographic mode. Examples highlight the energetic dependency of contrast. Validation has been performed, using for comparison semi-conductor detectors developed in our laboratory. An example of quantification of beam-hardening effect illustrates the influence of detector characteristics on reconstructed image quality.
引用
收藏
页码:2550 / 2554
页数:5
相关论文
共 13 条
  • [1] Simulation of charge collection processes in semiconductor CdZnTe γ-ray detectors
    Benoit, Mathieu
    Hamel, L. A.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 606 (03) : 508 - 516
  • [2] CdTe Linear Pixel X-Ray Detector With Enhanced Spectrometric Performance for High Flux X-Ray Imaging
    Brambilla, A.
    Ouvrier-Buffet, P.
    Rinkel, J.
    Gonon, G.
    Boudou, C.
    Verger, L.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (04) : 1552 - 1558
  • [3] Photon Counting Energy Dispersive Detector Arrays for X-ray Imaging
    Iwanczyk, Jan S.
    Nygard, Einar
    Meirav, Oded
    Arenson, Jerry
    Barber, William C.
    Hartsough, Neal E.
    Malakhov, Nail
    Wessel, Jan C.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) : 535 - 542
  • [4] Kruger H., 2008, P SPIE PROGR BIOMEDI, V6913
  • [5] Simulation and design of orthogonal capacitive strip CdZnTe detectors
    Montemont, G.
    Gentet, M.-C.
    Monnet, O.
    Rustique, J.
    Verger, L.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2007, 54 (04) : 854 - 859
  • [6] Ramo S., 1939, PROC IRE, V27, P584
  • [7] Roessl E., 2011, IEEE NUCL SCI S
  • [8] Realistic CT simulation using the 4D XCAT phantom
    Segars, W. P.
    Mahesh, M.
    Beck, T. J.
    Frey, E. C.
    Tsui, B. M. W.
    [J]. MEDICAL PHYSICS, 2008, 35 (08) : 3800 - 3808
  • [9] An algorithm for Monte Carlo simulation of coupled electron-photon transport
    Sempau, J
    Acosta, E
    Baro, J
    FernandezVarea, JM
    Salvat, F
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1997, 132 (03) : 377 - 390
  • [10] Sourbelle K., 2007, THORAX PHANTOM