Evaluation of total corneal power measurements with a new optical biometer

被引:37
作者
Shajari, Mehdi [1 ,2 ]
Sonntag, Ruven [1 ]
Ramsauer, Michaela [1 ]
Kreutzer, Thomas [1 ]
Vounotrypidis, Efstathios [1 ]
Kohnen, Thomas [2 ]
Priglinger, Siegfried [1 ]
Mayer, Wolfgang J. [1 ]
机构
[1] Ludwig Maximilians Univ Munchen, Dept Ophthalmol, Mathildenstr 8, D-80336 Munich, Germany
[2] Goethe Univ, Dept Ophthalmol, Frankfurt, Germany
关键词
REPEATABILITY; ASTIGMATISM; AGREEMENT; KERATOMETRY; DEVICES;
D O I
10.1097/j.jcrs.0000000000000136
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate the repeatability and validity of total corneal power measurements (total keratometry [TK]) obtained with a recently introduced optical biometer (IOLMaster 700, Carl Zeiss Meditec AG) to a Scheimpflug device (Pentacam, Oculus). Setting: Department of Ophthalmology, Ludwig-MaximiliansUniversity, Munich, Germany. Design: Prospective randomized controlled trial. Methods: The inclusion criteria were a corneal astigmatism of 3.00 diopters (D) or less, no previous ocular surgery, no known corneal irregularities, and no known dry eyes. All eyes were measured 3 times using the optical biometer and the Scheimpflug device. The results were statistically compared using Bland-Altman, within- subject SD, and astigmatism vector analysis. Results: Ninety-three eyes of 93 subjects were included. Overall, the repeatability of all eyes measured was 0.42 for standard K, 0.40 for TK, 0.45 for total corneal refractive power (TCRP), 0.43 for true net power (TNP), and 0.39 for simulated K. Bland-Altman analysis showed no significant difference between the optical biometer's TK compared with standard K and the Scheimpflug device's simulated K, TNP, and TCRP, the P value exceeding.05 in each case. A difference of >0.50 diopters between astigmatism measurements between TK and TCRP was found in 10 cases compared with standard K and TNP with 0 and 2 cases. Conclusions: Repeatability of both devices was high, although measurements of meridian showed a great variability, suggesting that numerous measurements are needed to enhance accuracy. TCRP measurements between recently introduced TK and TCRP cannot be used interchangeably. Future studies are necessary to evaluate which measurement will result in a better outcome when respective measurements are used for toric intraocular lens calculations. Copyright (C) 2020 Published by Wolters Kluwer on behalf of ASCRS and ESCRS
引用
收藏
页码:675 / 681
页数:7
相关论文
共 20 条
  • [1] Pursuing perfection in intraocular lens calculations IV. Rethinking astigmatism analysis for intraocular lens-based surgery: Suggested terminology, analysis, and standards for outcome reports
    Abulafia, Adi
    Koch, Douglas D.
    Holladay, Jack T.
    Wang, Li
    Hill, Warren
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2018, 44 (10) : 1169 - 1174
  • [2] Accuracy of the Barrett True-K formula for intraocular lens power prediction after laser in situ keratomileusis or photorefractive keratectomy for myopia
    Abulafia, Adi
    Hill, Warren E.
    Koch, Douglas D.
    Wang, Li
    Barrett, Graham D.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (03) : 363 - 369
  • [3] Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500
    Akman, Ahmet
    Asena, Leyla
    Gungor, Sirel Gur
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2016, 100 (09) : 1201 - 1205
  • [4] [Anonymous], AST DOUBL ANGL PLOT
  • [5] Bland JM, 1996, BRIT MED J, V313, P744
  • [6] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [7] Keratometry versus total corneal refractive power: Analysis of measurement repeatability with 5 different devices in normal eyes with low astigmatism
    Fityo, Sylvia
    Buehren, Jens
    Shajari, Mehdi
    Kohnen, Thomas
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (04) : 569 - 576
  • [8] Haigis W, 2014, ANN M 32 C ESCRS LON
  • [9] Herrmann C, 2008, OPHTHALMOLOGE, V105, P193, DOI 10.1007/s00347-008-1696-0
  • [10] Enhanced Penetration for Axial Length Measurement of Eyes with Dense Cataracts Using Swept Source Optical Coherence Tomography: A Consecutive Observational Study
    Hirnschall, Nino
    Varsits, Ralph
    Doeller, Birgit
    Findl, Oliver
    [J]. OPHTHALMOLOGY AND THERAPY, 2018, 7 (01) : 119 - 124