Monte Carlo study of the scattered radiation field near the eyes of the operator in interventional procedures

被引:21
|
作者
Ferrari, Paolo [1 ]
Becker, Frank [2 ]
Carinou, Eleftheria [3 ]
Chumak, Vadim [4 ]
Farah, Jad [5 ]
Jovanovic, Zoran [6 ]
Krstic, Dragana [6 ]
Morgun, Artem [4 ]
Principi, Sara [7 ]
Teles, Pedro [8 ]
机构
[1] ENEA Radiat Protect Inst, 4 Via Martiri di Monte Sole, I-40129 Bologna, BO, Italy
[2] Karlsruhe Inst Technol, Inst Nucl Waste Disposal, Hermann von Helmholtz Pl 1, D-76344 Eggenstein Leopoldshafen, Germany
[3] GAEC Greek Atom Energy Commiss, Attiki 15310, Greece
[4] NAMS Ukraine, Natl Res Ctr Radiat Med, Melnikova 53, UA-04050 Kiev, Ukraine
[5] IRSN, PRP HOM, SDE, LDRI, BP 17, F-92262 Fontenay Aux Roses, France
[6] Univ Kragujevac, Fac Sci, 12 R Domanovic, Kragujevac 34000, Serbia
[7] Univ Politecn Cataluna, Inst Tecn Energet, Diagonal 647, E-08028 Barcelona, Spain
[8] Ctr Ciencias & Tecnol Nucl C2TN, Grp Protecao & Seguranca Radiol, CTN IST, Polo De Loures, Bobadela, Portugal
关键词
eye lens; interventional radiology; interventional cardiology; lead glasses; shielding; Monte Carlo; medical staff dose monitoring; LENS DOSIMETRY; CATARACT RISK; EXPOSURE; RADIOLOGY; PROTECTION; STAFF; CALIBRATION; QUANTITIES; PATIENT; EYEWEAR;
D O I
10.1088/0952-4746/36/4/902
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Interventional radiology and cardiology guarantee high benefits for patients, but are known to be associated with a high level of radiation exposure of medical staff. The recently suggested decrease of the annual dose limit for the eye lens, from 150 to 20 mSv, caused a need for a reconsideration of practices ensuring sufficient protection for the lens of the eyes of medical staff. In such context the study of the scattered radiation around the operator's head could help in finding the best solutions to be adopted for the ceiling-suspended shield and lead glasses in the most common situations in interventional practices. MCNPX Monte Carlo code was employed with anthropomorphic mathematical phantoms to simulate interventional practice projections. For each projection the effect of changing selected parameters on the evaluated scattered radiation towards the operator's head has been calculated. The variety of modelled situations provides plentiful material regarding the spatial distribution of the scattered radiation, useful to improve eye lens radiation protection, such as the following: (a) Glasses, which provide shielding from both lateral and bottom-up scattered radiation, can reduce by ten times the exposure to the most exposed eyes; (b) The ceiling-suspended shield offers valuable protection, but such effectiveness can diminish by 90% if the shielding is not correctly positioned; (c) The transition from femoral to radial access usually intensifies the scattered radiation toward the operator head (a factor of 1.5 for AP projection), but for RAO projections, a reduction of the order by two to three times, in the case of radial access, can be seen, due to the protection provided by the image receptor. The detailed fluence outcomes show that there is a preferential direction of the impinging scattered radiation that should be considered when radiation protection options are evaluated or when a dedicated eye lens dosemeter is used for monitoring.
引用
收藏
页码:902 / 921
页数:20
相关论文
共 50 条
  • [21] Occupational exposures during abdominal fluoroscopically guided interventional procedures for different patient sizes - A Monte Carlo approach
    Santos, William S.
    Belinato, Walmir
    Perini, Ana P.
    Caldas, Linda V. E.
    Galeano, Diego C.
    Santos, Carla J.
    Neves, Lucio P.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 45 : 35 - 43
  • [22] Monte Carlo study of scattered and random coincidences for MADPET-4
    Negar Omidvari
    Jorge Cabello
    Florian Schneider
    Stephan Paul
    Sibyle I Ziegler
    EJNMMI Physics, 2 (Suppl 1)
  • [23] Monte Carlo calculation of scattered radiation from applicators in low energy clinical electron beams
    Jabbari, Nasrollah
    Hashemi-Malayeri, Bijan
    Farajollahi, Ali Reza
    Kazemnejad, Anoshiravan
    NUKLEONIKA, 2007, 52 (03) : 97 - 103
  • [24] Algorithm for forcing scattered radiation to arbitrary convex regions in neutral particle Monte Carlo simulation
    Tickner, J. R.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2009, 267 (14) : 2361 - 2364
  • [25] Monte Carlo-based patient internal dosimetry in fluoroscopy-guided interventional procedures: A review
    Fum, Wilbur K. S.
    Wong, Jeannie Hsiu Ding
    Tan, Li Kuo
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 84 : 228 - 240
  • [26] Minimizing scattered radiation dose in cardiac catheterization laboratory during interventional procedures using lead free drape-MILD study
    Bhat, Keshavamurthy Ganapathy
    Guleria, Vivek Singh
    Singla, Manish
    Bohra, Vijay
    Kumar, J. Ratheesh
    Bharadwaj, Prashant
    Datta, Rajat
    Hasija, Pradeep Kumar
    INDIAN HEART JOURNAL, 2022, 74 (03) : 201 - 205
  • [27] Polarization imaging of multiply-scattered radiation based on integral-vector Monte Carlo method
    Gay, Benoit
    Vaillon, Rodolphe
    Menguc, M. Pinar
    JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 2010, 111 (02) : 287 - 294
  • [28] Technical Note: A Monte Carlo study of magnetic-field-induced radiation dose effects in mice
    Rubinstein, Ashley E.
    Liao, Zhongxing
    Melancon, Adam D.
    Guindani, Michele
    Followill, David S.
    Tailor, Ramesh C.
    Hazle, John D.
    Court, Laurence E.
    MEDICAL PHYSICS, 2015, 42 (09) : 5510 - 5516
  • [29] A Monte Carlo study of radiation transport through multileaf collimators
    Kim, JO
    Siebers, JV
    Keall, PJ
    Arnfield, MR
    Mohan, R
    MEDICAL PHYSICS, 2001, 28 (12) : 2497 - 2506
  • [30] Study of the gamma radiation fields of irradiators in a calibration laboratory using Monte Carlo codes
    Prado, Julio Dinis Ramalho
    Reynaldo, Sibele Reis
    Lacerda, Marco Aurelio de Sousa
    APPLIED RADIATION AND ISOTOPES, 2025, 220