High Early Failure Rate of the Columbus® Posterior Stabilized High-flexion Knee Prosthesis

被引:19
|
作者
Jung, Woon-Hwa [2 ]
Jeong, Jae-Heon [1 ]
Ha, Yong-Chan [1 ]
Lee, Young-Kyun [3 ]
Koo, Kyung-Hoi [3 ]
机构
[1] Chung Ang Univ, Dept Orthopaed Surg, Coll Med, Seoul 156755, South Korea
[2] Murup Hosp, Dept Orthopaed Surg, Chang Won, South Korea
[3] Seoul Natl Univ, Bundang Hosp, Dept Orthopaed Surg, Songnam, South Korea
关键词
RECURRENT DISLOCATION; CRUCIATE LIGAMENT; FEMORAL COMPONENT; ARTHROPLASTY; REPLACEMENT; DESIGN; MOTION; SYSTEM;
D O I
10.1007/s11999-011-2202-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background Most studies report high survivorship rates for TKAs, however, we observed higher than anticipated rates of dislocation and femoral component loosening after implanting a Columbus (R) posterior stabilized prosthesis. Questions/purpose We therefore determined (1) the incidence of dislocation and aseptic loosening that occurred after implantation of posterior stabilized high-flexion prostheses in TKAs, (2) the causative factors of dislocation and aseptic femoral component loosening when comparing two designs of prostheses, and (3) the mechanisms of dislocation. Methods We retrospectively reviewed 319 patients who underwent 384 TKAs from May 2007 to July 2008. These patients had been assigned alternately to receive a Scorpio (R) posterior stabilized knee prosthesis (Group I, 158 patients, 187 knees) or a Columbus (R) posterior stabilized knee prosthesis (Group II, 161 patients, 197 knees). We followed the patients with clinical and radiographic evaluations for a minimum of 24 months (mean, 26 months; range, 2438 months). Ten retrieved prostheses were examined visually. Results Ten dislocations (5.1%; 10 of 197 knees) and seven aseptic loosenings of femoral components (3.6%; seven of 197 knees) occurred in Group II at a mean of 10.9 months postoperatively. However, no dislocation or loosening occurred in patients in Group I. Most dislocations were associated with varus flexion or flexion rotation movements during normal daily activities. The cam jump distance at 908 flexion for the Columbus (R) prosthesis was lower than for the Scorpio (R) prosthesis. Conclusions We observed a high rate of early failure during short-term followup after implantation of a Columbus (R) posterior stabilized prosthesis. It appears that early failures of the Columbus (R) design were related to a different cam-post design attributable to a low jump distance during knee flexion. We no longer recommend using this device.
引用
收藏
页码:1472 / 1481
页数:10
相关论文
共 50 条
  • [1] Standard Versus High-Flexion Posterior Stabilized Total Knee Prostheses
    Li, Ning
    Li, Junwei
    Li, Peng
    Wang, Dan
    Liu, Ming
    Xia, Lei
    ORTHOPEDICS, 2015, 38 (03) : E206 - E212
  • [2] The Effect of Weight-Bearing Condition on Kinematics of a High-Flexion, Posterior-Stabilized Knee Prosthesis
    Shimizu, Norimasa
    Tomita, Tetsuya
    Yamazaki, Takaharu
    Yoshikawa, Hideki
    Sugamoto, Kazuomi
    JOURNAL OF ARTHROPLASTY, 2011, 26 (07) : 1031 - 1037
  • [3] Early Clinical Results of a High-Flexion, Posterior-Stabilized, Mobile-Bearing Total Knee Arthroplasty
    Scuderi, Giles R.
    Hedden, David R.
    Maltry, John A.
    Traina, Steven M.
    Sheinkop, Mitchell B.
    Hartzband, Mark A.
    JOURNAL OF ARTHROPLASTY, 2012, 27 (03) : 421 - 429
  • [4] Long-term survivorship of total knee arthroplasty with a single-radius, high-flexion posterior stabilized prosthesis
    Baek, Ji-Hoon
    Lee, Su Chan
    Choi, Kyungwon
    Ahn, Hye Sun
    Nam, Chang Hyun
    KNEE, 2021, 30 : 275 - 282
  • [5] In Vivo Kinematic Analysis of a High-Flexion, Posterior-Stabilized, Mobile-Bearing Knee Prosthesis in Deep Knee Bending Motion
    Tamaki, Masashi
    Tomita, Tetsuya
    Watanabe, Tetsu
    Yamazaki, Takaharu
    Yoshikawa, Hideki
    Sugamoto, Kazuomi
    JOURNAL OF ARTHROPLASTY, 2009, 24 (06) : 972 - 978
  • [6] Early results with the Genesis II Posterior Stabilized High Flexion knee prosthesis A one year follow-up study
    Zeh, Alexander
    Davis, John
    Laskin, Richard
    Klima, Stefan
    Wohlrab, David
    ACTA ORTHOPAEDICA BELGICA, 2009, 75 (06): : 792 - 800
  • [7] Is Highly Cross-Linked Polyethylene Safe for Use in High-Flexion Posterior Stabilized Total Knee Arthroplasty?
    Kim, Young-Hoo
    Park, Jang-Won
    Jang, Young-Soo
    Kim, Eun-Jung
    JOURNAL OF ARTHROPLASTY, 2023, 38 (02) : 286 - 292
  • [8] Improved knee flexion following high-flexion total knee arthroplasty
    Lionberger, David R.
    Eggers, Mitchell D.
    Brewer, Kathryn E.
    Fang, Li
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2012, 7
  • [9] Biomechanical analysis of posterior cruciate ligament retaining high-flexion total knee arthroplasty
    Zelle, J.
    Van der Zanden, A. C.
    Malefijt, M. De Waal
    Verdonschot, N.
    CLINICAL BIOMECHANICS, 2009, 24 (10) : 842 - 849
  • [10] Functional Outcome and Range of Motion of High-Flexion Posterior Cruciate-Retaining and High-Flexion Posterior Cruciate-Substituting Total Knee Prostheses A Prospective, Randomized Study
    Kim, Young-Hoo
    Choi, Yoowang
    Kwon, Oh-Ryong
    Kim, Jun-Shik
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (04) : 753 - 760