Screening of renal artery stenosis: Assessment with magnetic resonance angiography at 1.0 T

被引:9
|
作者
Laissy, JP [1 ]
Benyounes, M [1 ]
Limot, O [1 ]
Cinqualbre, A [1 ]
Benamer, H [1 ]
Kenouch, S [1 ]
HenryFeugeas, MC [1 ]
Falise, B [1 ]
Chillon, S [1 ]
Valere, PE [1 ]
SchoumanClaeys, E [1 ]
机构
[1] HOP BICHAT, DEPT RADIOL, F-75877 PARIS 18, FRANCE
关键词
renal arteries; MR studies; stenosis or obstruction; magnetic resonance (MR); comparative studies; vascular studies;
D O I
10.1016/S0730-725X(96)00097-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The results of MR angiography at 1.0 T with digital intraarterial angiography in the screening of patients with suspected renal hypertension were compared. In the first phase of the study, 10 volunteers underwent examination with both two-dimensional (2D) with traveling saturation time-of-flight (TOF) magnetic resonance angiography (MRA) with various parameters to develop a protocol for evaluation of the renal arteries, In the second phase, 36 patients with suspected renovascular hypertension underwent both 2D TOF MRA and intraarterial digital angiography to evaluate the clinical value of MRA. The degree of stenosis was graded with a two-point scale, In volunteers, using 2D acquisitions, C/N ratios indicated the best flip angle as being 55 degrees (p = .02), MRA showed 100% (70/70) of all main arteries and 86% (6/7) of all accessory renal arteries seen on angiography, MRA had a sensitivity of 94% (15/16) and a specificity of 98% (60/61) for detection of stenoses of greater than 50% present in 14 patients. 2D-TOF MRA at 1.0 T shows promise in the noninvasive diagnosis of patients with suspected renovascular hypertension. Copyright (C) 1996 Elsevier Science Inc.
引用
收藏
页码:1033 / 1041
页数:9
相关论文
共 50 条
  • [31] Assessment of gadobenate dimeglumine for magnetic resonance angiography - Phase I studies
    Knopp, MV
    Schoenberg, SO
    Rehm, C
    Floemer, F
    Von Tengg-Kobligk, H
    Bock, M
    Hentrich, HR
    INVESTIGATIVE RADIOLOGY, 2002, 37 (12) : 706 - 715
  • [32] Clinical and experimental investigation of pseudoaneurysm in the anterior communicating artery area on 3-dimensional time-of-flight cerebral magnetic resonance angiography
    Chung, TS
    Lee, YJ
    Kang, WS
    Kang, SK
    Rhim, YC
    Yoo, BG
    Park, IK
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2004, 28 (03) : 414 - 421
  • [33] Contrast-enhanced breath-hold three-dimensional magnetic resonance angiography in the evaluation of renal arteries: Optimization of technique and pitfalls
    Shetty, AN
    Bis, KG
    Kirsch, M
    Weintraub, J
    Laub, G
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2000, 12 (06) : 912 - 923
  • [34] Evaluation of renal artery stenosis:: Comparison of angiography with invasive blood pressure measurement and Doppler ultrasound.
    Sigmund, G
    Hettinger, M
    Blöck, T
    Stölben, A
    Gard, R
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (07): : 615 - 622
  • [35] Quality improvement guidelines for angiography, angioplasty, and stent placement in the diagnosis and treatment of renal artery stenosis in adults
    Martin, LG
    Rundback, JH
    Sacks, D
    Cardella, JF
    Rees, CR
    Matsumoto, AH
    Meranze, SG
    Schwartzberg, MS
    Silverstein, MI
    Lewis, CA
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2002, 13 (11) : 1069 - 1083
  • [37] High-resolution magnetic resonance angiography of the renal arteries using parallel imaging acquisition techniques at 3.0 T - Initial experience
    Kramer, U
    Nael, K
    Laub, G
    Nyborg, GK
    Fenchel, M
    Miller, S
    Claussen, CD
    Finn, JP
    INVESTIGATIVE RADIOLOGY, 2006, 41 (02) : 125 - 132
  • [38] Preoperative demonstration of the Adamkiewicz artery by magnetic resonance angiography in patients with descending or thoracoabdominal aortic aneurysms
    Yamada, N
    Okita, Y
    Minatoya, K
    Tagusari, O
    Ando, M
    Takamiya, M
    Kitamura, S
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2000, 18 (01) : 104 - 111
  • [39] ASSESSMENT OF THE RELIABILITY OF THE DETERMINATION OF CAROTID-ARTERY LUMEN SIZES BY QUANTITATIVE IMAGE-PROCESSING OF MAGNETIC-RESONANCE ANGIOGRAMS AND IMAGES
    BERR, SS
    HURT, NS
    AYERS, CR
    SNELL, JW
    MERICKEL, MB
    MAGNETIC RESONANCE IMAGING, 1995, 13 (06) : 827 - 835
  • [40] Time-of-flight magnetic resonance angiography of aorta and renal arteries: State-of-the-art
    Laissy, JP
    SchoumanClaeys, E
    EUROPEAN JOURNAL OF RADIOLOGY, 1996, 21 (03) : 201 - 211