GAMBE: multipurpose sandwich detector for neutrons and photons

被引:2
|
作者
Ahmed, A. [1 ]
Burdin, S. [1 ]
Casse, G. [1 ]
van Zalinge, H. [2 ]
Powel, S. [1 ]
Rees, J. [1 ]
Smith, A. [1 ]
Tsurin, I. [1 ]
机构
[1] Univ Liverpool, Fac Sci & Engn, Dept Phys, Oxford St, Liverpool L69 7ZE, Merseyside, England
[2] Univ Liverpool, Fac Sci & Engn, Dept Elect Engn & Elect, Brownlow Hill St, Liverpool L69 3GJ, Merseyside, England
关键词
Neutron detector; Semiconductor detector; Neutron detection; Neutron conversion;
D O I
10.1117/12.2236752
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Detectors made with semiconductors such as silicon can be efficiently used for detecting and imaging neutrons when coated with suitable materials. They detect the charged reaction products resulting from the interaction of thermal neutrons with materials with high capture cross section like B-10, Li-6, and (LiF)-Li-6. This work describes the performance of a thermal neutron detector system, GAMBE, which is based on silicon sensors and a layer of neutron-sensitive material, such as a lithium fluoride film or a lithium-6 foil, in a sandwich configuration. This arrangement has a total detection efficiency of 4 +/- 2 %, 7 +/- 1 %, and 12 +/- 1 % for 7 mu m (LiF)-Li-6 film, 40 mu m and 70 mu m Li-6 foil respectively. Also, it enhances the rejection of fake hits using a simple coincidence method. The coincidence that defines a true neutron hit is the simultaneous signal recorded by the two sensors facing the conversion layer (or foil). These coincidences provide a very good method for rejecting the spurious hits coming from gamma-rays, which are usually present in the neutron field under measurement. The GAMBE system yields a rejection factor at the level of 10(8) allowing very pure neutron detection in high gamma background conditions. However, the price to pay is a reduction of the detection efficiency of 1 +/- 1 % or 0.9 +/- 0.3 % for 7 mu m (LiF)-Li-6 film and 40 mu m Li-6 foil respectively.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] PERSONNEL MONITORING OF NEUTRONS, PHOTONS, AND BETA PARTICLES WITH A NEW MULTIPURPOSE BADGE
    DISTENFELD, C
    HEALTH PHYSICS, 1978, 35 (06): : 893 - 893
  • [2] Response of a diamond detector sandwich to 14 MeV neutrons
    Osipenko, M.
    Ripani, M.
    Ricco, G.
    Caiffi, B.
    Pompili, F.
    Pillon, M.
    Verona-Rinati, G.
    Cardarelli, R.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 817 : 19 - 25
  • [3] GAMBE: Thermal neutron detection system based on a sandwich configuration of silicon semiconductor detector coupled with neutron reactive material
    Omar, A.
    Burdin, S.
    Casse, G.
    van Zalinge, H.
    Powel, S.
    Rees, J.
    Smith, A.
    Tsurin, I.
    RADIATION MEASUREMENTS, 2019, 122 : 121 - 125
  • [4] MULTIENERGY NEUTRON DETECTOR FOR COUNTING THERMAL-NEUTRONS, HIGH-ENERGY NEUTRONS, AND GAMMA PHOTONS SEPARATELY
    CHILES, MM
    BAUER, ML
    MCELHANEY, SA
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (03) : 1348 - 1351
  • [5] Neutron spectrometry in a mixed field of neutrons and protons with a phoswich neutron detector - Part I: response functions for photons and neutrons of the phoswich neutron detector
    Takada, M
    Taniguchi, S
    Nakamura, T
    Nakao, N
    Uwamino, Y
    Shibata, T
    Fujitaka, K
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 465 (2-3): : 498 - 511
  • [6] Beating neutrons into photons
    Hutheesing, N
    FORBES, 1996, 158 (06): : 176 - 177
  • [7] GAMBE: Thermal Neutron Detector for Directional Measurement of Neutron Flux
    Ahmed, A.
    Burdin, S.
    Casse, G.
    van Zalinge, H.
    Powel, S.
    Rees, J.
    Smith, A.
    Tsurin, I.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [8] EXPERIMENTS WITH NEUTRONS AND PHOTONS AT ELBE
    Schwengner, R.
    Beyer, R.
    Junghans, A. R.
    Massarczyk, R.
    Schramm, G.
    Bemmerer, D.
    Birgersson, E.
    Ferrari, A.
    Grosse, E.
    Hannaske, R.
    Kempe, M.
    Koegler, T.
    Matic, A.
    Schilling, K. D.
    Wagner, A.
    Rusev, G.
    Makinaga, A.
    Belgya, T.
    Kis, Z.
    Szentmiklosi, L.
    Weil, J.
    Becvar, F.
    Krticka, M.
    CAPTURE GAMMA-RAY SPECTROSCOPY AND RELATED TOPICS, 2013, : 465 - 474
  • [9] Field doses for photons and neutrons
    KeirimMarkus, IB
    Lebedev, VN
    Sannikov, AV
    RADIATION PROTECTION DOSIMETRY, 1996, 63 (03) : 165 - 174
  • [10] Micromegas, a multipurpose gaseous detector
    Charpak, G
    Derré, J
    Giomataris, Y
    Rebourgeard, P
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 478 (1-2): : 26 - 36