Noninvasive Control of the Power Transferred to an Implanted Device by an Ultrasonic Transcutaneous Energy Transfer Link

被引:27
作者
Shmilovitz, Doron [1 ]
Ozeri, Shaul [1 ]
Wang, Chua-Chin [2 ]
Spivak, Boaz [1 ]
机构
[1] Tel Aviv Univ, Sch Elect Engn, IL-69978 Tel Aviv, Israel
[2] Natl Sun Yat Sen Univ, Kaohsiung 80424, Taiwan
关键词
Contactless power transfer; hysteresis control; implanted medical device; ultrasonic transcutaneous energy transfer; voltage regulation; TISSUE; AIR;
D O I
10.1109/TBME.2013.2280460
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ultrasonic transcutaneous energy transfer is an effective method for powering implanted devices noninvasively. Nevertheless, the amount of power harvested by the implanted receiver is sensitive to the distance and orientation of the external transmitting transducer attached to the skin with respect to the implanted receiving transducer. This paper describes an ultrasonic power transfer link whose harvested power is controlled by an inductive link. A small (5 mu F) storage capacitor voltage, which is part of the implanted unit, is allowed to swing between 3.8 and 3.5 V using hysteretic control. The two control states are indicated by excitation (while the implanted storage capacitor voltage decreases) or the absence of excitation of an implanted coil that is magnetically coupled to an external coil attached to the skin surface. A 35 mW Ultrasonic Transcutaneous Energy Transfer link was fabricated using two piezoelectric transducers of equal size (Fuji Ceramics C-2 PZT disc 15 mm x 3 mm) operated at a vibration frequency of 720 kHz. By applying the proposed hysteretic control, the captured power was effectively regulated for implantation depths of up to 85 mm.
引用
收藏
页码:995 / 1004
页数:10
相关论文
共 21 条
  • [11] Ultrasonic transcutaneous energy transfer using a continuous wave 650 kHz Gaussian shaded transmitter
    Ozeri, Shaul
    Shmilovitz, Doron
    Singer, Sigmond
    Wang, Chua-Chin
    [J]. ULTRASONICS, 2010, 50 (07) : 666 - 674
  • [12] Ultrasonic transcutaneous energy transfer for powering implanted devices
    Ozeri, Shaul
    Shmilovitz, Doron
    [J]. ULTRASONICS, 2010, 50 (06) : 556 - 566
  • [13] Roes MGL, 2011, IEEE IND ELEC, P1238
  • [14] Acoustic Energy Transfer: A Review
    Roes, Maurice G. L.
    Duarte, Jorge L.
    Hendrix, Marcel A. M.
    Lomonova, Elena A.
    [J]. IEEE TRANSACTIONS ON INDUSTRIAL ELECTRONICS, 2013, 60 (01) : 242 - 248
  • [15] Shigeta Y, 2009, EURO WIREL TECHNOL C, P49
  • [16] *US DEP HOM SEC, 2008, IMM ENF ACT 2007, P1
  • [17] Wireless Implantable Electronic Platform for Chronic Fluorescent-Based Biosensors
    Valdastri, Pietro
    Susilo, Ekawahyu
    Foerster, Thilo
    Strohhoefer, Christof
    Menciassi, Arianna
    Dario, Paolo
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2011, 58 (06) : 1846 - 1854
  • [18] WANG CC, 2007, P IEEE INT S INT CIR, V1, P604
  • [19] Self-Sampled All-MOS ASK Demodulator for Lower ISM Band Applications
    Wang, Chua-Chin
    Chen, Chih-Lin
    Kuo, Ron-Chi
    Shmilovitz, Doron
    [J]. IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS II-EXPRESS BRIEFS, 2010, 57 (04) : 265 - 269
  • [20] Xue R. F., 2011, P 13 INT S INT CIRC, P224