A Spartan 6 FPGA-based data acquisition system for dedicated imagers in nuclear medicine

被引:14
作者
Fysikopoulos, E. [1 ,2 ]
Loudos, G. [2 ]
Georgiou, M. [2 ,3 ]
David, S. [2 ]
Matsopoulos, G. [1 ]
机构
[1] Natl Tech Univ Athens, Sch Elect & Comp Engn, Athens, Greece
[2] Technol Educ Inst Athens, Dept Med Instruments Technol, Athens, Greece
[3] Univ Thessaly, Sch Med, Dept Nucl Med, Larisa, Greece
关键词
field programmable gate array (FPGA); position-sensitive photomultiplier tube (PSPMT); silicon photomultiplier (SiPM); embedded development kit (EDK); data acquisition (DAQ);
D O I
10.1088/0957-0233/23/12/125403
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
We present the development of a four-channel low-cost hardware system for data acquisition, with application in dedicated nuclear medicine imagers. A 12 bit octal channel high-speed analogue to digital converter, with up to 65 Msps sampling rate, was used for the digitization of analogue signals. The digitized data are fed into a field programmable gate array (FPGA), which contains an interface to a bank of double data rate 2 (DDR2)-type memory. The FPGA processes the digitized data and stores the results into the DDR2. An ethernet link was used for data transmission to a personal computer. The embedded system was designed using Xilinx's embedded development kit (EDK) and was based on Xilinx's Microblaze soft-core processor. The system has been evaluated using two different discrete optical detector arrays (a position-sensitive photomultiplier tube and a silicon photomultiplier) with two different pixelated scintillator arrays (BGO, LSO:Ce). The energy resolution for both detectors was approximately 25%. A clear identification of all crystal elements was achieved in all cases. The data rate of the system with this implementation can reach 60 Mbits s(-1). The results have shown that this FPGA data acquisition system is a compact and flexible solution for single-photon-detection applications.
引用
收藏
页数:5
相关论文
共 18 条
  • [11] SensL, 2011, ARRAYSL 4 SCAL SIL P
  • [12] Pulse recording by free-running sampling
    Streun, M
    Brandenburg, G
    Larue, H
    Zimmermann, E
    Ziemons, K
    Halling, H
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (03) : 524 - 526
  • [13] Texas Instruments, 2010, 5 10 OUTP CLOCK GEN
  • [14] Xilinx Inc, 2009, MULT MEM CONTR MPMC
  • [15] Xilinx Inc, 2010, SP601 HARDW US GUID
  • [16] Xilinx Inc, 2009, XPS LL TEMAC V2 03A
  • [17] Xilinx Inc, 2009, CHIPSCOPE PRO 11 4 S
  • [18] Xilinx Inc, 2009, EDK CONC TOOLS TECHN