Design of ridge filters for spread-out Bragg peaks with Monte Carlo simulation in carbon ion therapy

被引:27
|
作者
Sakama, M. [1 ,2 ]
Kanai, T. [2 ]
Kase, Y. [3 ]
Yusa, K. [2 ]
Tashiro, M. [2 ]
Torikai, K. [2 ]
Shimada, H. [2 ]
Yamada, S. [2 ]
Ohno, T. [2 ]
Nakano, T. [2 ]
机构
[1] Nihon Univ, Dept Elect & Elect Engn, Coll Ind Technol, Narashino, Chiba 2758575, Japan
[2] Gunma Univ, Heavy Ion Med Res Ctr, Maebashi, Gunma 3718511, Japan
[3] Shizuoka Canc Ctr, Proton Therapy Div, Res Inst, Nagaizumi, Shizuoka 4118777, Japan
关键词
ACCURATE UNIVERSAL PARAMETERIZATION; ABSORPTION CROSS-SECTIONS; IRRADIATION SYSTEM; PROTON THERAPY; BEAM; GEANT4; MODELS; INJECTOR; AIR;
D O I
10.1088/0031-9155/57/20/6615
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Spread-out Bragg peaks made by ridge filters or wheel range modulators are used in charged particle therapy with passive methods to achieve uniform biological responses in irradiated tumors. Following the biological responses needed to design the ridge filters, which were developed at the National Institute of Radiological Sciences in Japan, new ridge filters were designed using recent developments in heavy-ion reactions and dosimetry. The Monte Carlo code of Geant4 was used to calculate the qualities of carbon ion beams in a water phantom. The results obtained from the simulation were corrected so that they agreed with the measurements of depth dose distributions. The calculations of biological responses to fragments other than carbon ions were assumed to be for helium ions. The measured dose distributions with the designed ridge filters were compared to the calculated distributions. A beam modifying system using this adaptable method was successively applied to carbon ion therapy at Gunma University.
引用
收藏
页码:6615 / 6633
页数:19
相关论文
共 22 条
  • [1] Comparison of methodologies for creating spread-out Bragg peaks in proton therapy using TOPAS and MCNP codes
    Branco, I. S. L.
    Burin, A. L.
    Pereira, J. J. N.
    Siqueira, P. T. D.
    Shorto, J. M. B.
    Yoriyaz, H.
    RADIATION PHYSICS AND CHEMISTRY, 2023, 211
  • [2] Simulation of spread-out bragg peaks in proton beams using Geant4/TOPAS
    Velten, Christian
    Tome, Wolfgang A.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (04):
  • [3] Secondary particle production and physical properties during dose enhancement for spread-out Bragg peaks
    Hwang, Chulhwan
    Kim, Jung Hoon
    TRANSLATIONAL CANCER RESEARCH, 2019, 8 (04) : 1449 - 1456
  • [4] Designing a range modulator wheel to spread-out the Bragg peak for a passive proton therapy facility
    Jia, S. Bijan
    Romano, F.
    Cirrone, Giuseppe A. P.
    Cuttone, G.
    Hadizadeh, M. H.
    Mowlavi, A. A.
    Raffaele, L.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 806 : 101 - 108
  • [5] "Broadscale" nanodosimetry: Nanodosimetric track structure quantities increase at distal edge of spread-out proton Bragg peaks
    Rabus, Hans
    Ngcezu, Sonwabile A.
    Braunroth, Thomas
    Nettelbeck, Heidi
    RADIATION PHYSICS AND CHEMISTRY, 2020, 166
  • [6] Calculation method using Clarkson integration for the physical dose at the center of the spread-out Bragg peak in carbon-ion radiotherapy
    Tajiri, Minoru
    Maeda, Takamasa
    Koba, Yusuke
    Isobe, Yoshiharu
    Kuroiwa, Toshitaka
    Fukuda, Shigekazu
    Tanimoto, Katsuyuki
    Shibayama, Koichi
    MEDICAL PHYSICS, 2013, 40 (07)
  • [7] Evaluation of the effect of soft tissue composition on the characteristics of spread-out Bragg peak in proton therapy
    Ghorbani, Mahdi
    Jia, Sayyed Bijan
    Khosroabadi, Mohsen
    Sadoughi, Hamid Reza
    Knaup, Courtney
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2017, 13 (06) : 974 - 980
  • [8] Spread-out Bragg peak proton FLASH irradiation using a clinical synchrocyclotron: Proof of concept and ion chamber characterization
    Darafsheh, Arash
    Hao, Yao
    Zhao, Xiandong
    Zwart, Townsend
    Wagner, Miles
    Evans, Tucker
    Reynoso, Francisco
    Zhao, Tianyu
    MEDICAL PHYSICS, 2021, : 4472 - 4484
  • [9] Semianalytical tool for spread-out Bragg peak optimization in an active energy-modulation proton therapy linac
    Ronsivalle, C.
    Nichelatti, E.
    Picardi, L.
    Montereali, R. M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2021, 988
  • [10] Correction method for the physical dose calculated using Clarkson integration at the center of the spread-out Bragg peak for asymmetric field in carbon-ion radiotherapy
    Tajiri, Minoru
    Maeda, Takamasa
    Isobe, Yoshiharu
    Tanimoto, Katsuyuki
    Shibayama, Koichi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (08): : 985 - 988