Short-term effects experienced during examinations in an actively shielded 7 T MR

被引:12
作者
Hansson, Boel [1 ,2 ]
Hoglund, Peter [3 ]
Bloch, Karin Markenroth [4 ]
Nilsson, Markus [2 ]
Olsrud, Johan [1 ,2 ]
Wilen, Jonna [5 ]
Bjorkman-Burtscher, Isabella M. [1 ,2 ,6 ]
机构
[1] Skane Univ Hosp, Dept Diagnost Radiol, SE-22185 Lund, Sweden
[2] Lund Univ, Dept Diagnost Radiol, Clin Sci, Lund, Sweden
[3] Lund Univ, Dept Clin Pharmacol, Lund, Sweden
[4] Lund Univ, Lund Univ Bioimaging Ctr LBIC, Lund, Sweden
[5] Umea Univ, Dept Radiat Sci, Umea, Sweden
[6] Univ Gothenburg, Sahlgrenska Acad, Inst Clin Sci, Dept Radiol, Gothenburg, Sweden
关键词
dizziness; peripheral nerve stimulation; motion; patient compliance; bias; STATIC MAGNETIC-FIELDS; OCCUPATIONAL-EXPOSURE; SUBJECTIVE ACCEPTANCE; STRAY FIELDS; SCANNERS; VERTIGO; SAFETY; TESLA;
D O I
10.1002/bem.22189
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
The objective of this study was to evaluate occurrence and strength of short-term effects experienced by study participants in an actively shielded (AS) 7 tesla (7 T) magnetic resonance (MR) scanner, to compare results with earlier reports on passively shielded (PS) 7 T MR scanners, and to outline possible healthcare strategies to improve patient compliance. Study participants (n = 124) completed a web-based questionnaire directly after being examined in an AS 7 T MR (n = 154 examinations). Most frequently experienced short-term effects were dizziness (84%) and inconsistent movement (70%), especially while moving into or out of the magnet. Peripheral nerve stimulation (PNS)-twitching-was experienced in 67% of research examinations and showed a dependence between strength of twitches and recorded predicted PNS values. Of the participants, 74% experienced noise levels as acceptable and the majority experienced body and room temperature as comfortable. Of the study participants, 95% felt well-informed and felt they had had good contact with the staff before the examination. Willingness to undergo a future 7 T examination was high (>90%). Our study concludes short-term effects are often experienced during examinations in an AS 7 T MR, leaving room for improvement in nursing care strategies to increase patient compliance. Bioelectromagnetics. 2019;9999:XX-XX. (c) 2019 The Authors. Bioelectromagnetics Published by Wiley Periodicals, Inc.
引用
收藏
页码:234 / 249
页数:16
相关论文
共 38 条
[1]   Static magnetic field effects on human subjects related to magnetic resonance imaging systems [J].
Chakeres, DW ;
de Vocht, F .
PROGRESS IN BIOPHYSICS & MOLECULAR BIOLOGY, 2005, 87 (2-3) :255-265
[2]   Effect of static magnetic field exposure of up to 8 Tesla on sequential human vital sign measurements [J].
Chakeres, DW ;
Kangarlu, A ;
Boudoulas, H ;
Young, DC .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 18 (03) :346-352
[3]   Randomized comparison of cognitive function in humans at 0 and 8 Tesla [J].
Chakeres, DW ;
Bornstein, R ;
Kangarlu, A .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 18 (03) :342-345
[4]   Short-term side-effects of brain MR examination at 7 T: a single-centre experience [J].
Cosottini, M. ;
Frosini, D. ;
Biagi, L. ;
Pesaresi, I. ;
Costagli, M. ;
Tiberi, G. ;
Symms, M. ;
Tosetti, M. .
EUROPEAN RADIOLOGY, 2014, 24 (08) :1923-1928
[5]   Prediction of peripheral nerve stimulation thresholds of MRI gradient coils using coupled electromagnetic and neurodynamic simulations [J].
Davids, Mathias ;
Guerin, Bastien ;
vom Endt, Axel ;
Schad, Lothar R. ;
Wald, Lawrence L. .
MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (01) :686-701
[6]  
Glover P., 2015, CURR RADIOL REP, V3, P29, DOI DOI 10.1007/S40134-015-0112-1
[7]   Magnetic-field-induced vertigo: A theoretical and experimental investigation [J].
Glover, P. M. ;
Cavin, I. ;
Qian, W. ;
Bowtell, R. ;
Gowland, P. A. .
BIOELECTROMAGNETICS, 2007, 28 (05) :349-361
[8]   Interaction of MRI field gradients with the human body [J].
Glover, P. M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (21) :R99-R115
[9]   A dynamic model of the eye nystagmus response to high magnetic fields [J].
Glover, Paul M. ;
Li, Yan ;
Antunes, Andre ;
Mian, Omar S. ;
Day, Brian L. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (03) :631-645
[10]   CONSIDERATIONS IN THE DESIGN OF MRI MAGNETS WITH REDUCED STRAY FIELDS [J].
HAWKSWORTH, DG ;
MCDOUGALL, IL ;
BIRD, JM ;
BLACK, D .
IEEE TRANSACTIONS ON MAGNETICS, 1987, 23 (02) :1309-1314